Comparison of body weight and composition measured by two different dual energy X-ray absorptiometry devices and three acquisition modes in obese women

被引:25
作者
Genton, Laurence [1 ]
Karsegard, Veronique L. [1 ]
Zawadynski, Sophie [1 ]
Kyle, Ursula G. [1 ]
Pichard, Claude [1 ]
Golay, Alain [1 ]
Hans, Didier B. [1 ]
机构
[1] Univ Hosp Geneva, CH-1211 Geneva 14, Switzerland
关键词
body composition; DXA; inter-device differences; reproducibility; obesity;
D O I
10.1016/j.clnu.2005.11.002
中图分类号
R15 [营养卫生、食品卫生]; TS201 [基础科学];
学科分类号
100403 ;
摘要
Background & aims: Weight measured by dual-energy X-ray (DXA) was shown to be increasingly underestimated in subjects over 75 kg compared to an electronic scale. This study compares body weight and composition measured by balance beam scale and three DXA acquisition modes in obese subjects. Methods: In 39 obese, body weight was measured by balance beam scale, and body weight and composition by DXA Hologic QDR450OA(R) in normal (NPM) and high power mode (HPM) (Enhanced v8.26 and v8.26* softwares) and DXA GE-Lunar Prodigy(R) (v6.5 software). To ensure linearity of body weight and composition measured by the different DXA acquisitions, we also measured 13 women with a body mass index (BMI) of 25-30 kg/m(2). Results: While QDR4500A HPM overestimates scale weight by about 2 kg over the whole BMI spectrum, QDR4500A NPM underestimates scale weight as a weight-dependent response (-1.7 +/- 1.8 kg overall, -4.1 +/- 1.6 kg in morbidly obese women). These results suggest switching from one mode to the other at a specific threshold, i.e. in our study a weight of 90 kg or a BMI of 34 kg/m(2). Prodigy gives weight about similar to scale (+0.5 +/- 0.8 kg). Both Hologic acquisition modes underestimate fat mass but overestimate lean body mass compared to Prodigy. Conclusions: The QDR4500A NPM is inappropriate in women over 90 kg. Unfortunately, the QDR4500A HPM overestimates body weight in the range of 90-150 kg. The difference between scale and Prodigy weight remains stable throughout weight ranges. To better assess their accuracies in terms of body composition, QDR4500A NPM, HPM and Prodigy should be tested against phantoms or in vivo multi-compartment models. (C) 2005 Elsevier Ltd and European Society for Clinical Nutrition and Metabolism. All rights reserved.
引用
收藏
页码:428 / 437
页数:10
相关论文
共 22 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   Reproducibility of DXA in obese women [J].
Cordero-MacIntyre, ZR ;
Peters, W ;
Libanati, CR ;
España, RC ;
Abila, SO ;
Howell, WH ;
Lohman, TG .
JOURNAL OF CLINICAL DENSITOMETRY, 2002, 5 (01) :35-44
[3]   Bone mineral measurements: A comparison of delayed gamma neutron activation, dual-energy X-ray absorptiometry and direct chemical analysis [J].
Economos, CD ;
Nelson, ME ;
Singh, MAF ;
Kehayias, JJ ;
Dallal, GE ;
Heymsfield, SB ;
Wang, J ;
Yasumura, S ;
Ma, R ;
Pierson, RN .
OSTEOPOROSIS INTERNATIONAL, 1999, 10 (03) :200-206
[4]   Bone mineral and body composition measurements: Cross-calibration of pencil-beam and fan-beam dual-energy X-ray absorptiometers [J].
Ellis, KJ ;
Shypailo, RJ .
JOURNAL OF BONE AND MINERAL RESEARCH, 1998, 13 (10) :1613-1618
[5]   Dual-energy x-ray absorptiometry and body composition: Differences between devices and comparison with reference methods [J].
Genton, L ;
Hans, D ;
Kyle, UG ;
Pichard, C .
NUTRITION, 2002, 18 (01) :66-70
[6]  
GENTON L, 2003, CLIN NUTR, V21, P22
[7]   Assessment of research-based body composition techniques in healthy elderly men and women using the 4-compartment model as a criterion method [J].
Goran, MI ;
Toth, MJ ;
Poehlman, ET .
INTERNATIONAL JOURNAL OF OBESITY, 1998, 22 (02) :135-142
[8]   Comparison of dual-energy x-ray absorptiometry to four other methods to determine body composition in underweight patients with chronic gastrointestinal disease [J].
Haderslev, KV ;
Staun, M .
METABOLISM-CLINICAL AND EXPERIMENTAL, 2000, 49 (03) :360-366
[9]   DUAL-ENERGY X-RAY ABSORPTIOMETRY MEASUREMENTS OF BODY-COMPOSITION - EFFECTS OF DEPTH AND TISSUE THICKNESS, INCLUDING COMPARISONS WITH DIRECT ANALYSIS [J].
JEBB, SA ;
GOLDBERG, GR ;
JENNINGS, G ;
ELIA, M .
CLINICAL SCIENCE, 1995, 88 (03) :319-324
[10]   Body composition determined by six different methods in women bilaterally adrenalectomized for treatment of Cushing's disease [J].
Kemink, SAG ;
Frijns, JTM ;
Hermus, ARMM ;
Pieters, GFFM ;
Smals, AGH ;
Lichtenbelt, WDV .
JOURNAL OF CLINICAL ENDOCRINOLOGY & METABOLISM, 1999, 84 (11) :3991-3999