Accuracy of Infrared Ear Thermometry in Children: A Meta-Analysis and Systematic Review

被引:22
作者
Zhen, Chen [1 ,2 ]
Xia, Zhang [2 ]
Long, Li [1 ]
Pu, Yu [1 ]
机构
[1] Capital Inst Pediat, Beijing 100000, Peoples R China
[2] Peking Univ, Beijing 100871, Peoples R China
关键词
infrared ear thermometry; rectal thermometry; children; meta-analysis; systematic review; TYMPANIC MEMBRANE; TEMPERATURE-MEASUREMENTS; BODY-TEMPERATURE; RECTAL TEMPERATURES; AXILLARY; FEVER; INFANTS; AGE;
D O I
10.1177/0009922814536774
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Background. Accurate determination of temperature is important, especially in the diagnosis and treatment of febrile illnesses in the pediatric population. False negative measurement can lead to miss febrile and false positive measurement can cause excessive medical care. Temperatures can be measured at various sites, but we have not found the ideal thermometer yet. As a relatively new and popular alternative over traditional methods, infrared ear thermometry has many advantages, but its accuracy remains a major concern. Design. Systematic review and meta-analysis. Data Sources. Medline, Ovid, Elsevier, Google Scholar, Cochrane Library. Study Selection. Cross-sectional, prospective design. Data Extraction. Two investigators independently assessed selected studies and extracted data. Disagreements were resolved by discussion with other reviewers. Results. Mean tympanic temperature was always lower than rectal temperature. The overall pooled (random effects) mean difference between tympanic and rectal temperature was 0.22 degrees C (95% limits of agreement -0.44 degrees C to 1.30 degrees C), which is similar to the within rectal device groups (mercury, 0.21 degrees C, -0.44 degrees C to 1.27 degrees C; electronic, 0.24 degrees C, -0.46 degrees C to 1.34 degrees C). In febrile children group, the pooled mean temperature difference between tympanic and rectal temperature was 0.15 degrees C (95% limits of agreement -0.32 degrees C to 1.10 degrees C). Conclusion. The mean difference was large and the 95% limits of agreement was wide. The accuracy of infrared ear thermometry in children is poor, and it cannot replace rectal thermometry in clinical practice of children.
引用
收藏
页码:1158 / 1165
页数:8
相关论文
共 47 条
[1]   A comparative study of rectal tympanic and axillary thermometry in febrile children under 5 years of age in Nigeria [J].
Abdulkadir, Mohammed Baba ;
Johnson, Wahab Babatunde Rotimi .
PAEDIATRICS AND INTERNATIONAL CHILD HEALTH, 2013, 33 (03) :165-169
[2]   Tympanic thermometry in Nigerian children [J].
Akinyinka, OO ;
Omokhodion, SI ;
Olawuyi, JF ;
Olumese, PE ;
Brown, BJ .
ANNALS OF TROPICAL PAEDIATRICS, 2001, 21 (02) :169-174
[3]   Comparison of Rectal, Axillary, Tympanic, and Temporal Artery Thermometry in the Pediatric Emergency Room [J].
Batra, Prerna ;
Goyal, Sudhanshu .
PEDIATRIC EMERGENCY CARE, 2013, 29 (01) :63-66
[4]  
Beckstrand R L, 1996, Pediatr Nurs, V22, P436
[5]  
Bernardo L M, 1999, J Emerg Nurs, V25, P327, DOI 10.1016/S0099-1767(99)70063-2
[6]  
Bernardo L M, 1996, J Emerg Nurs, V22, P403, DOI 10.1016/S0099-1767(96)80160-7
[7]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[8]   RELIABILITY OF INFRARED TYMPANIC THERMOMETRY IN THE DETECTION OF RECTAL FEVER IN CHILDREN [J].
BRENNAN, DF ;
FALK, JL ;
ROTHROCK, SG ;
KERR, RB .
ANNALS OF EMERGENCY MEDICINE, 1995, 25 (01) :21-30
[9]   Temperature measured at the axilla compared with rectum in children and young people: systematic review [J].
Craig, JV ;
Lancaster, GA ;
Williamson, PR ;
Smyth, RL .
BRITISH MEDICAL JOURNAL, 2000, 320 (7243) :1174-1178
[10]   Infrared ear thermometry compared with rectal thermometry in children: a systematic review [J].
Craig, JV ;
Lancaster, GA ;
Taylor, S ;
Williamson, PR ;
Smyth, RL .
LANCET, 2002, 360 (9333) :603-609