Validation of the Hingmed WBP-02A device for ambulatory blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010

被引:2
作者
Fania, Claudio [1 ]
Vezzu, Leonardo [1 ]
Lazzaretto, Ilaria [1 ]
Palatini, Paolo
机构
[1] Casa Cura Villa Maria Hosp, Padua, Italy
关键词
ambulatory; blood pressure; device; validation;
D O I
10.1097/MBP.0000000000000371
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
The objective of this study was to determine the accuracy of the Hingmed WBP-02A device for ambulatory blood pressure (BP) monitoring according to the International Protocol of the European Society of Hypertension (ESH-IP). Device evaluation was carried out in 33 patients. The mean age was 69.8 +/- 13.8 years, the mean systolic BP was 137.6 +/- 27.0 mmHg (range: 96-172), the mean diastolic BP was 86.7 +/- 20.1 mmHg (range: 48-124) and the mean arm circumference was 30.1 +/- 6.2 cm (range: 20-43). The protocol requirements were followed precisely. Four different cuffs, small, standard, large, and extra-large were used. The device comfortably passed all ESH-IP requirements, as 84 out of 99 device-observer systolic BP differences and 97 of 99 diastolic BP differences were within +/- 5 mmHg thereby fulfilling the standards of the protocol. Device-observer BP differences were unrelated to arm circumference. These data show that the Hingmed WBP-02A device for ambulatory BP monitoring met the requirements for validation by the ESH-IP and can be recommended for clinical use in the adult population. Copyright (C) 2019 Wolters Kluwer Health, Inc. All rights reserved.
引用
收藏
页码:151 / 154
页数:4
相关论文
共 50 条
  • [41] Validation of the Pangao PG-800B5 for clinical use and self-measurement according to the European Society of Hypertension International Protocol revision 2010
    Chen, Wan
    Zeng, Zhaolin
    Li, Lizhi
    Wan, Xiaofen
    Wan, Yi
    [J]. BLOOD PRESSURE MONITORING, 2014, 19 (05) : 302 - 305
  • [42] Validation of the Grandway MD2301 digital automatic blood pressure monitor according to the European Society of Hypertension International Protocol
    Chen, Wan
    Zeng, Zhao-Lin
    Bing, Sen
    Li, Lin-Yi
    Wang, Rui
    Wan, Yi
    [J]. BLOOD PRESSURE MONITORING, 2016, 21 (04) : 259 - 261
  • [43] Validation of the Lloyds Pharmacy BP11 oscillometric blood pressure monitor according to the International Protocol of the European Society of Hypertension
    Coleman, Andrew
    Steel, Stephen
    DeGreef, Annemarie
    Shennan, Andrew
    [J]. BLOOD PRESSURE MONITORING, 2010, 15 (03) : 163 - 166
  • [44] Validation of the Andon KD-575 automated blood pressure monitor according to the European Society of Hypertension International Protocol in adults
    Pan, Feng
    Wan, Yi
    Chen, Wan
    Li, Ni
    Tong, Wei
    Xu, Yongyong
    [J]. BLOOD PRESSURE MONITORING, 2010, 15 (03) : 167 - 170
  • [45] Validation of the Omron HBP-1320 for professional use according to the ANSI/AAMI/ISO 81060-2:2013 protocol and the 2010 revision of the European Society of Hypertension International Protocol
    Saito, Kanako
    Hishiki, Yukiko
    Takahashi, Hakuo
    [J]. BLOOD PRESSURE MONITORING, 2020, 25 (03) : 162 - 166
  • [46] Performance of the UA-787 oscillometric blood pressure monitor according to the European Society of Hypertension protocol
    Longo, D
    Toffanin, G
    Garbelotto, R
    Zaetta, V
    Businaro, L
    Palatini, P
    [J]. BLOOD PRESSURE MONITORING, 2003, 8 (02) : 91 - 95
  • [47] Validation of the PHILIPS DL8760 upper arm blood pressure monitor, in oscillometry mode, for self-measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010
    Chahine, Mirna N.
    Bou Harb, Souad
    Saad, Abdel Rahman
    Sarkis, Patrick
    Azaki, Alaa
    Harb, Aya
    Allouch, Ali
    Asmar, Roland
    [J]. BLOOD PRESSURE MONITORING, 2021, 26 (03) : 237 - 241
  • [48] Validation of the Microlife WatchBP O3 device for clinic, home, and ambulatory blood pressure measurement, according to the International Protocol
    Ragazzo, Fabio
    Saladini, Francesca
    Palatini, Paolo
    [J]. BLOOD PRESSURE MONITORING, 2010, 15 (01) : 59 - 62
  • [49] Validation of the Omron HEM-7201 upper arm blood pressure monitor, for self-measurement in a high-altitude environment, according to the European Society of Hypertension International Protocol revision 2010
    K Cho
    M Tian
    Y Lan
    X Zhao
    L L Yan
    [J]. Journal of Human Hypertension, 2013, 27 : 487 - 491
  • [50] Validation of the DINAMAP ProCare blood pressure device according to the international protocol in an adult population
    Reinders, Annemarie
    Reggiori, Fayrouge
    Shennan, Andrew H.
    [J]. BLOOD PRESSURE MONITORING, 2006, 11 (05) : 293 - 296