Validation of the Hingmed WBP-02A device for ambulatory blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010

被引:2
作者
Fania, Claudio [1 ]
Vezzu, Leonardo [1 ]
Lazzaretto, Ilaria [1 ]
Palatini, Paolo
机构
[1] Casa Cura Villa Maria Hosp, Padua, Italy
关键词
ambulatory; blood pressure; device; validation;
D O I
10.1097/MBP.0000000000000371
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
The objective of this study was to determine the accuracy of the Hingmed WBP-02A device for ambulatory blood pressure (BP) monitoring according to the International Protocol of the European Society of Hypertension (ESH-IP). Device evaluation was carried out in 33 patients. The mean age was 69.8 +/- 13.8 years, the mean systolic BP was 137.6 +/- 27.0 mmHg (range: 96-172), the mean diastolic BP was 86.7 +/- 20.1 mmHg (range: 48-124) and the mean arm circumference was 30.1 +/- 6.2 cm (range: 20-43). The protocol requirements were followed precisely. Four different cuffs, small, standard, large, and extra-large were used. The device comfortably passed all ESH-IP requirements, as 84 out of 99 device-observer systolic BP differences and 97 of 99 diastolic BP differences were within +/- 5 mmHg thereby fulfilling the standards of the protocol. Device-observer BP differences were unrelated to arm circumference. These data show that the Hingmed WBP-02A device for ambulatory BP monitoring met the requirements for validation by the ESH-IP and can be recommended for clinical use in the adult population. Copyright (C) 2019 Wolters Kluwer Health, Inc. All rights reserved.
引用
收藏
页码:151 / 154
页数:4
相关论文
共 50 条
  • [31] Validation of the iHealth BP3 upper-arm blood pressure monitor, for clinic use and self-measurement, according to the European Society of Hypertension International Protocol revision 2010
    Chen, Can
    Shang, Fujun
    Wang, Jiepin
    Chen, Jianghong
    Ji, Na
    Wan, Yi
    BLOOD PRESSURE MONITORING, 2012, 17 (06) : 253 - 256
  • [32] Validation of the Medipro MediCare 100f upper arm blood pressure monitor, for self-measurement, according to the European Society of Hypertension International Protocol revision 2010
    Yi, Jun
    Wan, Yi
    Pan, Feng
    Yu, Xiaorong
    Zhao, Huadong
    Shang, Fujun
    Xu, Yongyong
    BLOOD PRESSURE MONITORING, 2011, 16 (04) : 197 - 199
  • [33] Validation of the UEBE Visomat Double Comfort upper arm blood pressure monitor, in auscultation mode, for clinic use and self-measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010
    Masiero, Serena
    Fania, Claudio
    Palatini, Paolo
    BLOOD PRESSURE MONITORING, 2011, 16 (04) : 208 - 210
  • [34] Validation of the UEBE Visomat Double Comfort upper arm blood pressure monitor, in oscillometric mode, for clinic use and self-measurement in a general population according to the European Society of Hypertension International Protocol, revision 2010
    Masiero, Serena
    Fania, Claudio
    Palatini, Paolo
    BLOOD PRESSURE MONITORING, 2011, 16 (05) : 262 - 264
  • [35] Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults
    O'Brien, E
    Pickering, T
    Asmar, R
    Myers, M
    Parati, G
    Staessen, J
    Mengden, T
    Imai, Y
    Waeber, B
    Palatini, P
    Atkins, N
    Gerin, W
    BLOOD PRESSURE MONITORING, 2002, 7 (01) : 3 - 17
  • [36] European Society of Hypertension International Protocol for the validation of blood pressure monitors: a critical review of its application and rationale for revision
    Stergiou, George S.
    Karpettas, Nikos
    Atkins, Neil
    O'Brien, Eoin
    BLOOD PRESSURE MONITORING, 2010, 15 (01) : 39 - 48
  • [37] Validation of four automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension
    Belghazi, Jalil
    El Feghali, Ramzi N.
    Moussalem, Therese
    Rejdych, Maya
    Asmar, Roland G.
    VASCULAR HEALTH AND RISK MANAGEMENT, 2007, 3 (04) : 389 - 400
  • [38] Validation of the Artsana CS 410 automated blood pressure monitor in adults according to the International Protocol of the European Society of Hypertension
    Pini, Claudio
    Natalizi, Anna
    Gerosa, Pietro Francesco
    Frigerioa, Mauro
    Omboni, Stefano
    Parati, Gianfranco
    BLOOD PRESSURE MONITORING, 2008, 13 (03) : 177 - 182
  • [39] Validation of the Artsana CSI 610 automated blood pressure monitor in adults according to the International Protocol of the European Society of Hypertension
    Pini, Claudio
    Pastori, Marco
    Baccheschi, Jordan
    Omboni, Stefano
    Parati, Gianfranco
    BLOOD PRESSURE MONITORING, 2007, 12 (03) : 179 - 184
  • [40] Validation of four automatic devices for self-measurement of blood pressure according to the international protocol of the European Society of Hypertension
    Topouchian, Jirar
    Agnoletti, Davide
    Blacher, Jacques
    Youssef, Ahmed
    Ibanez, Isabel
    Khabouth, Jose
    Khawaja, Salwa
    Beaino, Layale
    Asmar, Roland
    VASCULAR HEALTH AND RISK MANAGEMENT, 2011, 7 : 709 - 717