Separation of plasma from whole human blood in a continuous cross-flow in a molded microfluidic device

被引:180
作者
VanDelinder, Virginia [1 ]
Groisman, Alex [1 ]
机构
[1] Univ Calif San Diego, Dept Phys, La Jolla, CA 92093 USA
关键词
D O I
10.1021/ac060042r
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
We designed, fabricated, and tested a microfluidic device for separation of plasma from whole human blood by size exclusion in a cross-flow. The device is made of a single mold of a silicone elastomer poly( dimethylsiloxane) ( PDMS) sealed with a cover glass and is essentially disposable. When loaded with blood diluted to 20% hematocrit and driven with pulsatile pressure to prevent clogging of the channels with blood cells, the device can operate for at least 1 h, extracting similar to 8% of blood volume as plasma at an average rate of 0.65 mu L/min. The flow in the device causes very little hemolysis; the extracted plasma meets the standards for common assays and is delivered to the device outlet similar to 30 s after injection of blood to the inlet. Integration of the cross-flow microchannel array with on-chip assay elements would create a microanalysis system for point-of-care diagnostics, reducing costs, turn-around times, and volumes of blood sample and reagents required for the assays.
引用
收藏
页码:3765 / 3771
页数:7
相关论文
共 50 条
[31]   A microfluidic plasma separation device combined with a surface plasmon resonance biosensor for biomarker detection in whole blood [J].
Debnath, Nandini ;
Live, Ludovic Saiveng ;
Poudineh, Mahla .
LAB ON A CHIP, 2023, 23 (04) :572-579
[32]   Enhancement of microfluidic particle separation using cross-flow filters with hydrodynamic focusing [J].
Chiu, Yun-Yen ;
Huang, Chen-Kang ;
Lu, Yen-Wen .
BIOMICROFLUIDICS, 2016, 10 (01)
[33]   Continuous High-Throughput Plasma Separation for Blood Biomarker Sensing Using a Hydrodynamic Microfluidic Device [J].
Abouali, Hesam ;
Keyvani, Fatemeh ;
Hosseini, Seied Ali ;
Srikant, Sanjana ;
Poudineh, Mahla .
ADVANCED HEALTHCARE MATERIALS, 2025, 14 (09)
[34]   Establishment of CFD model for rotating cross-flow membrane separation device [J].
Chen, Zhihao ;
Li, Liuhuan ;
Cheng, Dazhou .
2018 INTERNATIONAL CONFERENCE ON CONSTRUCTION, AVIATION AND ENVIRONMENTAL ENGINEERING, 2019, 233
[35]   Microfluidic immunomagnetic cell separation from whole blood [J].
Gourikutty, Sajay Bhuvanendran Nair ;
Chang, Chia-Pin ;
Puiu, Poenar Daniel .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2016, 1011 :77-88
[36]   Disposable filtration device for plasma separation from whole blood clinical samples [J].
Fomovska, GN ;
Seeley, K .
CLINICAL CHEMISTRY, 2005, 51 :A205-A206
[37]   Pinched flow fractionation (PFF) for continuous particle separation in a microfluidic device [J].
Nakashima, M ;
Yamada, M ;
Seki, M .
MEMS 2004: 17TH IEEE INTERNATIONAL CONFERENCE ON MICRO ELECTRO MECHANICAL SYSTEMS, TECHNICAL DIGEST, 2004, :33-36
[38]   Integrated separation of blood plasma from whole blood for microfluidic paper-based analytical devices [J].
Yang, Xiaoxi ;
Forouzan, Omid ;
Brown, Theodore P. ;
Shevkoplyas, Sergey S. .
LAB ON A CHIP, 2012, 12 (02) :274-280
[39]   From cooperative to uncorrelated clogging in cross-flow microfluidic membranes [J].
R. van Zwieten ;
T. van de Laar ;
J. Sprakel ;
K. Schroën .
Scientific Reports, 8
[40]   From cooperative to uncorrelated clogging in cross-flow microfluidic membranes [J].
van Zwieten, R. ;
van de laar, T. ;
Sprakel, J. ;
Schroen, K. .
SCIENTIFIC REPORTS, 2018, 8