Separation of plasma from whole human blood in a continuous cross-flow in a molded microfluidic device

被引:178
|
作者
VanDelinder, Virginia [1 ]
Groisman, Alex [1 ]
机构
[1] Univ Calif San Diego, Dept Phys, La Jolla, CA 92093 USA
关键词
D O I
10.1021/ac060042r
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
We designed, fabricated, and tested a microfluidic device for separation of plasma from whole human blood by size exclusion in a cross-flow. The device is made of a single mold of a silicone elastomer poly( dimethylsiloxane) ( PDMS) sealed with a cover glass and is essentially disposable. When loaded with blood diluted to 20% hematocrit and driven with pulsatile pressure to prevent clogging of the channels with blood cells, the device can operate for at least 1 h, extracting similar to 8% of blood volume as plasma at an average rate of 0.65 mu L/min. The flow in the device causes very little hemolysis; the extracted plasma meets the standards for common assays and is delivered to the device outlet similar to 30 s after injection of blood to the inlet. Integration of the cross-flow microchannel array with on-chip assay elements would create a microanalysis system for point-of-care diagnostics, reducing costs, turn-around times, and volumes of blood sample and reagents required for the assays.
引用
收藏
页码:3765 / 3771
页数:7
相关论文
共 50 条
  • [1] Perfusion in microfluidic cross-flow: Separation of white blood cells from whole blood and exchange of medium in a continuous flow
    VanDelinder, Virginia
    Groisman, Alex
    ANALYTICAL CHEMISTRY, 2007, 79 (05) : 2023 - 2030
  • [3] Isolation of plasma from whole blood using a microfludic chip in a continuous cross-flow
    Chen Xing
    Cui DaFu
    Zhang LuLu
    CHINESE SCIENCE BULLETIN, 2009, 54 (02): : 324 - 327
  • [4] A Microfluidic Device for Continuous White Blood Cell Separation and Lysis From Whole Blood
    Kim, Myounggon
    Jung, Seung Mo
    Lee, Kyeong-Hwan
    Kang, Yang Jun
    Yang, Sung
    ARTIFICIAL ORGANS, 2010, 34 (11) : 996 - 1002
  • [5] A passive microfluidic device for plasma extraction from whole human blood
    Sollier, E.
    Cubizolles, M.
    Faivre, M.
    Fouillet, Y.
    Achard, J. L.
    2009 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-20, 2009, : 7030 - +
  • [6] A microfluidic device for continuous, real time blood plasma separation
    Yang, Sung
    Undar, Akif
    Zahn, Jeffrey D.
    LAB ON A CHIP, 2006, 6 (07) : 871 - 880
  • [7] CONTINUOUS SEPARATION OF WHITE BLOOD CELL FROM BLOOD IN A MICROFLUIDIC DEVICE
    Iliescu, Florina S.
    Sterian, Andreea P.
    Barbarini, Elena
    Avram, Marioara
    Iliescu, Ciprian
    UNIVERSITY POLITEHNICA OF BUCHAREST SCIENTIFIC BULLETIN-SERIES A-APPLIED MATHEMATICS AND PHYSICS, 2009, 71 (04): : 21 - 30
  • [8] Microfluidic Sedimentation System for Separation of Plasma from Whole Blood
    Kuroda, Chiaki
    Ohki, Yoshimichi
    Ashiba, Hiroki
    Fujimaki, Makoto
    Awazu, Koichi
    Tanaka, Torahiko
    Makishima, Makoto
    2014 IEEE SENSORS, 2014, : 1854 - 1857
  • [9] Simultaneous Separation, Metering, and Dilution of Plasma from Human Whole Blood in a Microfluidic System
    Tachi, Tomoya
    Kaji, Noritada
    Tokeshi, Manabu
    Baba, Yoshinobu
    ANALYTICAL CHEMISTRY, 2009, 81 (08) : 3194 - 3198
  • [10] Continuous Separation of Circulating Tumor Cells from Whole Blood Using a Slanted Weir Microfluidic Device
    Yoon, Yousang
    Lee, Jusin
    Ra, Moonsoo
    Gwon, Hyeokshin
    Lee, Seungwon
    Kim, Min Young
    Yoo, Ki-Chun
    Sul, Onejae
    Kim, Chul Geun
    Kim, Whoi-Yul
    Park, Jea-Gun
    Lee, Su-Jae
    Lee, Young Yiul
    Choi, Ho Soon
    Lee, Seung-Beck
    CANCERS, 2019, 11 (02):