A New Approach for Scatter Removal and Attenuation Compensation from SPECT/CT Images

被引:0
作者
Oloomi, Shabnam [1 ]
Eskandari, Hadi Noori [2 ]
Zakavi, Seyed Rasoul [3 ]
Knoll, Peter [4 ]
Kalantari, Faraz [5 ]
Saffar, Mohsen Hajizadeh [1 ]
机构
[1] Mashhad Univ Med Sci, Dept Med Phys, Mashhad, Iran
[2] Ferdowsi Univ Mashhad, Sch Math Sci, Dept Appl Math, Mashhad, Iran
[3] Mashhad Univ Med Sci, Fac Med, Imam Reza Hosp, Nucl Med Res Ctr, Mashhad, Iran
[4] Wilhelminenspital Vienna, Dept Nucl Med, Vienna, Austria
[5] Univ Tehran Med Sci, Res Ctr Nucl Med, Tehran, Iran
关键词
Attenuation correction; MLEM; Scatter correction; SPECT; CEREBRAL BLOOD-FLOW; MONTE-CARLO; PERFUSION SPECT; RECONSTRUCTION; MODEL; SYSTEM; PET;
D O I
暂无
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Objective(s): In SPECT, the sinogram contains scatter and lack of attenuated counts that degrade the reconstructed image quality and quantity. Many techniques for attenuation and scatter correction have been proposed. An acceptable method of correction is to incorporate effects into an iterative statistical reconstruction. Here, we propose new Maximum Likelihood Expectation Maximization (MLEM) formula to correct scattering and attenuating photons during reconstruction. Materials and Methods: In this work, scatters are estimated through Klein-Nishina formula in all iterations and CT images are used for accurate attenuation correction. Reconstructed images resulted from different MLEM reconstruction formula have been compared considering profile agreement, contrast, mean square error, signal-to-noise ratio, contrast-to-noise ratio and computational time. Results: The proposed formula has a good profile agreement, increased contrast, signal-to-noise (SNR) & contrast-to-noise ratio (CNR), computational time and decreased mean square error (MSE) compared with uncorrected images and/or images from conventional formula. Conclusion: In conclusion, by applying the proposed formula we were able to correct attenuation and scatter via MLEM and improve the image quality, which is a necessary step for both qualitative and quantitative SPECT images.
引用
收藏
页码:1181 / 1189
页数:9
相关论文
共 30 条
  • [11] Fast implementations of reconstruction-based scatter compensation in fully 3D SPECT image reconstruction
    Kadrmas, DJ
    Frey, EC
    Karimi, SS
    Tsui, BMW
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (04) : 857 - 873
  • [12] Kalantari F, 2011, HELL J NUCL MED, V14, P278
  • [13] A study to improve the image quality in low-dose computed tomography (SPECT) using filtration
    Kheruka, S. C.
    Naithani, U. C.
    Maurya, A. K.
    Painuly, N. K.
    Aggarwal, L. M.
    Gambhir, S.
    [J]. INDIAN JOURNAL OF NUCLEAR MEDICINE, 2011, 26 (01): : 14 - 21
  • [14] X-ray-based attenuation correction for positron emission tomography/computed tomography scanners
    Kinahan, PE
    Hasegawa, BH
    Beyer, T
    [J]. SEMINARS IN NUCLEAR MEDICINE, 2003, 33 (03) : 166 - 179
  • [15] King MA, 2004, EMISSION TOMOGRAPHY, P74
  • [16] Comparison of advanced iterative reconstruction methods for SPECT/CT
    Knoll, Peter
    Kotalova, Daniela
    Koechle, Gunnar
    Kuzelka, Ivan
    Minear, Greg
    Mirzaei, Siroos
    Samal, Martin
    Zadrazil, Ladislav
    Bergmann, Helmar
    [J]. ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2012, 22 (01): : 58 - 69
  • [17] Konik A.B., 2010, Evaluation of attenuation and scatter correction requirements in small animal PET and SPECT imaging
  • [18] Rapid optimization of SPECT scatter correction using model LROC observers
    Kulkarni, S.
    Khurd, R.
    Zhou, L.
    Gindi, G.
    [J]. 2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 3986 - +
  • [19] Fully 3D Monte Carlo reconstruction in SPECT: a feasibility study
    Lazaro, D
    El Bitar, Z
    Breton, V
    Hill, D
    Buvat, I
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (16) : 3739 - 3754
  • [20] Realistic CT simulation using the 4D XCAT phantom
    Segars, W. P.
    Mahesh, M.
    Beck, T. J.
    Frey, E. C.
    Tsui, B. M. W.