Optimization of Plasma Fluorogenic Thrombin-Generation Assays

被引:38
作者
Chandler, Wayne L. [1 ]
Roshal, Mikhail [1 ]
机构
[1] Univ Washington, Dept Lab Med, Seattle, WA 98195 USA
关键词
Thrombin generation; Fluorogenic; Tissue factor; Plasma; Coagulation; CALIBRATED AUTOMATED THROMBOGRAPHY; RISK;
D O I
10.1309/AJCP6AY4HTRAAJFQ
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
We optimized fluorogenic thrombin-generation assays with regard to sample volume, calibration, analytic corrections, and activation reagents. Lower sample volumes (40 vs 80 mu L) were associated with better recovery of thrombin activity lower interference due to absorbance of light, and higher total thrombin generation (area under the curve), even using internal standards to calibrate plasma samples. With lower sample volumes, there was no advantage to internal calibration of samples without obvious interference (hemolysis). Previously developed corrections for measured vs expected fluorescence units, residual thrombin-alpha(2)-macroglobulin activity, and hemolysis improved the analytic accuracy of the assay. An optimized assay with a 40-mu L sample volume, analytic corrections, and a corn trypsin inhibitor to block contact activation showed that 0.6 pmol/L tissue factor activator was better than 5 pmol/L at differentiating healthy subjects from patients with sepsis while demonstrating good reproducibility (area under the curve, 4% within-run and 7% between-run coefficient of variation).
引用
收藏
页码:169 / 179
页数:11
相关论文
共 15 条
  • [1] The measurement and application of thrombin generation
    Baglin, T
    [J]. BRITISH JOURNAL OF HAEMATOLOGY, 2005, 130 (05) : 653 - 661
  • [2] Chantarangkul V, 2003, HAEMATOLOGICA, V88, P547
  • [3] Global tests of haemostasis in critically ill patients with severe sepsis syndrome compared to controls
    Collins, Peter W.
    Macchiavello, Luis I.
    Lewis, Sarah J.
    Macartney, Nichola J.
    Saayman, Anton G.
    Luddington, Roger
    Baglin, Trevor
    Findlay, George P.
    [J]. BRITISH JOURNAL OF HAEMATOLOGY, 2006, 135 (02) : 220 - 227
  • [4] Effect of standardization and normalization on imprecision of calibrated automated thrombography: an international multicentre study
    Dargaud, Yesim
    Luddington, Rodger
    Gray, Elaine
    Negrier, Claude
    Lecompte, Thomas
    Petros, Sirak
    Hogwood, John
    Bordet, Jean-Claude
    Hogwood, John
    Bordet, Jean-Claude
    Regnault, Veronique
    Siegemund, Annelie
    Baglin, Trevor
    [J]. BRITISH JOURNAL OF HAEMATOLOGY, 2007, 139 (02) : 303 - 309
  • [5] Thrombin generation in plasma of healthy adults and children: Chromogenic versus fluorogenic thrombogram analysis
    Devreese, Katrien
    Wijns, Walter
    Combes, Isabelle
    Van kerckhoven, Soetkin
    Hoylaerts, Marc F.
    [J]. THROMBOSIS AND HAEMOSTASIS, 2007, 98 (03) : 600 - 613
  • [6] Thrombin generation, a function test of the haemostatic-thrombotic system
    Hemker, H. Coenraad
    Al Dieri, Raed
    De Smedt, Erik
    Beguin, Suzette
    [J]. THROMBOSIS AND HAEMOSTASIS, 2006, 96 (05) : 553 - 561
  • [7] The Calibrated Automated Thrombogram (CAT):: A universal routine test for hyper- and hypocoagulability
    Hemker, HC
    Giesen, P
    AlDieri, R
    Regnault, V
    de Smed, E
    Wagenvoord, R
    Lecompte, T
    Béguin, S
    [J]. PATHOPHYSIOLOGY OF HAEMOSTASIS AND THROMBOSIS, 2002, 32 (5-6) : 249 - 253
  • [8] Calibrated automated thrombin generation measurement in clotting plasma
    Hemker, HC
    Giesen, P
    Al Dieri, R
    Regnault, V
    de Smedt, E
    Wagenvoord, R
    Lecompte, T
    Béguin, S
    [J]. PATHOPHYSIOLOGY OF HAEMOSTASIS AND THROMBOSIS, 2003, 33 (01) : 4 - 15
  • [9] HEMKER HC, 1995, THROMB HAEMOSTASIS, V74, P134
  • [10] Identification of patients at low risk for recurrent venous thromboembolism by measuring thrombin generation
    Hron, Gregor
    Kollars, Marietta
    Binder, Bernd R.
    Eichinger, Sabine
    Kyrle, Paul A.
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2006, 296 (04): : 397 - 402