MEASUREMENT OF RADIATION DOSES TO THE EYE LENS DURING ORTHOPEDIC SURGERY USING AN C-ARM X-RAY SYSTEM

被引:14
|
作者
Suzuki, Akira [1 ]
Matsubara, Kosuke [2 ]
Sasa, Yuko [1 ]
机构
[1] Japanese Red Cross Sendai Hosp, Dept Radiol, 2-43-3 Yagiyama Honcho, Sendai, Miyagi 9828501, Japan
[2] Kanazawa Univ, Dept Quantum Med Technol, Fac Hlth Sci, 5-11-80 Kodatsuno, Kanazawa, Ishikawa 9200942, Japan
关键词
EXPOSURE;
D O I
10.1093/rpd/ncx250
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The present study aimed to determine doses delivered to the eye lenses of surgeons while using the inverted-C-arm technique and the protective effect of leaded spectacles during orthopedic surgery. The kerma in air was measured at five positions on leaded glasses positioned near the eye lens and on the neck using small optically stimulated luminescence (OSL) dosemeters. The lens equivalent dose was also measured at the neck using an OSL dosemeter. The maximum equivalent dose to the eye lens and the maximum kerma were 0.8 mSv/month and 0.66 mGy/month, respectively. The leaded glasses reduced the exposure by similar to 60%. Even if the surgeons are exposed to the maximum dose of X-ray radiation for 5 years, the equivalent doses to the eye lens will not exceed the present limit recommended by the ICRP.
引用
收藏
页码:189 / 195
页数:7
相关论文
共 20 条
  • [1] Analysis of radiation risk to patients from intraoperative use of the mobile X-ray system (C-arm)
    Lee, Yang-Sub
    Lee, Hae-Kag
    Cho, Jae-Hwan
    Kim, Ham-Gyum
    JOURNAL OF RESEARCH IN MEDICAL SCIENCES, 2015, 20 (01): : 7 - 12
  • [2] Radiation Doses of Cerebral Blood Volume Measurements Using C-Arm CT: A Phantom Study
    Chu, W. F.
    Lin, C. J.
    Chen, W. S.
    Hung, S. C.
    Chiu, C. F.
    Wu, T. H.
    Guo, W. Y.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2014, 35 (06) : 1073 - 1077
  • [3] The Dose Rate Measurements of C-arm Fluoroscopic X-ray Systems with and without Apron
    Tuncel, Nina
    1ST INTERNATIONAL CONFERENCE ON RADIATIONS AND APPLICATIONS (ICRA-2017), 2018, 1994
  • [4] Effect of X-ray protective visors on the eye lens dose during ureteroscopy
    Zoeller, G.
    Konstanz, V
    Broenner, J.
    Hoedlmoser, H.
    UROLOGE, 2020, 59 (10): : 1231 - 1236
  • [5] X-ray Radiation Effect of Image Intensfier (C-Arm) on Adipose Tissue-Mesenchymal Stem Cell Viability and Osteogenic Differentiation
    Rahyussalim, Ahmad J.
    Pawitan, Jeanne A.
    Maska, Hendra
    Kurniawati, Tri
    ADVANCED SCIENCE LETTERS, 2017, 23 (07) : 6970 - 6974
  • [6] Effect of metallic tools on scattered radiation dose during the use of C-arm fluoroscopy in orthopaedic surgery
    Sung, Ki Hyuk
    Jung, Young-Jun
    Cha, Hyemi
    Chung, Chin Youb
    Lee, Kisung
    Park, Moon Seok
    JOURNAL OF RADIATION RESEARCH, 2019, 60 (01) : 1 - 6
  • [7] Optimization of radiation doses for open lumbar spinal fusion using C-arm fluoroscopy and impact on radiation-induced cancer: a pilot study
    Ty, Claire Van Ngoc
    Fitton, Isabelle
    Arvieu, Robin
    Ferrero, Emmanuelle
    de Loubresse, Christian Garreau
    Khalife, Marc
    EUROPEAN SPINE JOURNAL, 2024, 33 (05) : 2043 - 2048
  • [8] Distribution of scatter radiation by C-arm cone-beam computed tomography in angiographic suite: measurement of doses and effectiveness of protection devices
    Yamaji, Mayako
    Ishiguchi, Tsuneo
    Koyama, Shuji
    Ikeda, Shuji
    Kitagawa, Akira
    Hagihara, Makiyo
    Itoh, Yuji
    Nakamura, Masaru
    Ota, Toyohiro
    Suzuki, Kojiro
    NAGOYA JOURNAL OF MEDICAL SCIENCE, 2021, 83 (02): : 277 - 286
  • [9] LEAD GLASSES SIGNIFICANTLY REDUCED RADIATION DOSES TO EYE LENSES OF UROLOGISTS DURING PROCEDURES INVOLVING OVER-COUCH X-RAY TUBES
    Suzuki, Akira
    Matsubara, Kosuke
    Chusin, Thunyarat
    Suzuki, Yuko
    RADIATION PROTECTION DOSIMETRY, 2020, 189 (01) : 28 - 34
  • [10] Radiation dose during interventional cardiology procedures: portable C-arm vs. a new generation fluoroscopy system
    Pierce, K. V.
    Scansen, B. A.
    Rao, S.