Experimental verification of a CT-based Monte Carlo dose-calculation method in heterogeneous phantoms

被引:23
作者
Wang, L [1 ]
Lovelock, M [1 ]
Chui, CS [1 ]
机构
[1] Mem Sloan Kettering Canc Ctr, Dept Med Phys, New York, NY 10021 USA
关键词
Monte Carlo; inhomogeneous phantom; dose calculation;
D O I
10.1118/1.598802
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To further validate the Monte Carlo dose-calculation method [Med. Phys. 25, 867-878 (1998)] developed at the Memorial Sloan-Kettering Cancer Center, we have performed experimental verification in various inhomogeneous phantoms. The phantom geometries included simple layered slabs, a simulated bone column, a simulated missing-tissue hemisphere, and an anthropomorphic head geometry (Alderson Rando Phantom). The densities of the inhomogeneity range from 0.14 to 1.86 g/cm(3) simulating both clinically relevant lunglike and bonelike materials. The data an reported as central axis depth doses, dose profiles, dose values at points of interest, such as points at the interface of two different media and in the "nasopharynx" region of the Rando head. The dosimeters used in the measurement included dosimetry film, TLD chips, and rods. The measured data were compared to that of Monte Carlo calculations for the same geometrical configurations. In the case of the Rando head phantom, a CT scan of the phantom was used to define the calculation geometry and to locate the points of interest. The agreement between the calculation and measurement is generally within 2.5%. This work validates the accuracy of the Monte Carlo method. While Monte Carlo, at present, is still too slow for routine treatment planning, it can be used as a benchmark against which other dose calculation methods can be compared. (C) 1999 American Association of Physicists in Medicine. [S0094-2405(99)01912-4].
引用
收藏
页码:2626 / 2634
页数:9
相关论文
共 19 条
  • [1] BATHO H F, 1964, J Can Assoc Radiol, V15, P79
  • [2] BRIESMEISTER JF, 1993, LA12625
  • [3] COX LJ, 1997, UCRLJC126731
  • [4] CT-based Monte Carlo simulation tool for dosimetry planning and analysis
    DeMarco, JJ
    Solberg, TD
    Smathers, JB
    [J]. MEDICAL PHYSICS, 1998, 25 (01) : 1 - 11
  • [5] HALBLEIB JA, 1992, 911634 SAND
  • [6] Hoban P. W., 1992, Australasian Physical and Engineering Sciences in Medicine, V15, P113
  • [7] ICRU, 1984, 37 ICRU
  • [8] KHAN FM, 1992, PHYSICS RAD THERAPY, P287
  • [9] A dual source photon beam model used in convolution/superposition dose calculations for clinical megavoltage x-ray beams
    Liu, HH
    Mackie, TR
    McCullough, EC
    [J]. MEDICAL PHYSICS, 1997, 24 (12) : 1960 - 1974
  • [10] MONTE-CARLO MODEL OF PHOTON BEAMS USED IN RADIATION-THERAPY
    LOVELOCK, DMJ
    CHUI, CS
    MOHAN, R
    [J]. MEDICAL PHYSICS, 1995, 22 (09) : 1387 - 1394