Comparison of metal release from various metallic biomaterials in vitro

被引:683
作者
Okazaki, Y
Gotoh, E
机构
[1] Natl Inst Adv Ind Sci & Technol, Inst Mech Syst Engn, Ecol Oriented Struct Mat Grp, Tsukuba, Ibaraki 3058564, Japan
[2] Natl Inst Technol & Evaluat, Tsukuba, Ibaraki 3050044, Japan
关键词
in vitro test; corrosion; metal release; stainless steel; cobalt-chromium-molybdenum alloy; titanium alloy; pH;
D O I
10.1016/j.biomaterials.2004.02.005
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
To investigate the metal release of each base and alloying elements in vitro, SUS316L stainless steel, Co-Cr-Mo casting alloy, commercially pure Ti grade 2, and Ti-6Al-4V, V-free Ti-6Al-7Nb and Ti-15Zr-4Nb-4Ta alloys were immersed in various solutions, namely, alpha-medium, PBS(-), calf serum, 0.9% NaCl, artificial saliva, 1.2 mass% L-Cysteine, 1 mass% lactic acid and 0.01 mass% HCl for 7 d. The difference in the quantity of Co released from the Co-Cr-Mo casting alloy was relatively small in all the solutions. The quantities of Ti released into a-medium, PBS(-), calf serum, 0.9% NaCl and artificial saliva were much lower than those released into 1.2% L-CySteine, 1% lactic acid and 0.01% HCl. The quantity of Fe released from SUS316L stainless steel decreased linearly with increasing pH. On the other hand, the quantity of Ti released from Ti materials increased with decreasing pH, and it markedly attenuated at pHs of approximately 4 and higher. The quantity of Ni released from stainless steel gradually decreased with increasing pH. The quantities of Al released from the Ti-6Al-4V and Ti-W-7M alloys gradually decreased with increasing pH. A small V release was observed in calf serum, PBS(-), artificial saliva, 1% lactic acid, 1.2% L-cysteine and 0.01% HCl. The quantity of Ti released from the Ti-15Zr-4Nb-4Ta alloy was smaller than those released from the Ti-6Al-4V and Ti6-Al-7Nb alloys in all the solutions. In particular, it was approximately 30% or smaller in 1% lactic acid, 1.2% L-cysteine and 0.01% HCl. The quantity of (Zr + Nb + Ta) released was also considerably lower than that of (Al + Nb) or (Al + V) released. Therefore, the Ti-15Zr-4Nb-4Ta alloy with its low metal release in vitro is considered advantageous for long-term implants. (C) 2004 Elsevier Ltd. All rights reserved.
引用
收藏
页码:11 / 21
页数:11
相关论文
共 43 条
  • [21] Kovacs P., 1996, ASTM STP, V1272, P163
  • [22] SYSTEMIC EFFECTS OF IMPLANTED PROSTHESES MADE OF COBALT-CHROMIUM ALLOYS
    MICHEL, R
    NOLTE, M
    REICH, M
    LOER, F
    [J]. ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1991, 110 (02) : 61 - 74
  • [23] Mishra AK, 1996, ASTM STP, P96
  • [24] Corrosion resistance of implant alloys in pseudo physiological solution and role of alloying elements in passive films
    Okazaki, Y
    Tateishi, T
    Ito, Y
    [J]. MATERIALS TRANSACTIONS JIM, 1997, 38 (01): : 78 - 84
  • [25] Effect of metal released from Ti alloy wear powder on cell viability
    Okazaki, Y
    Nishimura, E
    [J]. MATERIALS TRANSACTIONS JIM, 2000, 41 (09): : 1247 - 1255
  • [26] Effects of Mo and Pd on corrosion resistance of V-free titanium alloys for medical implants
    Okazaki, Y
    Kyo, K
    Ito, Y
    Tateishi, T
    [J]. MATERIALS TRANSACTIONS JIM, 1997, 38 (04): : 344 - 352
  • [27] Effects of Ti, Al and V concentrations on cell viability
    Okazaki, Y
    Rao, S
    Asao, S
    Tateishi, T
    Katsuda, S
    Furuki, Y
    [J]. MATERIALS TRANSACTIONS JIM, 1998, 39 (10): : 1053 - 1062
  • [28] Effects of metallic concentrations other than Ti, Al and V on cell viability
    Okazaki, Y
    Rao, S
    Asao, S
    Tateishi, T
    [J]. MATERIALS TRANSACTIONS JIM, 1998, 39 (10): : 1070 - 1079
  • [29] Effect of friction on anodic polarization properties of metallic biomaterials
    Okazaki, Y
    [J]. BIOMATERIALS, 2002, 23 (09) : 2071 - 2077
  • [30] A new Ti-15Zr-4Nb-4Ta alloy for medical applications
    Okazaki, Y
    [J]. CURRENT OPINION IN SOLID STATE & MATERIALS SCIENCE, 2001, 5 (01) : 45 - 53