Highly improved sensitivity of TS-FF-AAS for Cd(II) determination at ng L-1 levels using a simple flow injection minicolumn preconcentration system with multiwall carbon nanotubes

被引:91
作者
Teixeira Tarley, Cesar Ricardo [1 ]
Barbosa, Adriano Francisco
Segatelli, Mariana Gava
Figueiredo, Eduardo Costa
Luccas, Pedro Orival
机构
[1] Univ Fed Alfenas, Dept Exact Sci, BR-37130000 Alfenas, MG, Brazil
[2] Univ Estadual Campinas, Inst Chem, Dept Inorgan Chem, BR-13083970 Campinas, SP, Brazil
[3] Univ Estadual Campinas, Inst Chem, Dept Analyt Chem, BR-13083970 Campinas, SP, Brazil
关键词
D O I
10.1039/b606997e
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new method for cadmium determination at ng L-1 levels is described. The method is based on the on-line coupling of a flow preconcentration system using multiwall carbon nanotubes (MWCNT) as sorbent with TS-FF-AAS determination. Cadmium preconcentration was at pH 4.9 onto an MWCNT minicolumn (30 mg) for 2 min at a 5.0 mL min(-1) flow rate. The elution step is performed by using 0.5 mol L-1 HNO3 and the cadmium desorbed is directly pumped to a TS-FF-AAS. All experimental parameters that play important roles in system performance were evaluated and optimized by means of fractional factorial designs and response surface methodology. The excellent characteristics of MWCNT as sorbent, mainly owing to its high surface area, make it possible to obtain a preconcentration factor of 51-fold, thus improving the detection and quanti. cation limits in TS-FF-AAS, 11.4 and 38.1 ng L-1, respectively. When the flow preconcentration system, FI-TS-FF-AAS, was compared with FAAS alone, an increase in the sensitivity of 640-fold was obtained. The calibration graph was linear with a correlation coefficent higher than 0.999 from 38.1 to 1250 ng L-1. Repeatability of the measurements (n = 10), assessed as relative standard deviation (RSD), was found to be 6.5 and 2.1% for cadmium concentrations of 100 and 1000 ng L-1, respectively. Important parameters to characterize the flow preconcentration system were also evaluated, the consumption index being 0.196 mL, the concentration efficiency 25.5 min(-1) and the sample throughput 20 samples per hour. In order to demonstrate the accuracy of the system, addition and recovery studies in water samples (mineral water, tap water and river water) and cigarette samples were carried out. Moreover, for the same purpose, cadmium was determined in certified biological materials (Bovine Liver and Rye Grass), giving an average result in close agreement with the certified value.
引用
收藏
页码:1305 / 1313
页数:9
相关论文
共 36 条
[1]   Optimization of the preconcentration system of cadmium with 1(2-thiazolylazo)-p-cresol using a knotted reactor and flame atomic absorption spectrometric detection [J].
Cerutti, S ;
Ferreira, SLC ;
Gásquez, JA ;
Olsina, RA ;
Martinez, LD .
JOURNAL OF HAZARDOUS MATERIALS, 2004, 112 (03) :279-283
[2]   Use of Nb2O5-SiO2 in an automated on-line preconcentration system for determination of copper and cadmium by FAAS [J].
da Silva, EL ;
Ganzarolli, EM ;
Carasek, E .
TALANTA, 2004, 62 (04) :727-733
[3]   Immobilized salen (N,N′-bis (salicylidene) ethylenediamine) as a complexing agent for on-line sorbent extraction/preconcentration and flow injection-flame atomic absorption spectrometry [J].
Dadfarnia, S ;
Shabani, AMH ;
Tamaddon, F ;
Rezaei, M .
ANALYTICA CHIMICA ACTA, 2005, 539 (1-2) :69-75
[4]   Acid extraction and cloud point preconcentration as sample preparation strategies for cobalt determination in biological materials by thermospray flame furnace atomic absorption spectrometry [J].
Donati, GL ;
Nascentes, CC ;
Nogueira, ARA ;
Arruda, MAZ ;
Nóbrega, JS .
MICROCHEMICAL JOURNAL, 2006, 82 (02) :189-195
[5]   Determination of Cd in biological samples by flame AAS following on-line preconcentration by complexation with 0,0-diethyldithiophosphate and solid phase extraction with Amberlite XAD-4 [J].
dos Santos, EJ ;
Herrmann, AB ;
Ribeiro, AS ;
Curtius, AJ .
TALANTA, 2005, 65 (02) :593-597
[6]   Development of an on-line preconcentration system for zinc determination in biological samples [J].
Dutra, RL ;
Maltez, HF ;
Carasek, E .
TALANTA, 2006, 69 (02) :488-493
[7]  
Fang Z., 1993, FLOW INJECTION SEPAR
[8]   FUNDAMENTAL AND PRACTICAL CONSIDERATIONS IN THE DESIGN OF ONLINE COLUMN PRECONCENTRATION FOR FLOW-INJECTION ATOMIC SPECTROMETRIC SYSTEMS [J].
FANG, ZL ;
XU, SK ;
ZHANG, SC .
ANALYTICA CHIMICA ACTA, 1987, 200 (01) :35-49
[9]   The determination of molybdenum in water and biological samples by graphite furnace atomic spectrometry after polyurethane foam column separation and preconcentration [J].
Ferreira, SLC ;
dos Santos, HC ;
Campos, RC .
TALANTA, 2003, 61 (06) :789-795
[10]   Thermospray flame furnace atomic absorption spectrometry (TS-FF-AAS) -: a simple method for trace element determination with microsamples in the μg/1 concentration range [J].
Gáspár, A ;
Berndt, H .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2000, 55 (06) :587-597