Design and experimental evaluation of a teleoperated haptic robot-assisted catheter operating system

被引:79
作者
Yin, Xuanchun [1 ]
Guo, Shuxiang [2 ,3 ]
Hirata, Hideyuki [2 ]
Ishihara, Hidenori [2 ]
机构
[1] Kagawa Univ, Grad Sch Engn, Takamatsu, Kagawa 7610396, Japan
[2] Kagawa Univ, Dept Intelligent Mech Syst Engn, Takamatsu, Kagawa 7610396, Japan
[3] Beijing Inst Technol, Sch Life Sci, Beijing 100081, Peoples R China
关键词
Magnetorheological fluids; haptic interface; teleoperated haptic system; minimally invasive surgery; transparency; stability; catheter operating system; MR FLUID; DEVICES; DAMPER;
D O I
10.1177/1045389X14556167
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Minimally invasive surgery and therapy is popularly used both for diagnosis and for surgery. Teleoperation, a promising surgery, is used to protect the surgeon from X-ray radiation as well as to address the problem of lacking experienced surgeons in remote rural areas. However, surgery success ratio should be considered because the surgeon was separated from the patient remotely. A most effective addressing method to improve success ratio is design of a haptic interface as a master console, which can provide the "immersive'' operation to the surgeon. In this study, a haptic catheter operation system for teleoperation through exploiting magnetorheological fluids is proposed to solve the safety problem. The haptic sensation is provided by varying the viscosity of the magnetorheological fluids by adjusting the magnetic field, which is dependent on the force measured in the slave manipulator. Therefore, three parts of the haptic interface were designed and fabricated: magnetic field, magnetorheological fluids container and haptic performance calibration mechanism. Some preliminary experiments have been done to verify the effectiveness of this kind of haptic interface. Experimental results illustrated that the designed haptic catheter operation system can be used for teleoperation and for training the surgeon for the non-experience.
引用
收藏
页码:3 / 16
页数:14
相关论文
共 32 条
[1]  
Adams R. J., 1998, Proceedings of the ASME Dynamic Systems and Control Division-1998, P227
[2]  
Adams RJ, 1998, 1998 IEEE/RSJ INTERNATIONAL CONFERENCE ON INTELLIGENT ROBOTS AND SYSTEMS - PROCEEDINGS, VOLS 1-3, P1254, DOI 10.1109/IROS.1998.727471
[3]   Modeling of a magnetorheological actuator including magnetic hysteresis [J].
An, J ;
Kwon, DS .
JOURNAL OF INTELLIGENT MATERIAL SYSTEMS AND STRUCTURES, 2003, 14 (09) :541-550
[4]   Magnetorheological fluids: Materials, characterization, and devices [J].
Ashour, O ;
Rogers, CA ;
Kordonsky, W .
JOURNAL OF INTELLIGENT MATERIAL SYSTEMS AND STRUCTURES, 1996, 7 (02) :123-130
[5]   Analysis and design of an electromagnetic system for the characterization of magneto-rheological fluids for haptic interfaces [J].
Bicchi, A ;
Raugi, M ;
Rizzo, R ;
Sgambelluri, N .
IEEE TRANSACTIONS ON MAGNETICS, 2005, 41 (05) :1876-1879
[6]  
Butz T, 2002, Z ANGEW MATH MECH, V82, P3, DOI 10.1002/1521-4001(200201)82:1<3::AID-ZAMM3>3.0.CO
[7]  
2-O
[8]   MR fluid, foam and elastomer devices [J].
Carlson, JD ;
Jolly, MR .
MECHATRONICS, 2000, 10 (4-5) :555-569
[9]   Commercial magneto-rheological fluid devices [J].
Carlson, JD ;
Catanzarite, DM ;
StClair, KA .
INTERNATIONAL JOURNAL OF MODERN PHYSICS B, 1996, 10 (23-24) :2857-2865
[10]  
Carnahan H, 2006, P 1 CAN STUD C BIOM, P44