A Preliminary Assessment of Legged Mobility Provided by a Lower Limb Exoskeleton for Persons With Paraplegia

被引:152
作者
Farris, Ryan J. [1 ]
Quintero, Hugo A. [1 ]
Murray, Spencer A. [2 ]
Ha, Kevin H. [3 ]
Hartigan, Clare [4 ]
Goldfarb, Michael [2 ,3 ]
机构
[1] Vanderbilt Univ, Nashville, TN 37240 USA
[2] Vanderbilt Univ, Dept Elect Engn, Nashville, TN 37240 USA
[3] Vanderbilt Univ, Dept Mech Engn, Nashville, TN 37240 USA
[4] Shepherd Ctr, Atlanta, GA 30309 USA
基金
美国国家卫生研究院;
关键词
Assessment; assistive technology; knee-ankle-foot orthoses (KAFO); leg braces; legged mobility; lower limb exoskeleton; paraplegia; powered orthosis; spinal-cord injury (SCI); SPINAL-CORD-INJURY; ENERGY-EXPENDITURE; HYBRID ORTHOSIS; RECIPROCAL GAIT; OLDER-ADULTS; GO TEST; RELIABILITY; WALKING; VALIDITY; SUIT;
D O I
10.1109/TNSRE.2013.2268320
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper presents an assessment of a lower limb exoskeleton for providing legged mobility to people with paraplegia. In particular, the paper presents a single-subject case study comparing legged locomotion using the exoskeleton to locomotion using knee-ankle-foot orthoses (KAFOs) on a subject with a T10 motor and sensory complete injury. The assessment utilizes three assessment instruments to characterize legged mobility, which are the timed up-and-go test, the Ten-Meter Walk Test (10 MWT), and the Six-Minute Walk Test (6 MWT), which collectively assess the subject's ability to stand, walk, turn, and sit. The exertion associated with each assessment instrument was assessed using the Physiological Cost Index. Results indicate that the subject was able to perform the respective assessment instruments 25%, 70%, and 80% faster with the exoskeleton relative to the KAFOs for the timed up-and-go test, the 10 MWT, and the 6 MWT, respectively. Measurements of exertion indicate that the exoskeleton requires 1.6, 5.2, and 3.2 times less exertion than the KAFOs for each respective assessment instrument. The results indicate that the enhancement in speed and reduction in exertion are more significant during walking than during gait transitions.
引用
收藏
页码:482 / 490
页数:9
相关论文
共 46 条
[1]  
[Anonymous], P 3 INT S AMB MON
[2]   Gait Evaluation of a Novel Hip Constraint Orthosis With Implication for Walking in Paraplegia [J].
Audu, Musa L. ;
To, Curtis S. ;
Kobetic, Rudi ;
Triolo, Ronald J. .
IEEE TRANSACTIONS ON NEURAL SYSTEMS AND REHABILITATION ENGINEERING, 2010, 18 (06) :610-618
[3]  
Bailey M., 1995, Physiotherapy, V81, P618
[4]  
BRAUN Z, 1985, SCAND J REHABIL MED, P93
[5]   Clinical experience with functional electrical stimulation-assisted gait with parastep in spinal cord-injured patients [J].
Brissot, R ;
Gallien, P ;
Le Bot, MP ;
Beaubras, A ;
Laisné, D ;
Beillot, J ;
Dassonville, J .
SPINE, 2000, 25 (04) :501-508
[6]  
Brown-Triolo DL, 2002, J REHABIL RES DEV, V39, P659
[7]   Design and simulation of a pneumatic, stored-energy, hybrid orthosis for gait restoration [J].
Durfee, WK ;
Rivard, A .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2005, 127 (06) :1014-1019
[8]   Functional walk tests in individuals with stroke - Relation to perceived exertion and myocardial exertion [J].
Eng, JJ ;
Chu, KS ;
Dawson, AS ;
Kim, CM ;
Hepburn, KE .
STROKE, 2002, 33 (03) :756-761
[9]   The ReWalk Powered Exoskeleton to Restore Ambulatory Function to Individuals with Thoracic-Level Motor-Complete Spinal Cord Injury [J].
Esquenazi, Alberto ;
Talaty, Mukul ;
Packel, Andrew ;
Saulino, Michael .
AMERICAN JOURNAL OF PHYSICAL MEDICINE & REHABILITATION, 2012, 91 (11) :911-921
[10]   Preliminary evaluation of a controlled-brake orthosis for FES-aided gait [J].
Goldfarb, M ;
Korkowski, K ;
Harrold, B ;
Durfee, W .
IEEE TRANSACTIONS ON NEURAL SYSTEMS AND REHABILITATION ENGINEERING, 2003, 11 (03) :241-248