Selective mercury determination after membrane complexation and total reflection X-ray fluorescence analysis

被引:58
作者
Koulouridakis, PE [1 ]
Kallithrakas-Kontos, NG [1 ]
机构
[1] Tech Univ Crete, Analyt & Environm Chem Lab, GR-73100 Khania, Greece
关键词
D O I
10.1021/ac049780a
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new method for low mercury concentration analysis in drinking waters is presented. Membranes containing a few micrograms of various complexing reagents were produced on the surface of quartz glasses (reflectors). The reflectors were immersed in water solutions containing various concentrations of inorganic mercury salts at low concentrations (1-40 ng/mL). The membranes were left to equilibrate in 5-500 mL of solution for many hours; when the equilibration stage was finished they were cleaned with ultrapure water and left to dry. Analyses were performed by total reflection X-ray fluorescence (TXRF). The effects of various experimental parameters (complexing agent, equilibrium time, sample volume, etc.) as well as the selectivity of the membranes were studied. The complexing reagent dithizone with a PVC-based membrane gave the best results. The limit of quantitation was 0.8 ng/mL.
引用
收藏
页码:4315 / 4319
页数:5
相关论文
共 14 条
[1]   Determination of trace levels of mercury in water samples based on room temperature phosphorescence energy transfer [J].
de la Riva, BSV ;
Costa-Fernández, JM ;
Jin, WJ ;
Pereiro, R ;
Sanz-Medel, A .
ANALYTICA CHIMICA ACTA, 2002, 455 (02) :179-186
[2]   Sensitive determination of mercury in tap water by cloud point extraction pre-concentration and flow injection-cold vapor-inductively coupled plasma optical emission spectrometry [J].
de Wuilloud, JCA ;
Wuilloud, RG ;
Silva, MF ;
Olsina, RA ;
Martinez, LD .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2002, 57 (02) :365-374
[3]  
Jenkins R., 1988, CHEM ANAL SERIES MON, V99
[4]  
Klockenkamper R., 1996, TOTAL REFLECTION XRA
[5]  
KOULOURIDAKIS PE, IN PRESS MICROCHIM A
[6]   Nafion-based optode for the detection of metal ions in flow analysis [J].
Madden, JE ;
Cardwell, TJ ;
Cattrall, R ;
Deady, LW .
ANALYTICA CHIMICA ACTA, 1996, 319 (1-2) :129-134
[7]   A method for screening total mercury in water using a flow injection system with piezoelectric detection [J].
Manganiello, L ;
Ríos, A ;
Valcárcel, M .
ANALYTICAL CHEMISTRY, 2002, 74 (04) :921-925
[8]  
Martell A.E., 2001, NIST critically selected stability constants of metal complexes, version 6 Gaithersburg
[9]  
MERIAN E, 1991, METALS THEIR COMPOUN, P701
[10]   Lead ion selective PVC membrane electrode based on 5,5′-dithiobis-(2-nitrobenzoic acid) [J].
Rouhollahi, A ;
Ganjali, MR ;
Shamsipur, M .
TALANTA, 1998, 46 (06) :1341-1346