Impact of mask thickness on the imaging performance of the coded-aperture gamma camera

被引:2
作者
Zhu, B. [1 ]
Wang, Z. [1 ]
Zheng, Z. [1 ]
Zhou, R. [1 ]
Yang, C. [1 ]
机构
[1] Sichuan Univ, Coll Phys, Key Lab Radiat Phys & Technol, Minist Educ, Chengdu 610064, Peoples R China
基金
中国国家自然科学基金;
关键词
Detector modelling and simulations I (interaction of radiation with matter; interaction of photons with matter; interaction of hadrons with matter; etc); Gamma detectors (scintillators; CZT; HPGe; HgI etc); Gamma camera; SPECT; PET PET/CT coronary CT angiography (CTA); Interaction of radiation with matter;
D O I
10.1088/1748-0221/17/05/T05019
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Coded-aperture gamma imaging techniques have been applied widely in several radiation monitoring fields. However, in terms of the design of the coded-aperture mask, the category and rank of the mask are generally considered, but there is a lack of detailed research on selection of the appropriate mask thickness. Concretely, the imaging performance is affected by the mask thickness in terms of two factors: the full-field image quality of the overall field of view (FOV), which represents the background noise tolerance of the radioactive source; and the image brightness consistency, which represents consistency of the imaging efficiency. We have studied and analyzed these two imaging performance indexes in relation to the mask thickness based on the model of our custom-made gamma camera. Specifically, the full-field image quality initially improves and later deteriorates with the mask thickness increases from 0.1 mm to 30 mm because of the competition and the tradeoff between the transmission background and the oblique irradiation effect; the image brightness consistency initially descends, then improves, and finally decreases again under the same conditions as a result of the tradeoff between the improvement of the transmission background ratio and deterioration of the oblique irradiation coefficient ratio. Furthermore, these two indexes change trends are both related to gamma-ray energy. Finally, based on a comprehensive analysis of the imaging performance, the optimal thicknesses for the customized W-Cu alloy modified uniformly redundant array masks for 59.5 keV, 662 keV, 1250 keV gamma-ray imaging are 0.5 mm, 10 mm, and 16 mm, respectively.
引用
收藏
页数:16
相关论文
共 16 条
[1]   Optimal coded aperture patterns for improved SNR in nuclear medicine imaging [J].
Accorsi, R ;
Gasparini, F ;
Lanza, RC .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2001, 474 (03) :273-284
[2]  
Accorsi R., 2001, Thesis (Ph. D.
[3]   A comprehensive experimental characterization of the iPIX gamma imager [J].
Amgarou, K. ;
Paradiso, V. ;
Patoz, A. ;
Bonnet, F. ;
Handley, J. ;
Couturier, P. ;
Becker, F. ;
Menaa, N. .
JOURNAL OF INSTRUMENTATION, 2016, 11
[4]   Gamma-ray modulation properties of tungsten coded apertures for a novel mixed-field imaging system [J].
Cieslak, M. J. ;
Gamage, K. A. A. ;
Glover, R. ;
Taylor, C. J. .
JOURNAL OF INSTRUMENTATION, 2019, 14 (02)
[5]   CODED APERTURE IMAGING WITH UNIFORMLY REDUNDANT ARRAYS [J].
FENIMORE, EE ;
CANNON, TM .
APPLIED OPTICS, 1978, 17 (03) :337-347
[6]   GAMPIX: A new generation of gamma camera [J].
Gmar, M. ;
Agelou, M. ;
Carrel, F. ;
Schoepff, V. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 652 (01) :638-640
[7]   NEW FAMILY OF BINARY ARRAYS FOR CODED APERTURE IMAGING [J].
GOTTESMAN, SR ;
FENIMORE, EE .
APPLIED OPTICS, 1989, 28 (20) :4344-4352
[8]  
Hong J., 2006, THESIS HUAZHONG U SC
[9]   Comparison of gamma ray localization using system matrixes obtained by either MCNP simulations or ray-driven calculations for a coded-aperture imaging system [J].
Jeong, Manhee ;
Hammig, Mark D. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2020, 954
[10]   High performance γ-ray imager using dual anti-mask method for the investigation of high-energy nuclear materials [J].
Lee, Taewoong ;
Lee, Wonho .