Applied Taguchi method for fatigue testing of customized hip implant

被引:2
作者
Dharme, Mangesh R. [1 ]
Kuthe, Abhaykumar M. [1 ]
Deshmukh, Tushar R. [2 ]
机构
[1] Visvesvaraya Natl Inst Technol, Dept Mech Engn, Nagpur, Maharashtra, India
[2] Prof Ram Meghe Inst Technol & Res, Dept Mech Engn, Amravati, India
关键词
Experimental model; Fatigue; Finite element analysis; Gait cycle; Orthogonal array; Taguchi methods; FINITE-ELEMENT-ANALYSIS; 3-DIMENSIONAL SHAPE OPTIMIZATION; FEMORAL STEMS; INDIAN POPULATION; WALKING ACTIVITY; CONTACT FORCES; COBALT CHROME; ARTHROPLASTY; PROSTHESIS; JOINT;
D O I
10.5301/ijao.5000545
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Purpose: Human activities generate stresses, which vary with time and may result in fatigue failure of the customized hip implant. This study aims to investigate fatigue testing of customized hip implants using the minimum number of experiments by the Taguchi method, for 147 patients. This study was also useful to determine the influential geometrical parameters on the fatigue safety factor of customized hip implants. Methods: Horizontal offset (HO), vertical offset (VO) and neck shaft angle (NSA) of the hip joint of 147 patients were measured on computed tomography (CT) scanned images. Stress and strain of hip implants were calculated by finite element analysis and validated by in vitro experimental tests. Fatigue safety factors were calculated by Goodman, Soderberg and Gerber's fatigue theories and maximum stresses. Results: Analysis of variance results show that the highest impact on fatigue safety factors was equal to 54.38% for HO, 16.33% for VO, and was equal to 29.16% for NSA with reference to Goodman, Soderberg and Gerber's fatigue theories. The hip implant shape of experiment no. 8 has the highest safety factor value compared to all other hip implants. Conclusions: The results show that HO has the maximum influence on fatigue safety factors. The determination of influential geometric parameters may be useful to redesign customized hip implants in order to achieve the highest fatigue safety factor. The Taguchi method is suitable for fatigue testing of custom hip implant with a minimum number of experiments.
引用
收藏
页码:611 / 618
页数:8
相关论文
共 50 条
  • [21] A modified PMMA cement (Sub-cement) for accelerated fatigue testing of cemented implant constructs using cadaveric bone
    Race, Amos
    Miller, Mark A.
    Mann, Kenneth A.
    JOURNAL OF BIOMECHANICS, 2008, 41 (14) : 3017 - 3023
  • [22] Improved quantitative analysis method for evaluating fatigue cracks in thermal fatigue testing
    Qi, Yanfei
    Wang, Bo
    Li, Shudan
    Ren, Xiqiang
    Zhou, Jingyi
    Li, Yungang
    Mo, Jinjun
    MATERIALS LETTERS, 2019, 242 : 115 - 118
  • [23] Fatigue testing method and property changes of piezoelectric ceramics
    Sakai, T
    Kawamoto, H
    JOURNAL OF THE CERAMIC SOCIETY OF JAPAN, 1997, 105 (11) : 1032 - 1036
  • [24] Fatigue performance investigation on the automotive welded structure under damped loads using the Taguchi method
    Amarir, Imane
    Mounir, Hamid
    Rajad, Omar
    Amadane, Yassine
    HEAT TRANSFER, 2023, 52 (01) : 162 - 192
  • [25] Wear testing of a canine hip resurfacing implant that uses highly cross-linked polyethylene
    Warburton, Kevin J.
    Everingham, John B.
    Helms, Jillian L.
    Kazanovicz, Andrew J.
    Hollar, Katherine A.
    Brourman, Jeff D.
    Fox, Steven M.
    Lujan, Trevor J.
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2018, 36 (04) : 1196 - 1205
  • [26] Fracture strength of implant abutments after fatigue testing: A systematic review and a meta-analysis
    Coray, Rafaela
    Zeltner, Marco
    Ozcan, Mutlu
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2016, 62 : 333 - 346
  • [27] Dual Boundary Element Method applied in aircraft fuselage fatigue analysis
    Abreu de Oliveira, Thiago Arnaud
    IASS 60TH ANNIVERSARY SYMPOSIUM (IASS SYMPOSIUM 2019) - 9TH INTERNATIONAL CONFERENCE ON TEXTILE COMPOSITES AND INFLATABLE STRUCTURES (STRUCTURAL MEMBRANES 2019), 2019, : 1571 - 1578
  • [28] Modelling evaluation of the testing condition influence on the maximum stress induced in a hip prosthesis during ISO 7206 fatigue testing
    Raimondi, MT
    Pietrabissa, R
    MEDICAL ENGINEERING & PHYSICS, 1999, 21 (05) : 353 - 359
  • [29] Fatigue testing of materials under extremal conditions by acoustic method
    Baranov, VM
    Bibilashvili, YK
    Karasevich, VA
    Sarychev, GA
    TECHNICAL PHYSICS LETTERS, 2004, 30 (03) : 180 - 182
  • [30] Fatigue testing of materials under extremal conditions by acoustic method
    V. M. Baranov
    Yu. K. Bibilashvili
    V. A. Karasevich
    G. A. Sarychev
    Technical Physics Letters, 2004, 30 : 180 - 182