Consistency of Continuous Ambulatory Interstitial Glucose Monitoring Sensors

被引:4
|
作者
Wu, Pei T. [1 ,2 ]
Segovia, David E. [3 ,4 ]
Lee, Cathy C. [3 ,4 ]
Nguyen, Kim-Lien [1 ,2 ]
机构
[1] UCLA, David Geffen Sch Med, Div Cardiol, Los Angeles, CA 90024 USA
[2] VA Greater Los Angeles Healthcare Syst, Los Angeles, CA 90024 USA
[3] UCLA, David Geffen Sch Med, Dept Geriatr, Los Angeles, CA 90095 USA
[4] VA Greater Los Angeles Healthcare Syst, GRECC, Los Angeles, CA 90095 USA
来源
BIOSENSORS-BASEL | 2018年 / 8卷 / 02期
关键词
continuous glucose monitoring; interstitial glucose sensors; intra-subject agreement; postprandial glucose; BLOOD-GLUCOSE; ACCURACY; PERFORMANCE; SYSTEM; FLOW;
D O I
10.3390/bios8020049
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Aims: The abdominal region is the most common location for continuous glucose monitor (CGM) sensor insertion. However, a paucity of post-marketing data is available to demonstrate intra-individual consistency of CGM readings at different abdominal insertion sites. Methods: Healthy adults (fasting glucose (FG) < 5.5 mmol/L; BMI < 30 kg/m(2)) were recruited and a CGM sensor was placed on each side of the abdomen. Postprandial and continuous 48-h interstitial glucose levels were analyzed. Results: There was no significant difference in the 3-h postprandial glucose (PPG) level derived from the left versus right CGM, which remained non-significant after adjusting for waist circumference or FG. Among the glucose levels recorded over 48-h, values on the left site were greater in 3.6% of the data points (p < 0.05). After adjusting for waist circumference, only 0.5% of the glucose values remained significantly greater on the left (p < 0.05). When adjusted for FG, similar results were observed. For both PPG and 48-h readings, the mean absolute relative difference was not significant between the two abdominal sites. Conclusions: CGM-derived glucose measures were highly consistent between the left and right abdomen during both the postprandial and post-absorptive periods.
引用
收藏
页数:11
相关论文
共 50 条
  • [31] Continuous glucose monitoring in pediatrics:: Real life experience with DexCom™ sensors
    Mc Namara, Kim
    Barba, Laura A.
    Newfield, Ron S.
    DIABETES, 2008, 57 : A67 - A67
  • [33] Continuous Glucose Monitoring Sensors for Diabetes Management: A Review of Technologies and Applications
    Cappon, Giacomo
    Vettoretti, Martina
    Sparacino, Giovanni
    Facchinetti, Andrea
    DIABETES & METABOLISM JOURNAL, 2019, 43 (04) : 383 - 397
  • [34] Italian Contributions to the Development of Continuous Glucose Monitoring Sensors for Diabetes Management
    Sparacino, Giovanni
    Zanon, Mattia
    Facchinetti, Andrea
    Zecchin, Chiara
    Maran, Alberto
    Cobelli, Claudio
    SENSORS, 2012, 12 (10) : 13753 - 13780
  • [35] Fault Detection in Continuous Glucose Monitoring Sensors for Artificial Pancreas Systems
    Yu, Xia
    Rashid, Mudassir
    Feng, Jianyuan
    Hobbs, Nicole
    Hajizadeh, Iman
    Samadi, Sediqeh
    Sevil, Mert
    Lazaro, Caterina
    Maloney, Zacharie
    Cinar, Ali
    IFAC PAPERSONLINE, 2018, 51 (18): : 714 - 719
  • [36] Non-invasive wearable electrochemical sensors for continuous glucose monitoring
    Ameen, Sameera Sh. Mohammed
    Omer, Khalid M.
    Mansour, Fotouh R.
    Bedair, Alaa
    Hamed, Mahmoud
    ELECTROCHEMISTRY COMMUNICATIONS, 2025, 173
  • [37] Impact of Optical Fiber Characteristics in SPR Sensors for Continuous Glucose Monitoring
    Fallauto, Carmelo
    Perrone, Guido
    Vallan, Alberto
    2014 IEEE INTERNATIONAL SYMPOSIUM ON MEDICAL MEASUREMENTS AND APPLICATIONS (MEMEA), 2014, : 293 - 297
  • [38] Continuous Glucose Monitoring Sensors: Past, Present and Future Algorithmic Challenges
    Facchinetti, Andrea
    SENSORS, 2016, 16 (12)
  • [39] CONTINUOUS AMBULATORY URODYNAMIC MONITORING
    BHATIA, NN
    BRADLEY, WE
    HALDEMAN, S
    JOHNSON, BK
    BRITISH JOURNAL OF UROLOGY, 1982, 54 (04): : 357 - 359
  • [40] Estimating Plasma Glucose from Interstitial Glucose: The Issue of Calibration Algorithms in Commercial Continuous Glucose Monitoring Devices
    Rossetti, Paolo
    Bondia, Jorge
    Vehi, Josep
    Fanelli, Carmine G.
    SENSORS, 2010, 10 (12) : 10936 - 10952