Consistency of Continuous Ambulatory Interstitial Glucose Monitoring Sensors

被引:4
|
作者
Wu, Pei T. [1 ,2 ]
Segovia, David E. [3 ,4 ]
Lee, Cathy C. [3 ,4 ]
Nguyen, Kim-Lien [1 ,2 ]
机构
[1] UCLA, David Geffen Sch Med, Div Cardiol, Los Angeles, CA 90024 USA
[2] VA Greater Los Angeles Healthcare Syst, Los Angeles, CA 90024 USA
[3] UCLA, David Geffen Sch Med, Dept Geriatr, Los Angeles, CA 90095 USA
[4] VA Greater Los Angeles Healthcare Syst, GRECC, Los Angeles, CA 90095 USA
来源
BIOSENSORS-BASEL | 2018年 / 8卷 / 02期
关键词
continuous glucose monitoring; interstitial glucose sensors; intra-subject agreement; postprandial glucose; BLOOD-GLUCOSE; ACCURACY; PERFORMANCE; SYSTEM; FLOW;
D O I
10.3390/bios8020049
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Aims: The abdominal region is the most common location for continuous glucose monitor (CGM) sensor insertion. However, a paucity of post-marketing data is available to demonstrate intra-individual consistency of CGM readings at different abdominal insertion sites. Methods: Healthy adults (fasting glucose (FG) < 5.5 mmol/L; BMI < 30 kg/m(2)) were recruited and a CGM sensor was placed on each side of the abdomen. Postprandial and continuous 48-h interstitial glucose levels were analyzed. Results: There was no significant difference in the 3-h postprandial glucose (PPG) level derived from the left versus right CGM, which remained non-significant after adjusting for waist circumference or FG. Among the glucose levels recorded over 48-h, values on the left site were greater in 3.6% of the data points (p < 0.05). After adjusting for waist circumference, only 0.5% of the glucose values remained significantly greater on the left (p < 0.05). When adjusted for FG, similar results were observed. For both PPG and 48-h readings, the mean absolute relative difference was not significant between the two abdominal sites. Conclusions: CGM-derived glucose measures were highly consistent between the left and right abdomen during both the postprandial and post-absorptive periods.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Relationship between interstitial glucose variability in ambulatory glucose profile and standardized continuous glucose monitoring metrics; a pilot study
    Tokutsu, Akemi
    Okada, Yosuke
    Torimoto, Keiichi
    Tanaka, Yoshiya
    DIABETOLOGY & METABOLIC SYNDROME, 2020, 12 (01):
  • [2] Relationship between interstitial glucose variability in ambulatory glucose profile and standardized continuous glucose monitoring metrics; a pilot study
    Akemi Tokutsu
    Yosuke Okada
    Keiichi Torimoto
    Yoshiya Tanaka
    Diabetology & Metabolic Syndrome, 12
  • [3] Fluorescent Sensors for Continuous Glucose Monitoring
    Heo, Yun Jung
    2019 13TH IEEE INTERNATIONAL CONFERENCE ON NANO/MOLECULAR MEDICINE & ENGINEERING (IEEE-NANOMED 2019), 2019, : 60 - 60
  • [4] Optical sensors for continuous glucose monitoring
    Yu, Ziyi
    Jiang, Nan
    Kazarian, Sergei G.
    Tasoglu, Savas
    Yetisen, Ali K.
    PROGRESS IN BIOMEDICAL ENGINEERING, 2021, 3 (02):
  • [5] Continuous Monitoring of Interstitial Glucose: When and For What?
    Esteves, Cesar
    REVISTA PORTUGUESA DE ENDOCRINOLOGIA DIABETES E METABOLISMO, 2021, 16 (03) : 40 - 40
  • [6] Wearable textile sensors for continuous glucose monitoring
    Sunstrum, Frederique N.
    Khan, Jawairia Umar
    Li, Nga-Wun
    Welsh, Alec W.
    BIOSENSORS & BIOELECTRONICS, 2025, 273
  • [7] OVERVIEW OF MODERN SENSORS FOR CONTINUOUS GLUCOSE MONITORING
    Momynaliev, Kuvat T.
    Prokopiev, Maxim V.
    Ivanov, Igor V.
    DIABETES MELLITUS, 2023, 26 (06): : 575 - 584
  • [8] Differences in glucose readings between the continuous glucose monitoring calibration free interstitial sensors versus capillary blood glucose monitoring by glucometer: An analysis of two cases
    Quah, Phaik Ling
    Chai, Sally Mun Hua
    Tan, Kok Hian
    METABOLISM OPEN, 2024, 22
  • [9] Interstitial glucose concentration and glycemia: implications for continuous subcutaneous glucose monitoring
    Aussedat, B
    Dupire-Angel, M
    Gifford, R
    Klein, JC
    Wilson, GS
    Reach, G
    AMERICAN JOURNAL OF PHYSIOLOGY-ENDOCRINOLOGY AND METABOLISM, 2000, 278 (04): : E716 - E728
  • [10] Patent analysis of digital sensors for continuous glucose monitoring
    Litvinova, Olena
    Eitenberger, Magdalena
    Bilir, Aylin
    Yeung, Andy Wai Kan
    Parvanov, Emil D. D.
    MohanaSundaram, ArunSundar
    Horbanczuk, Jaroslaw Olav
    Atanasov, Atanas G. G.
    Willschke, Harald
    FRONTIERS IN PUBLIC HEALTH, 2023, 11