Study of a multi-wire proportional chamber with a cathode strip and delay-line readout

被引:3
|
作者
Han Li-Ying [1 ]
Li Qi-Te
Faisal, Q.
Ge Yu-Cheng
Liu Hong-Tao
Ye Yan-Liu
机构
[1] Peking Univ, Sch Phys, Beijing 100871, Peoples R China
关键词
multi-wire proportional chamber; delay-line readout; cathode strip; position resolution;
D O I
10.1088/1674-1137/33/5/009
中图分类号
O57 [原子核物理学、高能物理学];
学科分类号
070202 ;
摘要
The design principle for a multi-wire proportional chamber with a cathode strip and delay-line readout is described. A prototype chamber of a size of 10 cm x 10 cm was made together with the readout electronics circuit. A very clean signal with very low background noise was obtained by applying a transformer between the delay-line and the pre-amplifier in order to match the resistance. Along the anode wire direction a position resolution of less than 0.5 mm was achieved with a Fe-55-5.9 keV X ray source. The simple structure, large effective area and high position resolution allow the application of a gas chamber of this kind to many purposes.
引用
收藏
页码:364 / 368
页数:5
相关论文
共 50 条
  • [1] Study of a multi-wire proportional chamber with a cathode strip and delay-line readout
    韩丽英
    李奇特
    葛愉成
    刘洪涛
    叶沿林
    中国物理C, 2009, 33 (05) : 364 - 368
  • [2] Delay line readout of a large dimension multi-wire proportional chamber
    黎先利
    张云龙
    钱浩
    许咨宗
    汪晓莲
    中国物理C, 2012, (06) : 519 - 524
  • [3] Delay line readout of a large dimension multi-wire proportional chamber
    Li Xian-Li
    Zhang Yun-Long
    Qian Hao
    Xu Zi-Zong
    Wang Xiao-Lian
    CHINESE PHYSICS C, 2012, 36 (06) : 519 - 524
  • [4] DELAY-LINE READOUT OF ANODE WIRES IN A MULTIWIRE PROPORTIONAL CHAMBER
    BRYMAN, D
    CRESSWELL, J
    SKEGG, R
    NUCLEAR INSTRUMENTS & METHODS, 1977, 141 (03): : 573 - 576
  • [5] ADAPTATION OF MULTI-WIRE PROPORTIONAL CHAMBERS WITH DELAY-LINE READOUTS FOR NEUTRON RADIOGRAPHIC IMAGING
    VALENTINE, K
    KAUFMAN, L
    KAPLAN, S
    PEREZMEN.V
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1972, NS19 (01) : 374 - +
  • [6] AN ELECTROMAGNETIC SHOWER DETECTOR USING PROPORTIONAL WIRE CHAMBERS WITH CATHODE PLANE DELAY-LINE READOUT
    NUNAMAKER, TA
    HEISTERBERG, RH
    MO, LW
    NUCLEAR INSTRUMENTS & METHODS, 1980, 175 (2-3): : 331 - 334
  • [7] Readout method for two-dimensional multi-wire proportional chamber
    Wen Zhi-Wen
    Qi Hui-Rong
    Wang Yan-Feng
    Sun Zhi-Jia
    Zhang Yu-Lian
    Wang Hai-Yun
    Zhang Jian
    Ouyang Qun
    Chen Yuan-Bo
    Li Yu-Hong
    ACTA PHYSICA SINICA, 2017, 66 (07)
  • [8] DELAY-LINE READOUT FOR PROPORTIONAL CHAMBERS
    RINDI, A
    PEREZMEN.V
    WALLACE, RI
    NUCLEAR INSTRUMENTS & METHODS, 1970, 77 (02): : 325 - +
  • [9] The Study of Boron-Lined Honeycomb Neutron Detector with a Multi-Wire Proportional Chamber Readout
    Liu, Lixing
    Yu, Yangyi
    Fang, Zhujun
    Deng, Chao
    Zhang, Zhi
    Yang, Yigang
    2019 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2019,
  • [10] 2-DIMENSIONAL READOUT OF A MULTIWIRE PROPORTIONAL CHAMBER USING A HELICAL DELAY-LINE
    SITAR, B
    STUBBS, RJ
    BREARE, JM
    NUCLEAR INSTRUMENTS & METHODS, 1976, 134 (02): : 267 - 270