Design, Control, and Preliminary Assessment of a Multifunctional Semipowered Ankle Prosthesis

被引:24
作者
Bartlett, Harrison L. [1 ]
Lawson, Brian E. [1 ]
Goldfarb, Michael [2 ]
机构
[1] Synchro Mot LLC, Franklin, TN 37064 USA
[2] Vanderbilt Univ, Dept Mech Engn, Nashville, TN 37240 USA
基金
美国国家科学基金会;
关键词
Amputation; ankle; biomechanics; control; design; prostheses; FOOT PROSTHESIS; STAIR AMBULATION; WALKING; PREVALENCE; DESCENT; PEOPLE; RISK; FEAR;
D O I
10.1109/TMECH.2019.2918685
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
This paper describes the design, control, and preliminary assessment of a novel microprocessor-controlled multifunction ankle prosthesis that provides three microprocessor-controlled behaviors-a selectable stiffness equilibrium angle, lockable conformal damping, and swing-phase repositioning. Following a description of the motivation for providing these behaviors, the authors provide a detailed description of the device and walking controller design. This device utilizes a power-asymmetric linear actuator to provide the desired functionality in a compact and lightweight package through a combination of both hydraulic and electromechanical actuation approaches. The device is controlled for level ground walking via a finite-state machine. The functionality of the prosthesis is demonstrated by a set of benchtop experiments that characterize the ability of the prosthesis to provide the three desired behaviors and by an experiment in which the prosthesis was worn by a transtibial amputee during walking. Both sets of experiments indicate that the prosthesis provides the functionality for which it was designed.
引用
收藏
页码:1532 / 1540
页数:9
相关论文
共 47 条
[1]   Kinematics and kinetics with an adaptive ankle foot system during stair ambulation of transtibial amputees [J].
Alimusaj, Merkur ;
Fradet, Laetitia ;
Braatz, Frank ;
Gerner, Hans J. ;
Wolf, Sebastian I. .
GAIT & POSTURE, 2009, 30 (03) :356-363
[2]   Powered ankle-foot prosthesis to assist level-ground and stair-descent gaits [J].
Au, Samuel ;
Berniker, Max ;
Herr, Hugh .
NEURAL NETWORKS, 2008, 21 (04) :654-666
[3]  
Au SK, 2008, IEEE ROBOT AUTOM MAG, V15, P52, DOI 10.1109/MRA.2008.927697
[4]   Powered Ankle-Foot Prosthesis Improves Walking Metabolic Economy [J].
Au, Samuel K. ;
Weber, Jeff ;
Herr, Hugh .
IEEE TRANSACTIONS ON ROBOTICS, 2009, 25 (01) :51-66
[5]  
Bartlett HL, 2017, INT C REHAB ROBOT, P1531, DOI 10.1109/ICORR.2017.8009465
[6]   Optimal Transmission Ratio Selection for Electric Motor Driven Actuators With Known Output Torque and Motion Trajectories [J].
Bartlett, Harrison L. ;
Lawson, Brian E. ;
Goldfarb, Michael .
JOURNAL OF DYNAMIC SYSTEMS MEASUREMENT AND CONTROL-TRANSACTIONS OF THE ASME, 2017, 139 (10)
[7]   SPARKy 3: Design of an Active Robotic Ankle Prosthesis with Two Actuated Degrees of Freedom Using Regenerative Kinetics [J].
Bellman, Ryan D. ;
Holgate, Matthew A. ;
Sugar, Thomas G. .
2008 2ND IEEE RAS & EMBS INTERNATIONAL CONFERENCE ON BIOMEDICAL ROBOTICS AND BIOMECHATRONICS (BIOROB 2008), VOLS 1 AND 2, 2008, :511-+
[8]   Concept Through Preliminary Bench Testing of a Powered Lower Limb Prosthetic Device [J].
Bergelin, Bryan J. ;
Mattos, Javier O. ;
Wells, Joseph G., Jr. ;
Voglewede, Philip A. .
JOURNAL OF MECHANISMS AND ROBOTICS-TRANSACTIONS OF THE ASME, 2010, 2 (04)
[9]  
Burger H., 1997, DISABIL REHABIL, V19, P272
[10]  
Cempini M, 2017, IEEE INT C INT ROBOT, P1064, DOI 10.1109/IROS.2017.8202276