Technical Note: Quantitative accuracy evaluation for spectral images from a detector-based spectral CT scanner using an iodine phantom

被引:18
作者
Duan, Xinhui [1 ]
Arbique, Gary [1 ]
Guild, Jeffrey [1 ]
Xi, Yin [1 ]
Anderson, Jon [1 ]
机构
[1] Univ Texas Southwestern Med Ctr Dallas, Dept Radiol, Dallas, TX 75390 USA
关键词
dual-layer detector; phantom; quantitative evaluation; spectral CT; DUAL-ENERGY CT; COMPUTED-TOMOGRAPHY; QUANTIFICATION; EXPERIENCE; SYSTEM;
D O I
10.1002/mp.12834
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeThe purpose of this study was to evaluate the quantitative accuracy of spectral images from a detector-based spectral CT scanner using a phantom with iodine-loaded inserts. MethodsA 40-cm long-body phantom with seven iodine inserts (2-20 mg/ml of iodine) was used in the study. The inserts could be placed at 5.5 or 10.5 cm from the phantom axis. The phantom was scanned five times for each insert configuration using 120 kVp tube voltage. A set of iodine, virtual noncontrast, effective atomic number, and virtual monoenergetic spectral CT images were generated and measurements were made for all the iodine rods. Measured values were compared with reference values calculated from the chemical composition information provided by the phantom manufacturer. Radiation dose from the spectral CT was compared to a conventional CT using a CTDI (32 cm) phantom. ResultsGood agreement between measurements and reference values was achieved for all types of spectral images. The differences ranged from -0.46 to 0.1 mg/ml for iodine concentration, -9.95 to 6.41 HU for virtual noncontrast images, 0.12 to 0.35 for effective Z images, and -17.7 to 55.7 HU for virtual monoenergetic images. For a similar CTDIvol, image noise from the conventional CT was 10% lower than the spectral CT. ConclusionsThe detector-based spectral CT can achieve accurate spectral measurements on iodine concentration, virtual non-contrast images, effective atomic numbers, and virtual monoenergetic images.
引用
收藏
页码:2048 / 2053
页数:6
相关论文
共 18 条
[1]   ENERGY-SELECTIVE RECONSTRUCTIONS IN X-RAY COMPUTERIZED TOMOGRAPHY [J].
ALVAREZ, RE ;
MACOVSKI, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1976, 21 (05) :733-744
[2]  
[Anonymous], 2010, XCOM: Photon Cross Section Database (version 1.5)
[3]   Iodine Quantification With Dual-Energy CT: Phantom Study and Preliminary Experience With Renal Masses [J].
Chandarana, Hersh ;
Megibow, Alec J. ;
Cohen, Benjamin A. ;
Srinivasan, Ramya ;
Kim, Danny ;
Leidecker, Christianne ;
Macari, Michael .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2011, 196 (06) :W693-W700
[4]   First performance evaluation of a dual-source CT (DSCT) system [J].
Flohr, TG ;
McCollough, CH ;
Bruder, H ;
Petersilka, M ;
Gruber, K ;
Süss, C ;
Grasruck, M ;
Stierstorfer, K ;
Krauss, B ;
Raupach, R ;
Primak, AN ;
Küttner, A ;
Achenbach, S ;
Becker, C ;
Kopp, A ;
Ohnesorge, BM .
EUROPEAN RADIOLOGY, 2006, 16 (02) :256-268
[5]   Accuracies of the synthesized monochromatic CT numbers and effective atomic numbers obtained with a rapid kVp switching dual energy CT scanner [J].
Goodsitt, Mitchell M. ;
Christodoulou, Emmanuel G. ;
Larson, Sandra C. .
MEDICAL PHYSICS, 2011, 38 (04) :2222-2232
[6]   Density and atomic number measurements with spectral x-ray attenuation method [J].
Heismann, BJ ;
Leppert, J ;
Stierstorfer, K .
JOURNAL OF APPLIED PHYSICS, 2003, 94 (03) :2073-2079
[7]   COMPUTERIZED TRANSVERSE AXIAL SCANNING (TOMOGRAPHY) .1. DESCRIPTION OF SYSTEM [J].
HOUNDFIELD, GN .
BRITISH JOURNAL OF RADIOLOGY, 1973, 46 (552) :1016-1022
[8]   Material differentiation by dual energy CT: initial experience [J].
Johnson, Thorsten R. C. ;
Krauss, Bernhard ;
Sedlmair, Martin ;
Grasruck, Michael ;
Bruder, Herbert ;
Morhard, Dominik ;
Fink, Christian ;
Weckbach, Sabine ;
Lenhard, Miriam ;
Schmidt, Bernhard ;
Flohr, Thomas ;
Reiser, Maximilian F. ;
Becker, Christoph R. .
EUROPEAN RADIOLOGY, 2007, 17 (06) :1510-1517
[9]   EVALUATION OF A PROTOTYPE DUAL-ENERGY COMPUTED TOMOGRAPHIC APPARATUS .1. PHANTOM STUDIES [J].
KALENDER, WA ;
PERMAN, WH ;
VETTER, JR ;
KLOTZ, E .
MEDICAL PHYSICS, 1986, 13 (03) :334-339
[10]   Accuracy of dual-energy computed tomography for the quantification of iodine in a soft tissue-mimicking phantom [J].
Li, Jung-Hui ;
Du, Yeh-Ming ;
Huang, Hsuan-Ming .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (05) :418-426