How thin can you go? Performance of thin copper and aluminum RF coil conductors

被引:8
作者
Barta, Radim [1 ]
Volotovskyy, Vyacheslav [2 ]
Wachowicz, Keith [1 ,3 ]
Fallone, B. Gino [1 ,3 ,4 ]
De Zanche, Nicola [1 ,3 ]
机构
[1] Univ Alberta, Div Med Phys, Dept Oncol, 11560 Univ Ave, Edmonton, AB T6G 1Z2, Canada
[2] MagnetTx Oncol Solut Ltd, Edmonton, AB, Canada
[3] Cross Canc Inst, Dept Med Phys, Edmonton, AB, Canada
[4] Univ Alberta, Dept Phys, Edmonton, AB, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
alternate conductors; aluminum; efficiency; quality factor; RF coils; SNR; thin conductors; TO-NOISE RATIO; ARRAY;
D O I
10.1002/mrm.28540
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose To evaluate the impact of emerging conductor technology on RF coils. Performance and resulting image quality of thin or alternate conductors (eg, aluminum instead of copper) and thicknesses (9-600 mu m) are compared in terms of SNR. Methods Eight prototype RF coils (15 cm x 15 cm square loops) were constructed and bench-tested to measure quality factor. The coils used 6-mm-wide conducting strips of either copper or aluminum of a few different thicknesses (copper: 17, 32, 35, 127, 600 mu m; aluminum: 9, 13, 20, 127 mu m) on acetate projector sheets for backing. Corresponding image SNR was measured at 0.48 tesla (20.56 MHz). Results The coils spanned a range of unloaded quality factors from 89 to 390 and a fivefold range of losses. The image SNRs were consistent with the coils' bench-measured efficiencies (0.33-0.73). Thin aluminum conductors (9 mu m) led to the highest reduction in SNR (65% that of 127 mu m copper). Thin copper (<32 mu m) conductors lead to a much smaller decrease in SNR (approximately 10%) compared to 127 mu m copper. No performance difference was observed between 127 mu m thick copper and aluminum. The much thicker 600 mu m copper bars only yield a 5% improvement in SNR. Conclusion Even at 0.48 tesla, copper RF coil conductors much thinner than those in conventional construction can be used while maintaining SNR greater than 50% that of thick copper. These emerging coil conductor technologies enable RF coil functionality that cannot be achieved otherwise.
引用
收藏
页码:2327 / 2333
页数:7
相关论文
共 20 条
[1]  
[Anonymous], 2004, Physics for scientist and engineers
[2]  
Aster R., 2005, Parameter estimation and inverse problems
[3]   Impact of a parallel magnetic field on radiation dose beneath thin copper and aluminum foils [J].
Barta, Radim ;
Ghila, Andrei ;
Rathee, Satyapal ;
Fallone, B. Gino ;
De Zanche, Nicola .
BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2020, 6 (03)
[4]  
Chen C.N., 1989, Biomedical Magnetic Resonance Technology
[5]   Screen-printed flexible MRI receive coils [J].
Corea, Joseph R. ;
Flynn, Anita M. ;
Lechene, Balthazar ;
Scott, Greig ;
Reed, Galen D. ;
Shin, Peter J. ;
Lustig, Michael ;
Arias, Ana C. .
NATURE COMMUNICATIONS, 2016, 7
[6]   THE INTRINSIC SIGNAL-TO-NOISE RATIO IN NMR IMAGING [J].
EDELSTEIN, WA ;
GLOVER, GH ;
HARDY, CJ ;
REDINGTON, RW .
MAGNETIC RESONANCE IN MEDICINE, 1986, 3 (04) :604-618
[7]   The Rotating Biplanar Linac-Magnetic Resonance Imaging System [J].
Fallone, Biagio Gino .
SEMINARS IN RADIATION ONCOLOGY, 2014, 24 (03) :200-202
[8]   Conductor geometry and capacitor quality for performance optimization of low-frequency birdcage coils [J].
Giovannetti, G ;
Francesconi, R ;
Landini, L ;
Santarelli, MF ;
Positano, V ;
Viti, V ;
Benassi, A .
CONCEPTS IN MAGNETIC RESONANCE PART B-MAGNETIC RESONANCE ENGINEERING, 2004, 20B (01) :9-16
[9]   Estimation of losses in strip and circular wire conductors of radiofrequency planar surface coil by using the finite element method [J].
Giovannetti, Giulio ;
Fontana, Nunzia ;
Monorchio, Agostino ;
Tosetti, Michela ;
Tiberi, Gianluigi .
CONCEPTS IN MAGNETIC RESONANCE PART B-MAGNETIC RESONANCE ENGINEERING, 2017, 47B (03)
[10]   Classical and lateral skin effect contributions estimation in strip MR coils [J].
Giovannetti, Giulio ;
Hartwig, Valentina ;
Landini, Luigi ;
Santarelli, Maria Filomena .
CONCEPTS IN MAGNETIC RESONANCE PART B-MAGNETIC RESONANCE ENGINEERING, 2012, 41B (02) :57-61