Solvent selectivity and strength in reversed-phase liquid chromatography separation of peptides

被引:17
作者
Gilar, Martin [1 ]
Jaworski, Aleksander
McDonald, Thomas S. [1 ]
机构
[1] Waters Corp, Milford, MA 01757 USA
关键词
Peptides; Separation selectivity; Eluotropic strength; Reversed-phase; Method transfer; lsoeluotropic conditions; SOLVATION ENERGY RELATIONSHIPS; RETENTION TIMES; AMINO-ACID; SIDE-CHAINS; COLUMN SELECTIVITY; ORGANIC MODIFIERS; STATIONARY PHASES; GRADIENT ELUTION; GROWTH-HORMONE; MODEL PEPTIDES;
D O I
10.1016/j.chroma.2014.02.048
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A set of tryptic peptides was analyzed in reversed-phase liquid chromatography using gradient elution with acetonitrile, methanol, or isopropanol. We used these retention data as training sets to develop retention prediction models of peptides for the three organic eluents used. The coefficients of determination, R-2, between predicted and observed data were approximately 0.95 for all systems. Retention coefficient values of twenty amino acids calculated from a model were utilized to investigate differences in separation selectivity between acetonitrile, methanol, or isopropanol eluents. The experimentally observed difference in separation selectivity appears to be a complex interplay of multiple amino acids, each contributing to a different degree to overall peptide retention. While retention contribution of hydrophilic amino acids was higher in methanol than acetonitrile, peptides containing aromatic amino acids (tyrosine, phenylalanine, tryptophan) exhibit relatively lower retention in methanol compared to acetonitrile. The differences between acetonitrile and isopropanol eluents were less pronounced. We also compared the relative elution strength of the three organic eluents for peptides. The relationship between the elution strength of two solvents is not linear, rather it was best fitted by a cubic polynomial function. Three solvents can be arranged in the order of increasing elution power as methanol < acetonitrile < isopropanol. The equations for relative solvent strength conversion were proposed. (C) 2014 Elsevier B.V. All rights reserved.
引用
收藏
页码:140 / 146
页数:7
相关论文
共 57 条
[1]  
[Anonymous], 1989, MULTIVARIATE CALIBRA
[2]   Replacement of acetonitrile by ethanol as solvent in reversed phase chromatography of biomolecules [J].
Brettschneider, F. ;
Jankowski, V. ;
Guenthner, T. ;
Salem, S. ;
Nierhaus, M. ;
Schulz, A. ;
Zidek, W. ;
Jankowski, J. .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2010, 878 (9-10) :763-768
[3]   Use of an integrated MS-multiplexed MS/MS data acquisition strategy for high-coverage peptide mapping studies [J].
Chakraborty, Asish B. ;
Berger, Scott J. ;
Gebler, John C. .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2007, 21 (05) :730-744
[4]  
Chang JP, 1997, RAPID COMMUN MASS SP, V11, P1266, DOI 10.1002/(SICI)1097-0231(199708)11:12<1266::AID-RCM977>3.0.CO
[5]  
2-3
[6]   Optimum concentration of trifluoroacetic acid for reversed-phase liquid chromatography of peptides revisited [J].
Chen, Y ;
Mehok, AR ;
Mant, CT ;
Hodges, RS .
JOURNAL OF CHROMATOGRAPHY A, 2004, 1043 (01) :9-18
[7]   Temperature selectivity effects in reversed-phase liquid chromatography due to conformation differences between helical and non-helical peptides [J].
Chen, YX ;
Mant, CT ;
Hodges, RS .
JOURNAL OF CHROMATOGRAPHY A, 2003, 1010 (01) :45-61
[8]   COMPUTER-SIMULATION AS A TOOL FOR THE RAPID OPTIMIZATION OF THE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC SEPARATION OF A TRYPTIC DIGEST OF HUMAN GROWTH-HORMONE [J].
CHLOUPEK, RC ;
HANCOCK, WS ;
SNYDER, LR .
JOURNAL OF CHROMATOGRAPHY, 1992, 594 (1-2) :65-73
[9]   FACTORS INFLUENCING THE PERFORMANCE OF PEPTIDE-MAPPING BY REVERSED-PHASE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
DONG, MW ;
TRAN, AD .
JOURNAL OF CHROMATOGRAPHY, 1990, 499 :125-139
[10]   Characterizing the selectivity of stationary phases and organic modifiers in reversed-phase high-performance liquid chromatographic systems by a general solvation equation using gradient elution [J].
Du, CM ;
Valko, K ;
Bevan, C ;
Reynolds, D ;
Abraham, MH .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 2000, 38 (11) :503-511