Ionization chamber correction factors for MR-linacs

被引:47
作者
Pojtinger, Stefan [1 ,2 ]
Dohm, Oliver Steffen [2 ]
Kapsch, Ralf-Peter [1 ]
Thorwarth, Daniela [2 ]
机构
[1] Phys Tech Bundesanstalt, Braunschweig, Germany
[2] Univ Hosp Tubingen, Dept Radiat Oncol, Sect Biomed Phys, Tubingen, Germany
关键词
MR-guided radiotherapy; reference dosimetry; Monte Carlo; EGSnrc; MR-linac; ionization chamber; MONTE-CARLO SIMULATIONS; QUALITY CORRECTION FACTORS; MAGNETIC-FIELD CORRECTION; ENERGY PHOTON BEAMS; REFERENCE DOSIMETRY; RADIOTHERAPY; ACCELERATOR; ALGORITHM; TRANSPORT; PROSTATE;
D O I
10.1088/1361-6560/aac4f2
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Previously, readings of air-filled ionization chambers have been described as being influenced by magnetic fields. To use these chambers for dosimetry in magnetic resonance guided radiotherapy (MRgRT), this effect must be taken into account by introducing a correction factor k(B). The purpose of this study is to systematically investigate k(B) for a typical reference setup for commercially available ionization chambers with different magnetic field strengths. The Monte Carlo simulation tool EGSnrc was used to simulate eight commercially available ionization chambers in magnetic fields whose magnetic flux density was in the range of 0-2.5 T. To validate the simulation, the influence of the magnetic field was experimentally determined for a PTW30013 Farmer-type chamber for magnetic flux densities between 0 and 1.425 T. Changes in the detector response of up to 8% depending on the magnetic flux density, on the chamber geometry and on the chamber orientation were obtained. In the experimental setup, a maximum deviation of less than 2% was observed when comparing measured values with simulated values. Dedicated values for two MR-linac systems (ViewRay MRldian, ViewRay Inc, Cleveland, United States, 0.35 T/ 6 MV and Elekta Unity, Elekta AB, Stockholm, Sweden, 1.5 T/7 MV) were determined for future use in reference dosimetry. Simulated values for thimble-type chambers are in good agreement with experiments as well as with the results of previous publications. After further experimental validation, the results can be considered for definition of standard protocols for purposes of reference dosimetry in MRgRT.
引用
收藏
页数:8
相关论文
共 27 条
[1]   Evaluation of a commercial MRI Linac based Monte Carlo dose calculation algorithm with GEANT4 [J].
Ahmad, Syed Bilal ;
Sarfehnia, Arman ;
Paudel, Moti Raj ;
Kim, Anthony ;
Hissoiny, Sami ;
Sahgal, Arjun ;
Keller, Brian .
MEDICAL PHYSICS, 2016, 43 (02) :894-907
[2]  
Andreo P, 2006, Technical reports series
[3]   Application of an adapted Fano cavity test for Monte Carlo simulations in the presence of B-fields [J].
de Pooter, J. A. ;
de Prez, L. A. ;
Bouchard, H. .
PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (24) :9313-9327
[4]   A real-time MRI guided external beam radiotherapy delivery system [J].
Dempsey, J. ;
Dionne, B. ;
Fitzsimmons, J. ;
Haghigat, A. ;
Li, J. ;
Low, D. ;
Mutic, S. ;
Palta, J. ;
Romeijn, H. ;
Sjoden, G. .
MEDICAL PHYSICS, 2006, 33 (06) :2254-2254
[5]  
Deutsches Institut fur Normung, 2008, 68002 DIN
[6]   The Rotating Biplanar Linac-Magnetic Resonance Imaging System [J].
Fallone, Biagio Gino .
SEMINARS IN RADIATION ONCOLOGY, 2014, 24 (03) :200-202
[7]  
Kawrakow I, 2017, PIRS701 NRCC
[8]   The Australian Magnetic Resonance Imaging-Linac Program [J].
Keall, Paul J. ;
Barton, Michael ;
Crozier, Stuart .
SEMINARS IN RADIATION ONCOLOGY, 2014, 24 (03) :203-206
[9]   Charged particle transport in magnetic fields in EGSnrc [J].
Malkov, V. N. ;
Rogers, D. W. O. .
MEDICAL PHYSICS, 2016, 43 (07) :4447-4458
[10]   Sensitive volume effects on Monte Carlo calculated ion chamber response in magnetic fields [J].
Malkov, Victor N. ;
Rogers, D. W. O. .
MEDICAL PHYSICS, 2017, 44 (09) :4854-4858