Pacemaker-mediated tachycardia: Management by pacemaker interrogation/reprogramming in the ED

被引:4
|
作者
Sobel, RM
Donaldson, PR
Dhruva, N
机构
[1] Emory Univ, Sch Med, Dept Emergency Med, Atlanta, GA USA
[2] Cardiol Dis Specialists, Atlanta, GA USA
来源
AMERICAN JOURNAL OF EMERGENCY MEDICINE | 2002年 / 20卷 / 04期
关键词
pacemaker malfunction; pacemaker-mediated tachycardia; pacemaker interrogation and programming; atrial flutter; overdrive pacing; emergency department;
D O I
10.1053/ajem.2002.33780
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
We report a case of a patient with symptomatic pacemaker-mediated tachycardia (PMT). PMT should be suspected when asynchronous pacing induced by magnet application causes temporary resolution of an unknown tachycardia in a patient with a pacemaker. The underlying mechanism of PMT is either reentrant tachycardia or atrial triggering. PMT is better discerned with an intracardiac electrocardiogram (ECG) obtained through pacemaker interrogation (PI) than the standard 12-lead ECG. In the case reported, the interrogator was used to demonstrate that the initiating dysrhythmia was atrial flutter. Next, the programmer was used to overdrive the atrium restoring normal sinus rhythm. After pacemaker reprogramming, the patient was discharged from the emergency department (ED) in improved condition. We believe this to be the first reported case of a patient with PMT diagnosed, treated, and discharged effectively after PI, overdrive pacing and reprogramming from the ED. The pacemaker interrogator/programmer is a rapid noninvasive device that has practical applications in the ED. Copyright 2002, Elsevier Science (USA). All rights reserved.
引用
收藏
页码:336 / 339
页数:4
相关论文
共 37 条
  • [31] Endless loop tachycardia below the upper tracking rate of a pacemaker: A case report
    Sakamoto, Atsushi
    Takeuchi, Ryosuke
    Hosoya, Natsuko
    Kageyama, Shigetaka
    Kajihara, Jun
    Takahashi, Kosuke
    Kurabe, Takashi
    Murata, Koichiro
    Nawada, Ryuzo
    Onodera, Tomoya
    Takizawa, Akinori
    Nomura, Ryota
    Nakai, Masanao
    JOURNAL OF ARRHYTHMIA, 2012, 28 (06) : 356 - 359
  • [32] Unusual atrial-paced tachycardia after pacemaker implantation: What is the mechanism?
    Viswanathan, Mohan N.
    Tseng, Zian H.
    Godschlager, Nora
    Lee, Byron K.
    HEART RHYTHM, 2007, 4 (06) : 800 - 802
  • [33] PACEMAKER MEDIATED TACHYCARDIAS IN SINGLE CHAMBER RATE RESPONSIVE PACING
    LAU, CP
    TAI, YT
    FONG, PC
    CHENG, CH
    CHUNG, FLW
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (12): : 1575 - 1579
  • [34] When two hearts do not beat as one - An unusual cause of pacemaker related tachycardia
    Zormpas, Christos
    Mueller-Leisse, Johanna
    Duncker, David
    Veltmann, Christian
    JOURNAL OF ELECTROCARDIOLOGY, 2019, 57 : 6 - 9
  • [35] Initial clinical experience with a new automatic PMT detection and termination algorithm to discriminate between pacemaker mediated tachycardia due to ventriculo-atrial conduction and normal sinus tachycardia.
    Cameron, DA
    Gentzler, RD
    Love, CJ
    Griffin, J
    Fletcher, R
    Hummel, J
    Manubens, S
    Gibbs, K
    McElroy, B
    HEARTWEB, 1996, 2 (01): : U263 - U268
  • [36] Anesthetic management of an AAI pacemaker patient with paroxysmal atrial fibrillation during colorectal surgery
    Lin, Jui-An
    Borel, Cecil O.
    Wang, Wen-Been
    Wong, Chih-Shung
    Yeh, Chun-Chang
    Yang, Chih-Ping
    Wu, Ching-Tang
    JOURNAL OF CLINICAL ANESTHESIA, 2006, 18 (05) : 372 - 375
  • [37] "PVC response Atrial-Pace," an algorithm designed for preventing pacemaker-induced tachycardia after premature ventricular contractions, triggers atrial high rate episodes
    Sane, Markus
    Marjamaa, Annukka
    Kuusisto, Jouni
    Raatikainen, Pekka
    Karvonen, Jarkko
    HEART RHYTHM, 2024, 21 (04) : 495 - 496