Pacemaker-mediated tachycardia: Management by pacemaker interrogation/reprogramming in the ED

被引:4
作者
Sobel, RM
Donaldson, PR
Dhruva, N
机构
[1] Emory Univ, Sch Med, Dept Emergency Med, Atlanta, GA USA
[2] Cardiol Dis Specialists, Atlanta, GA USA
关键词
pacemaker malfunction; pacemaker-mediated tachycardia; pacemaker interrogation and programming; atrial flutter; overdrive pacing; emergency department;
D O I
10.1053/ajem.2002.33780
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
We report a case of a patient with symptomatic pacemaker-mediated tachycardia (PMT). PMT should be suspected when asynchronous pacing induced by magnet application causes temporary resolution of an unknown tachycardia in a patient with a pacemaker. The underlying mechanism of PMT is either reentrant tachycardia or atrial triggering. PMT is better discerned with an intracardiac electrocardiogram (ECG) obtained through pacemaker interrogation (PI) than the standard 12-lead ECG. In the case reported, the interrogator was used to demonstrate that the initiating dysrhythmia was atrial flutter. Next, the programmer was used to overdrive the atrium restoring normal sinus rhythm. After pacemaker reprogramming, the patient was discharged from the emergency department (ED) in improved condition. We believe this to be the first reported case of a patient with PMT diagnosed, treated, and discharged effectively after PI, overdrive pacing and reprogramming from the ED. The pacemaker interrogator/programmer is a rapid noninvasive device that has practical applications in the ED. Copyright 2002, Elsevier Science (USA). All rights reserved.
引用
收藏
页码:336 / 339
页数:4
相关论文
共 15 条
[1]  
Barold S S, 1985, Cardiol Clin, V3, P565
[2]   PACEMAKER-MEDIATED TACHYCARDIA - ENGINEERING SOLUTIONS [J].
CALFEE, RV .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (11) :1917-1928
[3]   Permanent cardiac pacemakers: Issues relevant to the emergency physician, part I [J].
Cardall, TY ;
Chan, TC ;
Brady, WJ ;
Perry, JC ;
Vilke, GM ;
Rosen, P .
JOURNAL OF EMERGENCY MEDICINE, 1999, 17 (03) :479-489
[4]   Permanent cardiac pacemakers: Issues relevant to the emergency physician, part II [J].
Cardall, TY ;
Brady, WJ ;
Chan, TC ;
Perry, JC ;
Vilke, GM ;
Rosen, P .
JOURNAL OF EMERGENCY MEDICINE, 1999, 17 (04) :697-709
[5]   TERMINATION OF PACEMAKER-MEDIATED TACHYCARDIA BY ADENOSINE [J].
CONTI, JB ;
CURTIS, AB ;
HILL, JA ;
RAYMENANTS, ER .
CLINICAL CARDIOLOGY, 1994, 17 (01) :47-48
[6]   PREVENTION AND TERMINATION OF PACEMAKER-MEDIATED TACHYCARDIA IN A NEW DDD PACING SYSTEM (SIEMENS-PACESETTER MODEL-2010T) [J].
DUNCAN, JL ;
CLARK, MF .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (11) :1679-1683
[7]   Variable effects of adenosine on retrograde conduction in patients with atrioventricular nodal reentry tachycardia [J].
Fishberger, SB ;
Mehta, D ;
Rossi, AF ;
Langan, MN ;
Marx, SO ;
Gomes, JA .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (06) :1254-1257
[8]   STEPWISE APPROACH TO THE UNKNOWN PACEMAKER ECG [J].
GARSON, A .
AMERICAN HEART JOURNAL, 1990, 119 (04) :924-941
[9]   Runaway pacemaker: A case report and review [J].
Griffin, J ;
Smithline, H ;
Cook, J .
JOURNAL OF EMERGENCY MEDICINE, 2000, 19 (02) :177-181
[10]  
LAU CP, 1994, AM HEART J, V121, P595