Evaluation of 12 Blood Glucose Monitoring Systems for Self-Testing: System Accuracy and Measurement Reproducibility

被引:43
作者
Freckmann, Guido [1 ]
Baumstark, Annette [1 ]
Schmid, Christina [1 ]
Pleus, Stefan [1 ]
Link, Manuela [1 ]
Haug, Cornelia [1 ]
机构
[1] Univ Ulm, Inst Diabet Technol Forsch & Entwicklungsgesell m, D-89081 Ulm, Germany
关键词
QUALITY;
D O I
10.1089/dia.2013.0208
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Systems for self-monitoring of blood glucose (SMBG) have to provide accurate and reproducible blood glucose (BG) values in order to ensure adequate therapeutic decisions by people with diabetes. Materials and Methods: Twelve SMBG systems were compared in a standardized manner under controlled laboratory conditions: nine systems were available on the German market and were purchased from a local pharmacy, and three systems were obtained from the manufacturer (two systems were available on the U.S. market, and one system was not yet introduced to the German market). System accuracy was evaluated following DIN EN ISO (International Organization for Standardization) 15197:2003. In addition, measurement reproducibility was assessed following a modified TNO (Netherlands Organization for Applied Scientific Research) procedure. Comparison measurements were performed with either the glucose oxidase method (YSI 2300 STAT Plus glucose analyzer; YSI Life Sciences, Yellow Springs, OH) or the hexokinase method (cobas((R)) c111; Roche Diagnostics GmbH, Mannheim, Germany) according to the manufacturer's measurement procedure. Results: The 12 evaluated systems showed between 71.5% and 100% of the measurement results within the required system accuracy limits. Ten systems fulfilled with the evaluated test strip lot minimum accuracy requirements specified by DIN EN ISO 15197:2003. In addition, accuracy limits of the recently published revision ISO 15197:2013 were applied and showed between 54.5% and 100% of the systems' measurement results within the required accuracy limits. Regarding measurement reproducibility, each of the 12 tested systems met the applied performance criteria. Conclusions: In summary, 83% of the systems fulfilled with the evaluated test strip lot minimum system accuracy requirements of DIN EN ISO 15197:2003. Each of the tested systems showed acceptable measurement reproducibility. In order to ensure sufficient measurement quality of each distributed test strip lot, regular evaluations are required.
引用
收藏
页码:113 / 122
页数:10
相关论文
共 23 条
[1]  
[Anonymous], 2003, 151972003 DIN EN ISO
[2]  
[Anonymous], 2008, DTSCH ARZTEBLATT
[3]  
[Anonymous], 1994, 57251 ISO
[4]  
[Anonymous], DIABETES TECHNOL THE
[5]  
[Anonymous], 2013, 1519720132013 ISO
[6]  
Bamberg Richard, 2005, Clin Lab Sci, V18, P203
[7]  
Baumstark Annette, 2012, J Diabetes Sci Technol, V6, P1076
[8]   The role of self-monitoring of blood glucose in the care of people with diabetes: report of a global consensus conference [J].
Bergenstal, RM ;
Gavin, JR .
AMERICAN JOURNAL OF MEDICINE, 2005, 118 :1S-6S
[9]   Current evidence regarding the value of self-monitored blood glucose testing [J].
Blonde, L ;
Karter, AJ .
AMERICAN JOURNAL OF MEDICINE, 2005, 118 :20S-26S
[10]  
Brazg Ronald L, 2013, J Diabetes Sci Technol, V7, P144