Detached multipinhole small animal SPECT device with real-time calibration

被引:13
|
作者
DiFilippo, Frank P.
Riffe, Matthew J.
Harsch, Kelley M.
McCabe, N. Patrick
Heston, Warren D.
机构
[1] Cleveland Clin Fdn, Dept Mol & Funct Imaging, Cleveland, OH 44195 USA
[2] Case Western Reserve Univ, Dept Biomed Engn, Cleveland, OH 44106 USA
[3] Cleveland Clin Fdn, Lerner Res Inst, Cleveland, OH 44195 USA
基金
美国国家卫生研究院;
关键词
calibration; pinhole; single photon emission computed tomography (SPECT); small animal;
D O I
10.1109/TNS.2006.882748
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
High-resolution small animal single photon emission computed tomography (SPECT) with pinhole collimation has become well-established in preclinical research. Conventional scanner designs use custom-fixtures that precisely attach and align pinholes to dedicated detectors. As a lower-cost alternative, we have developed a compact small animal SPECT device with integrated pinholes that is physically detached from gamma camera detectors and uses real-time calibration. The device consists of a shielded lead box with multiple integrated tungsten pinholes on the sides and with a rotating platform at the bottom. The animal or phantom to be imaged is positioned within a hollow cylinder mounted to the platform, with point source markers attached to the cylinder's outer surface. The entire device is placed on the patient table of a standard dual-head gamma camera system with collimators removed. Projection data are acquired with multiple energy windows to separate the point source data from the emission data. The complete geometric calibration is derived through analysis of the point source data. The system matrix is computed based on the calibration parameters and then applied to reconstruction of the emission data. Real-time calibration was found to be robust and accurate, even in the presence of downscatter from Tc-99m emission data or dual-isotope Tc-99m/In-111 emission data. Successful calibration was consistently achieved in all pinhole configurations tried (up to seven pinholes per detector). Accuracy of real-time calibration was demonstrated in phantom studies and animal studies, where structures as small as 0.8 mm could be resolved in reconstructed images. Images appeared to be free of artifacts, and no cross-contamination from the calibration point sources was observed. Real-time calibration is feasible for pinhole SPECT, thus enabling standalone pinhole devices to be used without special alignment fixtures. The calibration accuracy is sufficient for high-resolution small animal studies. The hardware is compatible with nearly all gamma camera models and represents a lower-cost entry-level solution to the field of small animal SPECT.
引用
收藏
页码:2605 / 2612
页数:8
相关论文
共 50 条
  • [21] Real-time self-calibration of a tracked augmented reality display
    Baum, Zachary
    Lasso, Andras
    Ungi, Tamas
    Fichtinger, Gabor
    MEDICAL IMAGING 2016: IMAGE-GUIDED PROCEDURES, ROBOTIC INTERVENTIONS, AND MODELING, 2016, 9786
  • [22] Novel real-time alignment and calibration of the LHCb detector and its performance
    Borghi, S.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2017, 845 : 560 - 564
  • [23] Characterization of Sampled Value Streams in Non Real-Time Calibration Systems
    Agustoni, Marco
    Frigo, Guglielmo
    ENERGIES, 2022, 15 (09)
  • [24] Error characterization and calibration of real-time magnetic field measurement systems
    Grech, Christian
    Amodeo, Maria
    Beaumont, Anthony
    Buzio, Marco
    Di Capua, Vincenzo
    Giloteaux, David
    Sammut, Nicholas
    Wallbank, Joseph Vella
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2021, 990
  • [25] Sensitivity of a real-time freeway crash prediction model to calibration optimality
    Amir Samimi
    Bruce Hellinga
    Samimi, A. (asamimi@sharif.edu), 1600, Springer Verlag (04): : 167 - 174
  • [26] Real-time calibration of space zoom cameras based on fixed stars
    Xia, Qi
    Qi, Naiming
    Ye, Dong
    Guo, Yubo
    Wang, Tianye
    IET COMPUTER VISION, 2017, 11 (05) : 342 - 349
  • [27] Real-Time Water Distribution System Calibration Using Genetic Algorithm
    Cai, Ziyuan
    Dziedzic, Rebecca
    Li, S. Samuel
    PROCEEDINGS OF THE CANADIAN SOCIETY OF CIVIL ENGINEERING ANNUAL CONFERENCE 2022, VOL 1, CSCE 2022, 2023, 363 : 749 - 762
  • [28] Calibration of Visibility Samples for Real-Time Passive Millimeter Wave Imaging
    Zhao, Yan
    Hu, Anyong
    Si, Weikang
    Guo, Xiangzhou
    Miao, Jungang
    IEEE ACCESS, 2021, 9 : 106441 - 106450
  • [29] A Real-Time Rotation Calibration for Interchannel Offset Mismatch in Time-Interleaved SAR ADCs
    Luo, Yixiao
    Liang, Hongzhi
    Peng, Zeyu
    Yu, Yukui
    Liu, Shubin
    Ding, Ruixue
    Zhu, Zhangming
    IEEE TRANSACTIONS ON VERY LARGE SCALE INTEGRATION (VLSI) SYSTEMS, 2025, 33 (03) : 897 - 901
  • [30] Online Calibration Method of DVL Error Based on Real-Time Deformation Compensation
    Gao, Zhaoqiang
    He, Jiazhou
    Li, Xinde
    IEEE SENSORS JOURNAL, 2023, 23 (20) : 25086 - 25095