Comparison of two portable solid state detectors with an improved collimation and alignment device for mammographic x-ray spectroscopy

被引:26
作者
Bottigli, U.
Golosio, B.
Masala, G. L.
Oliva, P.
Stumbo, S. [1 ]
Delogu, P.
Fantacci, M. E.
Abbene, L.
Fauci, F.
Raso, G.
机构
[1] Univ Sassari, Struttura Dipartimento Matemat & Fis, Cagliari, Italy
[2] Ist Nazl Fis Nucl, Cagliari, Italy
[3] Univ Pisa, Dipartimento Fis, Pisa, Italy
[4] Univ Palermo, Dipartimento Fis & Tecnnol Relat, Palermo, Italy
关键词
D O I
10.1118/1.2229431
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We describe a portable system for mammographic x-ray spectroscopy, based on a 2 x 2 x 1 mm(3) cadmium telluride (CdTe) solid state detector, that is greatly improved over a similar system based on a 3 x 3 x 2 mm(3) cadmium zinc telluride (CZT) solid state detector evaluated in an earlier work. The CdTe system utilized new pinhole collimators and an alignment device that facilitated measurement of mammographic x-ray spectra. Marnmographic x-ray spectra acquired by each system were comparable. Half value layer measurements obtained using an ion chamber agreed closely with those derived from the x-ray spectra measured by either detector. The faster electronics and other features of the CdTe detector allowed its use with a larger pinhole collimator than could be used with the CZT detector. Additionally, the improved pinhole collimator and alignment features of the apparatus permitted much more rapid setup for acquisition of x-ray spectra than was possible on the system described in the earlier work. These improvements in detector technology, collimation and ease of alignment, as well as low cost, make this apparatus attractive as a tool for both laboratory research and advanced mammography quality control. (c) 2006 American Association of Physicists in Medicine.
引用
收藏
页码:3469 / 3477
页数:9
相关论文
共 9 条
[1]  
*AMPT INC, US CDTE DIOD DET CZT
[2]   Accurate method for computer-generating tungsten anode x-ray spectra from 30 to 140 kV [J].
Boone, JM ;
Seibert, JA .
MEDICAL PHYSICS, 1997, 24 (11) :1661-1670
[3]  
CRANLEY K, 1997, 78 IEPM I PHYS ENG M
[4]  
Del Sordo S, 2004, NUOVO CIM B, V119, P257, DOI [10.1393/ncb/i2004-10075-o, 10.1393/ncb/i2004-10075-0]
[5]   AN ALGORITHM FOR LEAST-SQUARES ESTIMATION OF NONLINEAR PARAMETERS [J].
MARQUARDT, DW .
JOURNAL OF THE SOCIETY FOR INDUSTRIAL AND APPLIED MATHEMATICS, 1963, 11 (02) :431-441
[6]   CdZnTe detector in mammographic x-ray spectroscopy [J].
Miyajima, S ;
Imagawa, K .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (22) :3959-3972
[7]   CdZnTe detector in diagnostic x-ray spectroscopy [J].
Miyajima, S ;
Imagawa, K ;
Matsumoto, M .
MEDICAL PHYSICS, 2002, 29 (07) :1421-1429
[8]   Direct analysis of molybdenum target generated x-ray spectra with a portable device [J].
Stumbo, S ;
Bottigli, U ;
Golosio, B ;
Oliva, P ;
Tangaro, S .
MEDICAL PHYSICS, 2004, 31 (10) :2763-2770
[9]  
WAGNER LK, 1990, MED PHYS, V17, P989, DOI 10.1118/1.596444