A practical EPR dosimetry system for routine use in radiotherapy: uncertainty analysis of lithium formate dosimeters at the therapeutic dose level

被引:8
作者
Hoefel, Sebastian [1 ,2 ,3 ]
Stehle, Michael [1 ,2 ]
Zwicker, Felix [3 ]
Fix, Michael K. [4 ,5 ,6 ]
Drescher, Malte [1 ,2 ]
机构
[1] Univ Konstanz, Dept Chem, Constance, Germany
[2] Univ Konstanz, Konstanz Res Sch Chem Biol, Constance, Germany
[3] Klinikum Konstanz, Klin & Praxis Strahlentherapie, Constance, Germany
[4] Bern Univ Hosp, Inselspital, Div Med Radiat Phys, Bern, Switzerland
[5] Bern Univ Hosp, Inselspital, Dept Radiat Oncol, Bern, Switzerland
[6] Univ Bern, Bern, Switzerland
关键词
EPR dosimetry; alanine; lithium formate; radiotherapy; uncertainty;
D O I
10.1088/1361-6560/ab9414
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In electron paramagnetic resonance (EPR) dosimetry, solid dosimeter materials such as alanine (AL) or, more recently, lithium formate monohydrate (LFM) are typically used. These materials offer high potential for applications in radiotherapy based on their favorable dosimetric properties. Nevertheless, EPR dosimetry is not widespread in the clinics. This work presents an uncertainty analysis of EPR dosimetry in the dose range from 1 to 70 Gy using a compact spectrometer and applying a practical procedure being suitable for routine use in radiotherapy. The performances of self-pressed LFM pellets and commercial AL pellets are compared side by side. All pellets had a diameter of 4 mm and a height of 2 mm (AL) or 4 mm (LFM). The mean pellet mass was 35.81 mg and 73.81 mg for AL and LFM, respectively. Before irradiation, the pellets were stored for at least 8 weeks at 34 +/- 2% relative humidity. For irradiation, the pellets were put inside an airtight capsule. In total, 25 pellets per material were examined. The pellets were irradiated at a temperature of 25 +/- 2.5 (2 sigma) degrees C to doses of either 1, 5, 20, 50 or 70 Gy (five pellets per dose value and material) by a clinical 6 MV photon beam. Measurement uncertainties were obtained from five independent readouts per pellet within five weeks following irradiation using a benchtop EPR spectrometer. The measurement time of a single readout was restricted to 10 min per pellet. Dose values were derived from EPR signal amplitudes using a specifically developed spectral fitting procedure. Signal fading characteristics were analyzed and taken into account during evaluation. The relative dose uncertainties (1 sigma) for a single readout at doses >= 5 Gy are below 2.8% (AL) and 1.1% (LFM) but increase to 12.3% (AL) and 2.6% (LFM) at 1 Gy. By averaging five independent readouts, the uncertainties at 1 Gy decrease to 2.6% (AL) and 0.8% (LFM). In terms of dose uncertainty, the LFM pellets are superior to the commercial AL pellets owing to their narrower EPR spectrum and approximately doubled mass resulting in higher EPR signal intensities. In case of the LFM pellets, the EPR dosimetry system shows a high level of precision (< 3%) down to 1 Gy being preferable for applications in radiotherapy. The uncertainties can be further decreased by averaging multiple dose values from independent readouts.
引用
收藏
页数:15
相关论文
共 50 条
[41]   Dosimetry of small radiation field in inhomogeneous medium using alanine/EPR minidosimeters and PENELOPE Monte Carlo simulation [J].
Vega Ramirez, J. L. ;
Chen, F. ;
Nicolucci, P. ;
Baffa, O. .
RADIATION MEASUREMENTS, 2011, 46 (09) :941-944
[42]   Electron paramagnetic resonance (EPR) dosimetry using lithium formate in radiotherapy: comparison with thermoluminescence (TL) dosimetry using lithium fluoride rods [J].
Vestad, TA ;
Malinen, E ;
Olsen, DR ;
Hole, EO ;
Sagstuen, E .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (20) :4701-4715
[43]   EPR dosimetric properties of formates [J].
Vestad, TA ;
Malinen, E ;
Lund, A ;
Hole, EO ;
Sagstuen, E .
APPLIED RADIATION AND ISOTOPES, 2003, 59 (2-3) :181-188
[44]   In vivo alanine/electron spin resonance (ESR) dosimetry in radiotherapy of prostate cancer:: A feasibility study [J].
Wagner, Daniela ;
Anton, Mathias ;
Vorwerk, Hilke ;
Gsaenger, Tammo ;
Christiansen, Hans ;
Poppe, Bjoern ;
Hess, Clemens Friedrich ;
Hermann, Robert Michael .
RADIOTHERAPY AND ONCOLOGY, 2008, 88 (01) :140-147
[45]   Feasibility study of entrance and exit dose measurements at the contra lateral breast with alanine/electron spin resonance dosimetry in volumetric modulated radiotherapy of breast cancer [J].
Wagner, Daniela M. ;
Huettenrauch, Petra ;
Anton, Mathias ;
von Voigts-Rhetz, Philip ;
Zink, Klemens ;
Wolff, Hendrik A. .
PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (13) :5462-5472
[46]   Review of the dose-to-water energy dependence of alanine and lithium formate EPR dosimeters and LiF TL-dosimeters - Comparison with Monte Carlo simulations [J].
Waldeland, Einar ;
Malinen, Eirik .
RADIATION MEASUREMENTS, 2011, 46 (09) :945-951
[47]   Characterization of lithium formate EPR dosimeters for high dose applications - Comparison with alanine [J].
Waldeland, Einar ;
Helt-Hansen, Jakob ;
Malinen, Eirik .
RADIATION MEASUREMENTS, 2011, 46 (02) :213-218
[48]   Dosimetry of stereotactic radiosurgery using lithium formate EPR dosimeters [J].
Waldeland, Einar ;
Horling, Magnus ;
Hole, Eli Olaug ;
Sagstuen, Einar ;
Malinen, Eirik .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (08) :2307-2316
[49]   THE INFLUENCE OF NON-RADIATION INDUCED ESR BACKGROUND SIGNAL FROM PARAFFIN-ALANINE PROBES FOR DOSIMETRY IN THE RADIOTHERAPY DOSE RANGE [J].
WIESER, A ;
LETTAU, C ;
FILL, U ;
REGULLA, DF .
APPLIED RADIATION AND ISOTOPES, 1993, 44 (1-2) :59-65
[50]  
WIESER A, 1989, APPL RADIAT ISOTOPES, V40, P911, DOI 10.1016/0883-2889(89)90016-6