Novel endoscopic over-the-scope clip system

被引:32
作者
Armellini, Elia [1 ]
Crino, Stefano Francesco [1 ]
Orsello, Marco [1 ]
Ballare, Marco [1 ]
Tari, Roberto [1 ]
Saettone, Silvia [1 ]
Montino, Franco [1 ]
Occhipinti, Pietro [1 ]
机构
[1] Azienda Ospedaliero Univ Maggiore della Carita, Div Gastroenterol, I-28100 Novara, Italy
关键词
Therapeutic endoscopy; Over-the-scope clip; Non variceal gastrointestinal bleeding; Endoscopic hemostasis; Respiratory-oesophageal fistula; Fistula closure; PADLOCK-G CLIP; MANAGEMENT; SURVIVAL; PORCINE; CLOSURE;
D O I
10.3748/wjg.v21.i48.13587
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
This paper reports our experience with a new over-the-scope clip in the setting of recurrent bleeding and oesophageal fistula. We treated five patients with the over-the-scope Padlock Clip (TM). It is a nitinol ring, with six inner needles preassembled on an applicator cap, thumb press displaced by the Lock-It (TM) delivery system. The trigger wire is located alongside the shaft of the endoscope, and does not require the working channel. Three patients had recurrent bleeding lesions (bleeding rectal ulcer, post polypectomy delayed bleeding and duodenal Dieulafoy's lesion) and two patients had a persistent respiratory-esophageal fistula. In all patients a previous endoscopic attempt with standard techniques had been useless. All procedures were conducted under conscious sedation but for one patient that required general anaesthesia due to multiple comorbidities. We used one Padlock Clip (TM) for each patient in a single session. Simple suction was enough in all of our patients to obtain tissue adhesion to the instrument tip. A remarkably short application time was recorded for all cases (mean duration of the procedure: 8 min). We obtained technical and immediate clinical success for every patient. No major immediate, early or late (within 24 h, 7 d or 4 wk) adverse events were observed, over follow-up durations lasting a mean of 109.4 d. One patient, treated for duodenal bulb bleeding from a Dieulafoy's lesion, developed signs of mild pancreatitis 24 h after the procedure. The new over-the-scope Padlock Clip (TM) seems to be simple to use and effective in different clinical settings, particularly in "difficult" scenarios, like recurrent bleeding and respiratory-oesophageal fistulas.
引用
收藏
页码:13587 / 13592
页数:6
相关论文
共 12 条
[1]   Use of the Over-The-Scope Clip for treatment of refractory upper gastrointestinal bleeding: a case series [J].
Chan, Shannon M. ;
Chiu, Philip W. Y. ;
Teoh, Anthony Y. B. ;
Lau, James Y. W. .
ENDOSCOPY, 2014, 46 (05) :428-431
[2]   Gastrotomy Closure with the Lock-It System and the Padlock-G clip: A Survival Study in a Porcine Model [J].
Desilets, David J. ;
Romanelli, John R. ;
Earle, David B. ;
Chapman, Christopher N. .
JOURNAL OF LAPAROENDOSCOPIC & ADVANCED SURGICAL TECHNIQUES, 2010, 20 (08) :671-676
[3]   Yes, we can: reliable colonic closure with the Padlock-G clip in a survival porcine study (with video) [J].
Guarner-Argente, Carlos ;
Cordova, Henry ;
Martinez-Palli, Graciela ;
Navarro, Ricard ;
Cuatrecasas, Miriam ;
Rodriguez de Miguel, Cristina ;
Beltran, Mireia ;
Lacy, Antonio M. ;
Gines, Angels ;
Pellise, Maria ;
Llach, Josep ;
Fernandez-Esparrach, Gloria .
GASTROINTESTINAL ENDOSCOPY, 2010, 72 (04) :841-844
[4]   International multicenter experience with an over-the-scope clipping device for endoscopic management of GI defects (with video) [J].
Haito-Chavez, Yamile ;
Law, Joanna K. ;
Kratt, Thomas ;
Arezzo, Alberto ;
Verra, Mauro ;
Morino, Mario ;
Sharaiha, Reem Z. ;
Poley, Jan-Werner ;
Kahaleh, Michel ;
Thompson, Christopher C. ;
Ryan, Michele B. ;
Choksi, Neel ;
Elmunzer, B. Joseph ;
Gosain, Sonia ;
Goldberg, Eric M. ;
Modayil, Rani J. ;
Stavropoulos, Stavros N. ;
Schembre, Drew B. ;
DiMaio, Christopher J. ;
Chandrasekhara, Vinay ;
Hasan, Muhammad K. ;
Varadarajulu, Shyam ;
Hawes, Robert ;
Gomez, Victoria ;
Woodward, Timothy A. ;
Rubel-Cohen, Sergio ;
Fluxa, Fernando ;
Vleggaar, Frank P. ;
Akshintala, Venkata S. ;
Raju, Gottumukkala S. ;
Khashab, Mouen A. .
GASTROINTESTINAL ENDOSCOPY, 2014, 80 (04) :610-622
[5]   The role of endoscopy in the management of acute non-variceal upper GI bleeding [J].
Hwang, Joo Ha ;
Fisher, Deborah A. ;
Ben-Menachem, Tamir ;
Chandrasekhara, Vinay ;
Chathadi, Krishnavel ;
Decker, G. Anton ;
Early, Dayna S. ;
Evans, John A. ;
Fanelli, Robert D. ;
Foley, Kimberly ;
Fukami, Norio ;
Jain, Rajeev ;
Jue, Terry L. ;
Khan, Kahlid M. ;
Lightdale, Jenifer ;
Malpas, Phyllis M. ;
Maple, John T. ;
Pasha, Shabana ;
Saltzman, John ;
Sharaf, Ravi ;
Shergill, Amandeep K. ;
Dominitz, Jason A. ;
Cash, Brooks D. .
GASTROINTESTINAL ENDOSCOPY, 2012, 75 (06) :1132-1138
[6]   Endoscopic resection of hilar papillomatosis after Whipple procedure for ampullary adenoma [J].
Law, Ryan ;
Topazian, Mark ;
Baron, Todd H. .
GASTROINTESTINAL ENDOSCOPY, 2013, 78 (02) :226-226
[7]   Over-the-scope clip (OTSC) represents an effective endoscopic treatment for acute GI bleeding after failure of conventional techniques [J].
Manta, Raffaele ;
Galloro, Giuseppe ;
Mangiavillano, Benedetto ;
Conigliaro, Rita ;
Pasquale, Luigi ;
Arezzo, Alberto ;
Masci, Enzo ;
Bassotti, Gabrio ;
Frazzoni, Marzio .
SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2013, 27 (09) :3162-3164
[8]  
Merchea A, 2010, SURGERY, V148, P876, DOI 10.1016/j.surg.2010.07.010
[9]   Complications of Gastric Endoscopic Submucosal Dissection [J].
Oda, Ichiro ;
Suzuki, Haruhisa ;
Nonaka, Satoru ;
Yoshinaga, Shigetaka .
DIGESTIVE ENDOSCOPY, 2013, 25 :71-78
[10]  
Reed Michael F, 2003, Chest Surg Clin N Am, V13, P271, DOI 10.1016/S1052-3359(03)00030-9