A continuous surface reconstruction method on point cloud captured from a 3D surface photogrammetry system

被引:9
作者
Liu, Wenyang [1 ]
Cheung, Yam [2 ]
Sabouri, Pouya [2 ]
Arai, Tatsuya J. [2 ]
Sawant, Amit [2 ]
Ruan, Dan [1 ,3 ]
机构
[1] Univ Calif Los Angeles, Dept Bioengn, Los Angeles, CA 90095 USA
[2] Univ Texas Southwestern, Dept Radiat Oncol, Dallas, TX 75390 USA
[3] Univ Calif Los Angeles, Dept Radiat Oncol, Los Angeles, CA 90095 USA
关键词
surface reconstruction; implicit representation; photogrammetry; variational formulation; LEVEL SET METHOD; RADIATION-THERAPY;
D O I
10.1118/1.4933196
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To accurately and efficiently reconstruct a continuous surface from noisy point clouds captured by a surface photogrammetry system (VisionRT). Methods: The authors have developed a level-set based surface reconstruction method on point clouds captured by a surface photogrammetry system (VisionRT). The proposed method reconstructs an implicit and continuous representation of the underlying patient surface by optimizing a regularized fitting energy, offering extra robustness to noise and missing measurements. By contrast to explicit/discrete meshing-type schemes, their continuous representation is particularly advantageous for subsequent surface registration and motion tracking by eliminating the need for maintaining explicit point correspondences as in discrete models. The authors solve the proposed method with an efficient narrowband evolving scheme. The authors evaluated the proposed method on both phantom and human subject data with two sets of complementary experiments. In the first set of experiment, the authors generated a series of surfaces each with different black patches placed on one chest phantom. The resulting VisionRT measurements from the patched area had different degree of noise and missing levels, since VisionRT has difficulties in detecting dark surfaces. The authors applied the proposed method to point clouds acquired under these different configurations, and quantitatively evaluated reconstructed surfaces by comparing against a high-quality reference surface with respect to root mean squared error (RMSE). In the second set of experiment, the authors applied their method to 100 clinical point clouds acquired from one human subject. In the absence of ground-truth, the authors qualitatively validated reconstructed surfaces by comparing the local geometry, specifically mean curvature distributions, against that of the surface extracted from a high-quality CT obtained from the same patient. Results: On phantom point clouds, their method achieved submillimeter reconstruction RMSE under different configurations, demonstrating quantitatively the faith of the proposed method in preserving local structural properties of the underlying surface in the presence of noise and missing measurements, and its robustness toward variations of such characteristics. On point clouds from the human subject, the proposed method successfully reconstructed all patient surfaces, filling regions where raw point coordinate readings were missing. Within two comparable regions of interest in the chest area, similar mean curvature distributions were acquired from both their reconstructed surface and CT surface, with mean and standard deviation of (mu(recon) = -2.7x10(-3) mm(-1), sigma(recon) = 7.0x10(-3) mm(-1)) and (mu(CT) = -2.5x10(-3) mm(-1), sigma(CT) = 5.3x10(-3) mm(-1)), respectively. The agreement of local geometry properties between the reconstructed surfaces and the CT surface demonstrated the ability of the proposed method in faithfully representing the underlying patient surface. Conclusions: The authors have integrated and developed an accurate level-set based continuous surface reconstruction method on point clouds acquired by a 3D surface photogrammetry system. The proposed method has generated a continuous representation of the underlying phantom and patient surfaces with good robustness against noise and missing measurements. It serves as an important first step for further development of motion tracking methods during radiotherapy. (C) 2015 American Association of Physicists in Medicine.
引用
收藏
页码:6564 / 6571
页数:8
相关论文
共 22 条
[1]   A FAST LEVEL SET METHOD FOR PROPAGATING INTERFACES [J].
ADALSTEINSSON, D ;
SETHIAN, JA .
JOURNAL OF COMPUTATIONAL PHYSICS, 1995, 118 (02) :269-277
[2]   Surface reconstruction by Voronoi filtering [J].
Amenta, N ;
Bern, M .
DISCRETE & COMPUTATIONAL GEOMETRY, 1999, 22 (04) :481-504
[3]  
Bajaj Chandrajit., 1997, Introduction to implicit surfaces
[4]   3-DIMENSIONAL ALPHA-SHAPES [J].
EDELSBRUNNER, H ;
MUCKE, EP .
ACM TRANSACTIONS ON GRAPHICS, 1994, 13 (01) :43-72
[5]   Validation of the use of photogrammetry to register preprocedure MR images to intra-procedure patient position for image-guided cardiac catheterization procedures [J].
Gao, Gang ;
Tarte, Segolene ;
King, Andy ;
Ma, Yingliang ;
Chinchapatnam, Phani ;
Schaeffter, Tobias ;
Razavi, Reza ;
Hawkes, Dave ;
Hill, Derek ;
Rhode, Kawal .
MEDICAL IMAGING 2008: VISUALIZATION, IMAGE-GUIDED PROCEDURES, AND MODELING, PTS 1 AND 2, 2008, 6918
[6]   Evaluation of the Accuracy of a 3D Surface Imaging System for Patient Setup in Head and Neck Cancer Radiotherapy [J].
Gopan, Olga ;
Wu, Qiuwen .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 84 (02) :547-552
[7]   Clinical use of electronic portal imaging: Report of AAPM radiation therapy committee Task Group 58 [J].
Herman, MG ;
Balter, JM ;
Jaffray, DA ;
McGee, KP ;
Munro, P ;
Shalev, S ;
Van Herk, M ;
Wong, JW .
MEDICAL PHYSICS, 2001, 28 (05) :712-737
[8]  
Lefohn AE, 2003, LECT NOTES COMPUT SC, V2878, P564
[9]   Cone-beam-CT guided radiation therapy:: technical implementation [J].
Létourneau, D ;
Wong, JW ;
Oldham, M ;
Gulam, M ;
Watt, L ;
Jaffray, DA ;
Siewerdsen, JH ;
Martinez, AA .
RADIOTHERAPY AND ONCOLOGY, 2005, 75 (03) :279-286
[10]   A unified variational segmentation framework with a level-set based sparse composite shape prior [J].
Liu, Wenyang ;
Ruan, Dan .
PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (05) :1865-1877