Double-range near-infrared acetylene detection using a dual spot-ring Herriott cell (DSR-HC)

被引:64
作者
Dong, Ming [1 ]
Zheng, Chuantao [1 ,2 ]
Yao, Dan [1 ]
Zhong, Guoqiang [1 ]
Miao, Shuzhuo [1 ]
Ye, Weilin [2 ,3 ]
Wang, Yiding [1 ]
Tittel, Frank K. [2 ]
机构
[1] Jilin Univ, Coll Elect Sci & Engn, State Key Lab Integrated Optoelect, 2699 Qianjin St, Changchun 130012, Jilin, Peoples R China
[2] Rice Univ, Dept Elect & Comp Engn, 6100 Main St, Houston, TX 77005 USA
[3] Shantou Univ, Coll Engn, 243 Daxue Rd, Shantou 515063, Peoples R China
基金
美国国家科学基金会; 中国国家自然科学基金;
关键词
LASER-ABSORPTION-SPECTROSCOPY; INTERBAND CASCADE LASER; MIDINFRARED METHANE SENSOR; MU-M; GAS; SPECTROMETER; LINES;
D O I
10.1364/OE.26.012081
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Design and fabrication of a dual spot-ring Herriott cell (DSR-HC) were proposed. The sealed Herriott cell with a dimensional size of 5.5 cm x 9.2 cm x 32.1 cm, possessed two input/output coupling holes leading to two absorption path lengths of similar to 20 m and similar to 6 m, respectively. An acetylene (C2H2) sensor system with a double-range was developed using the DSR-HC and wavelength modulation spectroscopy (WMS) technique. A near-infrared distributed feedback (DFB) laser was employed for targeting a C2H2 absorption line at 6521.2 cm(-1). C2H2 concentration measurements were carried out by modulating the laser at a 5 kHz frequency and demodulating the detector signal with LabVIEW software. An Allan-Werle deviation analysis indicated that the limit of detection (LoD) for the two absorption path lengths of 20 m and 6 m are 7.9 parts-per-million in volume (ppmv) and 4.0 ppmv, respectively. The DSR-HC concept can be used to fabricate similar cells for single-gas detection requiring two different detection ranges as well as for dual-gas detection requiring different absorption path lengths. (C) 2018 Optical Society of America under the terms of the OSA Open Access Publishing Agreement
引用
收藏
页码:12081 / 12091
页数:11
相关论文
共 25 条
[1]   LASER RAMAN SENSOR FOR MEASUREMENT OF TRACE-HYDROGEN GAS [J].
ADLERGOLDEN, SM ;
GOLDSTEIN, N ;
BIEN, F ;
MATTHEW, MW ;
GERSH, ME ;
CHENG, WK ;
ADAMS, FW .
APPLIED OPTICS, 1992, 31 (06) :831-835
[2]   2-MIRROR MULTIPASS ABSORPTION CELL [J].
ALTMANN, J ;
BAUMGART, R ;
WEITKAMP, C .
APPLIED OPTICS, 1981, 20 (06) :995-999
[3]   Acetylene detection based on diode laser QEPAS: combined wavelength and residual amplitude modulation [J].
Cao, Y. ;
Jin, W. ;
Ho, H. L. ;
Qi, L. ;
Yang, Y. H. .
APPLIED PHYSICS B-LASERS AND OPTICS, 2012, 109 (02) :359-366
[4]   Precise multipass Herriott cell design: Derivation of controlling design equations [J].
Engel, Gregory S. ;
Moyer, Elisabeth J. .
OPTICS LETTERS, 2007, 32 (06) :704-706
[5]   ICL-based TDLAS sensor for real-time breath gas analysis of carbon monoxide isotopes [J].
Ghorbani, Ramin ;
Schmidt, Florian M. .
OPTICS EXPRESS, 2017, 25 (11) :12743-12752
[6]   Repetitively Mode-Locked Cavity-Enhanced Absorption Spectroscopy (RML-CEAS) for Near-Infrared Gas Sensing [J].
He, Qixin ;
Lou, Minhan ;
Zheng, Chuantao ;
Ye, Weilin ;
Wang, Yiding ;
Tittel, Frank K. .
SENSORS, 2017, 17 (12)
[7]   OFF-AXIS PATHS IN SPHERICAL MIRROR INTERFEROMETERS [J].
HERRIOTT, D ;
KOMPFNER, R ;
KOGELNIK, H .
APPLIED OPTICS, 1964, 3 (04) :523-&
[8]   FOLDED OPTICAL DELAY LINES [J].
HERRIOTT, DR ;
SCHULTE, HJ .
APPLIED OPTICS, 1965, 4 (08) :883-&
[9]   Detection of acetylene impurities in ethylene and polyethylene manufacturing processes using tunable diode laser spectroscopy in the 3-μm range [J].
Kluczynski, P. ;
Jahjah, M. ;
Naehle, L. ;
Axner, O. ;
Belahsene, S. ;
Fischer, M. ;
Koeth, J. ;
Rouillard, Y. ;
Westberg, J. ;
Vicet, A. ;
Lundqvist, S. .
APPLIED PHYSICS B-LASERS AND OPTICS, 2011, 105 (02) :427-434
[10]   Compact TDLAS based optical sensor for ppb-level ethane detection by use of a 3.34 μm room-temperature CW interband cascade laser [J].
Li, Chunguang ;
Dong, Lei ;
Zheng, Chuantao ;
Tittel, Frank K. .
SENSORS AND ACTUATORS B-CHEMICAL, 2016, 232 :188-194