Vagus nerve stimulation therapy system:: In vitro evaluation of magnetic resonance imaging-related heating and function at 1.5 and 3 tesla

被引:32
作者
Shellock, Frank G.
Begnaud, Jason
Inman, D. Michael
机构
[1] Inst Magnet Resonance Safety Educ & Res, Los Angeles, CA 90045 USA
[2] Univ So Calif, Keck Sch Med, Los Angeles, CA USA
[3] Cyberonics Inc, Houston, TX USA
来源
NEUROMODULATION | 2006年 / 9卷 / 03期
关键词
magnetic resonance imaging; safety; vagus nerve stimulation;
D O I
10.1111/j.1525-1403.2006.00061.x
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Objectives. To evaluate magnetic resonance imaging-related (MRI-related) heating for the VNS Therapy System at 1.5 and 3 tesla (T) using various device configurations and MRI conditions and to assess device function before and after MRI. Methods. The VNS Therapy System (pulse generator, Model 102; leads Models 300 and 302; Cyberonics, Inc., Houston, Tex, USA) underwent assessment of MRI-related heating at 1.5 and 3 T using different positioning configurations, leads, transmit radiofrequency (RF) coils (body and head), RF power levels, and scans on different body regions. The function of the VNS Therapy System was evaluated before and after scanning. Results. At 1.5 T using a transmit RF body coil, excessive temperature changes were associated with scans of the C-spine/shoulder (+11.5 degrees C, complete system; +29.5 degrees C, lead without pulse generator). The lowest temperature change occurred for the scan of the L-spine. At 1.5 T using a transmit/receive RF head coil, temperature changes did not exceed +0.2 degrees C under the conditions studied. At 3 T using a transmit RF body coil, the highest temperature change occurred with the scan of the C-spine/shoulder (+14.5 degrees C) with the lead configured with no strain relief loops at the vagus nerve. MRI performed using various conditions at 1.5 and 3 T produced no significant alterations in the function of the VNS Therapy System. Conclusions. MRI-related heating was characterized for a variety of scenarios, identifying unsafe as well as safe conditions. Device function was unaffected by MRI procedures at 1.5 and 3 T. By following specific conditions, safety guidelines for the VNS Therapy System may be expanded beyond those currently indicated by the manufacturer.
引用
收藏
页码:204 / 213
页数:10
相关论文
共 17 条
[1]  
*ASTM, 2002, F218202 ASTM
[2]   MRI of the brain is safe in patients implanted with the vagus nerve stimulator [J].
Benbadis, SR ;
Nyhenhuis, J ;
Tatum, WO ;
Murtagh, FR ;
Gieron, M ;
Vale, FL .
SEIZURE-EUROPEAN JOURNAL OF EPILEPSY, 2001, 10 (07) :512-515
[3]   Bilateral neuro stimulation systems used for deep brain stimulation: in vitro study of MRI-related heating at 1.5 T and implications for clinical imaging of the brain [J].
Bhidayasiri, R ;
Bronstein, JM ;
Sinha, S ;
Krahl, SE ;
Ahn, S ;
Behnke, EJ ;
Cohen, MS ;
Frysinger, R ;
Shellock, FG .
MAGNETIC RESONANCE IMAGING, 2005, 23 (04) :549-555
[4]  
COSMAN ER, 1992, TXB STEREOTACTIC FUN, P973
[5]  
*CYB INC, 2003, PHYS MAN VNS THER PU
[6]   Active deep brain stimulation during MRI: A feasibility study [J].
Georgi, JC ;
Stippich, C ;
Tronnier, VM ;
Heiland, S .
MAGNETIC RESONANCE IN MEDICINE, 2004, 51 (02) :380-388
[7]   Vagal nerve stimulation: a review of its applications and potential mechanisms that mediate its clinical effects [J].
Groves, DA ;
Brown, VJ .
NEUROSCIENCE AND BIOBEHAVIORAL REVIEWS, 2005, 29 (03) :493-500
[8]   Prevention of mechanical failures in implanted spinal cord stimulation systems [J].
Henderson, JM ;
Oakley, JC ;
Schade, CM ;
Sasaki, J ;
Caraway, DL .
NEUROSURGERY, 2005, 57 (02) :414-414
[9]   Beyond the treatment of epilepsy: New applications of vagus nerve stimulation in psychiatry [J].
Kosel, M ;
Schlaepfer, TE .
CNS SPECTRUMS, 2003, 8 (07) :515-521
[10]   Vagus nerve stimulation (VNS) synchronized BOLD fMRI suggests that VNS in depressed adults has frequency/dose dependent effects [J].
Lomarev, M ;
Denslow, S ;
Nahas, Z ;
Chae, JH ;
George, MS ;
Bohning, DE .
JOURNAL OF PSYCHIATRIC RESEARCH, 2002, 36 (04) :219-227