Performance of a High Speed and High Density Data Acquisition System for Multiple Gamma-ray Detection

被引:0
|
作者
Kimura, Atsushi [1 ]
Toh, Yosuke [1 ]
Koizumi, Mitsuo [1 ]
Furutaka, Kazuyoshi [1 ]
Kin, Tadahiro [1 ]
Oshima, Masumi [1 ]
机构
[1] Japan Atom Energy Agcy, Res Grp Innovat Nucl Sci, Tokai, Ibaraki, Japan
关键词
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To accelerate the data acquisition system (DAQ), which was reported in NSS 2004[1], the authors improved the Main ADC Modules in the DAQ. The new Main ADC Module has 8 input channels. Each channel consists of analog circuits for pulse shaping, a high-speed ADC (14bit 125MSPS), and a high-power Digital Signal Processor (DSP: 8000MIPS). Shaped input signals are digitized with the ADC and analyzed by the DSP to calculate the energy data. Total energy resolution of this system was 2.5 keV in FWHM for 1.33-MeV gamma rays. Time resolution of this system was 120 ns in FWHM between two Ge detectors. These results show good performances as a conventional DAQ. The dead time of this system is only 4.3 us per event and the maximum event rate was more than 350k events/s. Thus, enough energy and time resolutions are achieved with short dead time, high channel density and high event rate.
引用
收藏
页码:1382 / 1386
页数:5
相关论文
共 50 条
  • [21] High Speed Gamma-Ray Tomography for Hydrocarbon Flow Applications
    Hjertaker, Bjorn Tore
    Johansen, Geir Anton
    CT 2008: TOMOGRAPHY CONFLUENCE, 2008, 1050 : 163 - 174
  • [22] REAL-TIME DATA ACQUISITION-SYSTEM FOR VERY-HIGH-ENERGY GAMMA-RAY ASTRONOMY EXPERIMENT
    BHAT, PN
    JOHN, AV
    UPADHYAYA, SS
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1990, 292 (02): : 494 - 504
  • [23] AN EXPERIMENTAL SYSTEM FOR DETECTION OF HIGH-ENERGY COSMIC GAMMA-RAY BURSTS
    GOPALAKRISHNAN, NV
    GUPTA, SK
    RAMANAMURTHY, PV
    TONWAR, SC
    NUCLEAR INSTRUMENTS & METHODS, 1980, 175 (2-3): : 555 - 559
  • [24] Design of high speed data acquisition system
    Hou, Xiaomin
    Wan, Yuanqin
    Li, Gang
    Ou, Hongwu
    Shu Ju Cai Ji Yu Chu Li/Journal of Data Acquisition & Processing, 2000, 15 (02): : 245 - 247
  • [25] Development of a prompt gamma-ray analysis combined with multiple gamma-ray detection
    Toh, Y
    Oshima, A
    Koizumi, A
    Osa, A
    Kimura, A
    CAPTURE GAMMA-RAY SPECTROSCOPY AND RELATED TOPICS, 2006, 819 : 603 - +
  • [26] ROLLING FRICTION OF HIGH DENSITY POLYETHYLENE IRRADIATED BY GAMMA-RAY
    WATANABE, M
    MATSUBAR.K
    JOURNAL OF JAPAN SOCIETY OF LUBRICATION ENGINEERS, 1970, 15 (02): : 99 - &
  • [27] A High Performance Solid-state Storage System for High Speed Data Acquisition Applications
    Yu, Xujin
    Wang, Xuguang
    Yan, Wei
    2014 IEEE INTERNATIONAL CONFERENCE ON ELECTRON DEVICES AND SOLID-STATE CIRCUITS (EDSSC), 2014,
  • [28] Multiple scattering correction for high Gamma-Ray compton experiments
    Hamouda, S. A.
    Alabyad, A. M.
    2ND INTERNATIONAL CONFERENCE ON MATERIALS PHYSICS AND APPLICATIONS (JIPMA 2009/MATERIAUX 2009), 2010, 13
  • [29] Development of data acquisition and control software for gamma-ray process tomography system
    Yussup, N.
    Rahman, N. A. A.
    Ibrahim, M. M.
    Soh-Shaari, S. C.
    Saad, N.
    Hassan, H.
    Abdullah, J.
    INTERNATIONAL NUCLEAR SCIENCE TECHNOLOGY AND ENGINEERING CONFERENCE 2019 (INUSTEC2019), 2020, 785
  • [30] A HIGH RESOLUTION (N,N'GAMMA) GAMMA-RAY SPECTROMETER SYSTEM
    WILLIAMS, GH
    TUCKER, WE
    ASHE, JB
    BULLETIN OF THE AMERICAN PHYSICAL SOCIETY, 1968, 13 (06): : 873 - &