Impaction grafting with morsellised allograft and tricalcium phosphate-hydroxyapatite: incorporation within ovine metaphyseal bone defects

被引:22
作者
Pratt, JNJ
Griffon, DJ
Dunlop, DG
Smith, N
Howie, CR
机构
[1] Univ Edinburgh, Royal Dick Sch Vet Studies, Easter Bush Vet Ctr, Roslin EH42 9RG, Midlothian, Scotland
[2] Princess Margaret Rose Orthopaed Hosp, Edinburgh EH10 7ED, Midlothian, Scotland
[3] Pfizer Hosp Prod Ltd, Howmedica, Staines TW18 4AN, Middx, England
关键词
impaction grafting; revision hip arthroplasty; tricalcium phosphate; hydroxyapatite;
D O I
10.1016/S0142-9612(02)00018-2
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
An ovine model was used to investigate the in vivo properties of impacted tricalcium phosphate-hydroxyapatitc (TCP-HA) aggregates, varying in chemical composition (ratio of TCP to HA) and particle size distribution (8 versus 3 particle size ranges). All pellets were impacted to a standard compactive effort. Eight sheep underwent implantation of pellets in 4 metaphyseal defects in both rear limbs. Treatment groups consisted of: (1) allograft (clinical control), (2) 50/50 allograft/80%, HA/20%,, TCP in 8 particle size ranges, (3) 50/50 allograft/80% TCP/20% HA in 8 sizes and (4) 50/50 allograft/80% HA/20% TCP in only 3 sizes of particles. Healing of defects was evaluated at 14 weeks with computed tomography, histology and histomorphometry. The computer tomography (CT) density measured in all defects containing synthetic agents was higher than in defects filled with allograft alone (p<0.01). Defects containing 8 sizes of 80% HA/ 20% TCP granules (group 2) achieved lower histological scores and contained less bone than the clinical control (p<0.05), whereas groups 3 and 4 did not differ from the control. Although all synthetic agents were osteoconductive, our results suggest that increasing the ratio of TCP over HA and limiting the number of particle size ranges to 3 instead of 8 improve the performance of impacted aggregates Lis graft expanders. Evaluation under loading conditions of morsellised allograft expanded with 80% TCP/20% HA (BoneSave") in 3 particle size ranges is warranted. (C) 2002 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:3309 / 3317
页数:9
相关论文
共 36 条
[1]   Bone grafts and bone substitutes in hip and knee surgery [J].
Behairy, Y ;
Jasty, M .
ORTHOPEDIC CLINICS OF NORTH AMERICA, 1999, 30 (04) :661-+
[2]  
BLOM AW, 2001, J BONE JOINT SURG RH, P96
[3]   Mechanical considerations in impaction bone grafting [J].
Brewster, NT ;
Brewster, WJ ;
Howie, CR ;
Madabhushi, SPG ;
Usmani, AS ;
Fairbairn, DR .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1999, 81B (01) :118-124
[4]  
BURWELL RG, 1994, BONE GRAFTS DERIVATI, P5
[5]   Osteoconductive materials and their role as substitutes for autogenous bone grafts [J].
Cornell, CN .
ORTHOPEDIC CLINICS OF NORTH AMERICA, 1999, 30 (04) :591-+
[6]  
Craig R. F., 1993, SOIL MECH
[7]   Massive early subsidence following femoral impaction grafting [J].
Eldridge, JDJ ;
Smith, EJ ;
Hubble, MJ ;
Whitehouse, SL ;
Learmonth, ID .
JOURNAL OF ARTHROPLASTY, 1997, 12 (05) :535-540
[8]   OSSEOINTEGRATION OF MACROPOROUS CALCIUM-PHOSPHATE CERAMICS HAVING A DIFFERENT CHEMICAL-COMPOSITION [J].
FRAYSSINET, P ;
TROUILLET, JL ;
ROUQUET, N ;
AZIMUS, E ;
AUTEFAGE, A .
BIOMATERIALS, 1993, 14 (06) :423-429
[9]   Supply and demand of bone allograft for revision hip surgery in Scotland [J].
Galea, G ;
Kopman, D ;
Graham, BJM .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1998, 80B (04) :595-599
[10]   IMPACTED CANCELLOUS ALLOGRAFTS AND CEMENT FOR REVISION TOTAL HIP-ARTHROPLASTY [J].
GIE, GA ;
LINDER, L ;
LING, RSM ;
SIMON, IP ;
SLOOFF, TJJH ;
TIMPERLEY, AJ .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1993, 75 (01) :14-21