Does the Implantation Technique for Totally Implantable Venous Access Ports (TIVAPs) Influence Long-Term Outcome?

被引:10
作者
Biacchi, Daniele [1 ]
Sammartino, Paolo [1 ]
Sibio, Simone [1 ]
Accarpio, Fabio [1 ]
Cardi, Maurizio [1 ]
Sapienza, Paolo [1 ]
De Cesare, Alessandro [1 ]
Atta, Joseph Maher Fouad [1 ]
Impagnatiello, Alessio [1 ]
Di Giorgio, Angelo [1 ]
机构
[1] Univ Roma La Sapienza, Dept Surg Pietro Valdoni, I-00161 Rome, Italy
关键词
PINCH-OFF SYNDROME; COMPLICATIONS; DEVICES; CHEMOTHERAPY; COSTS;
D O I
10.1007/s00268-015-3233-z
中图分类号
R61 [外科手术学];
学科分类号
摘要
Totally implantable venous access ports (TIVAP) are eventually explanted for various reasons, related or unrelated to the implantation technique used. Having more information on long-term explantation would help improve placement techniques. From a series of 1572 cancer patients who had TIVAPs implanted in our center with the cutdown technique or Seldinger technique, we studied the 542 patients who returned to us to have their TIVAP explanted after 70 days or more. As outcome measures we distinguished between TIVAPs explanted for long-term complications (infection, catheter-, reservoir-, and patient-related complications) and TIVAPs no longer needed. Univariate and multivariate analyses were run to investigate the reasons for explantation and their possible correlation with implantation techniques. The most common reason for explantation was infection (47.6 %), followed by catheter-related (20.8 %), patient-related (14.7 %), and reservoir-related complications (4.7 %). In the remaining 12.2 % of cases, the TIVAP was explanted complication free after the planned treatments ended. Infection correlated closely with longer TIVAP use. Univariate and multivariate analyses identified the Seldinger technique as a major risk factor for venous thrombosis and catheter dislocation. The need for long-term TIVAP explantation in about one-third of cancer patients is related to the implantation techniques used.
引用
收藏
页码:284 / 290
页数:7
相关论文
共 18 条
[1]   Totally implantable venous access ports: Retrospective review of long-term complications in 81 patients [J].
Bassi, K. K. ;
Giri, A. K. ;
Pattanayak, M. ;
Abraham, S. W. ;
Pandey, K. K. .
INDIAN JOURNAL OF CANCER, 2012, 49 (01) :114-118
[2]   Totally implantable central venous access ports for long-term chemotherapy - A prospective study analyzing complications and costs of 333 devices with a minimum follow-up of 180 days [J].
Biffi, R ;
de Braud, E ;
Orsi, E ;
Pozzi, S ;
Mauri, S ;
Goldhirsch, A ;
Nole, F ;
Andreoni, B .
ANNALS OF ONCOLOGY, 1998, 9 (07) :767-773
[3]   Totally implantable venous access ports systems for patients receiving chemotherapy for solid tissue malignancies: A randomized controlled clinical trial examining the safety, efficacy, costs, and impact on quality of life [J].
Bow, EJ ;
Kilpatrick, MG ;
Clinch, JJ .
JOURNAL OF CLINICAL ONCOLOGY, 1999, 17 (04) :1267-1273
[4]   Ultrasound-guided central venous catheterization in cancer patients improves the success rate of cannulation and reduces mechanical complications: A prospective observational study of 1,978 consecutive catheterizations [J].
Cavanna, Luigi ;
Civardi, Giuseppe ;
Vallisa, Daniele ;
Di Nunzio, Camilla ;
Cappucciati, Lorella ;
Berte, Raffaella ;
Cordani, Maria Rosa ;
Lazzaro, Antonio ;
Cremona, Gabriele ;
Biasini, Claudia ;
Muroni, Monica ;
Mordenti, Patrizia ;
Gorgni, Silvia ;
Zaffignani, Elena ;
Ambroggi, Massimo ;
Bidin, Livia ;
Palladino, Maria Angela ;
Rodino, Carmelina ;
Tibaldi, Laura .
WORLD JOURNAL OF SURGICAL ONCOLOGY, 2010, 8
[5]  
Di Carlo I, 2000, J SURG ONCOL, V73, P172, DOI 10.1002/(SICI)1096-9098(200003)73:3<172::AID-JSO11>3.0.CO
[6]  
2-Z
[7]   Totally implantable venous access devices implanted surgically - A retrospective study on early and late complications [J].
Di Carlo, I ;
Cordio, S ;
La Greca, G ;
Privitera, G ;
Russello, D ;
Puleo, S ;
Latteri, F .
ARCHIVES OF SURGERY, 2001, 136 (09) :1050-1053
[8]   Reasons for explantation of totally implantable access ports:: A multivariate analysis of 385 consecutive patients [J].
Fischer, Lars ;
Knebel, Phillip ;
Schroeder, Steffen ;
Bruckner, Thomas ;
Diener, Markus K. ;
Hennes, Roland ;
Buhl, Klaus ;
Schmied, Bruno ;
Seiler, Christoph M. .
ANNALS OF SURGICAL ONCOLOGY, 2008, 15 (04) :1124-1129
[9]   PINCH-OFF SYNDROME - A COMPLICATION OF IMPLANTABLE SUBCLAVIAN VENOUS ACCESS DEVICES [J].
HINKE, DH ;
ZANDTSTASTNY, DA ;
GOODMAN, LR ;
QUEBBEMAN, EJ ;
KRZYWDA, EA ;
ANDRIS, DA .
RADIOLOGY, 1990, 177 (02) :353-356
[10]   Insertion of Totally Implantable Venous Access Devices: An Expertise-Based, Randomized, Controlled Trial (NCT00600444) [J].
Knebel, Phillip ;
Lopez-Benitez, Ruben ;
Fischer, Lars ;
Radeleff, Boris A. ;
Stampfl, Ulrike ;
Bruckner, Thomas ;
Hennes, Roland ;
Kieser, Meinhard ;
Kauczor, Hans-Ulrich ;
Buechler, Markus W. ;
Seiler, Christoph M. .
ANNALS OF SURGERY, 2011, 253 (06) :1111-1117