External peer review quality assurance testing in von Willebrand disease: The recent experience of the United States College of American Pathologists proficiency testing program

被引:40
作者
Hayes, Timothy E.
Brandt, John T.
Chandler, Wayne L.
Eby, Charles S.
Kottke-Marchant, Kandice
Krishnan, Jayashree
Lefkowitz, Jerry B.
Olson, John D.
Rund, Chad R.
Van Cott, Elizabeth M.
Cunningham, Mark T.
机构
[1] Maine Med Ctr, Dept Pathol & Lab Med, Portland, ME 04102 USA
[2] Eli Lilly & Co, Lilly Corp Ctr, Indianapolis, IN 46285 USA
[3] Univ Washington, Dept Lab Med, Seattle, WA 98195 USA
[4] Washington Univ, Sch Med, Dept Pathol, St Louis, MO 63110 USA
[5] Cleveland Clin Fdn, Dept Clin Pathol, Cleveland, OH 44195 USA
[6] Washington Hosp Ctr, Dept Pathol, Washington, DC 20010 USA
[7] Univ Colorado, Hlth Sci Ctr, Dept Pathol, Denver, CO 80262 USA
[8] Univ Texas, Hlth Sci Ctr, Dept Pathol, San Antonio, TX 78284 USA
[9] Univ New Mexico, Dept Pathol, Hlth Sci Ctr, Albuquerque, NM 87131 USA
[10] Massachusetts Gen Hosp, Dept Pathol, Boston, MA 02114 USA
[11] Univ Kansas, Med Ctr, Dept Pathol, Kansas City, KS 66103 USA
关键词
von Willebrand factor (vWF); von Willebrand disease (vWD); hemostasis testing; quality control; quality assurance; proficiency testing;
D O I
10.1055/s-2006-947864
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The U.S. College of American Pathologists (CAP) has conducted a focused studythe proficiency testing for von Willebrand disease (MID) analysis from 2003 to 2005. This report summarizes the findings regarding the accuracy and precision of the various assays at different analyte levels, as well as the influence of the reference material used to construct the assay standard curve. The results show that testing of von Willebrand factor (vWF):antigen (vWF:Ag) and ristocetin cofactor activity (v W:RCo) is reasonably accurate, with all-method mean values falling within 3.2 and 5.6%, respectively, of the International Society on Thrombosis and Haemostasis Secondary Coagulation Standard (lot 2) assigned values. vWF:Ag measurements are reasonably precise (all-method coefficients of variation [CVs] = 10.7 to 15.1%), even at lower levels,of vWF. The highest precision was observed for immunoturbidometric assays (CVs, 6.3 to 9.7%). V W:RCo measurements are less precise (all-method CVs, 23.3 to 30.9%). The reference materials used in the standard curves for immunoturbidometric vWF:Ag assays appear to have accurately assigned v W values for the majority of commercial suppliers.
引用
收藏
页码:499 / 504
页数:6
相关论文
共 7 条
[1]  
ABILDGAARD CF, 1980, BLOOD, V56, P712
[2]  
Favaloro EJ, 1999, THROMB HAEMOSTASIS, V82, P1276
[3]  
JENNINGS I, 2003, J THROMB HAEMOST S1, V1
[4]  
Sadler JE, 2000, THROMB HAEMOSTASIS, V84, P160
[5]  
SADLER JE, 1994, THROMB HAEMOSTASIS, V71, P520
[6]   Von Willebrand disease type 1: a diagnosis in search of a disease [J].
Sadler, JE .
BLOOD, 2003, 101 (06) :2089-2093
[7]  
SADLER JE, 1994, BLOOD, V84, P676