Point dose profile measurements using solid-state detectors in characterization of Computed Tomography systems

被引:15
作者
Herrnsdorf, L. [1 ]
Bjork, M. [1 ]
Cederquist, B. [1 ]
Mattsson, C. G. [2 ]
Thungstrom, G. [2 ]
Frojdh, C. [2 ]
机构
[1] RTI Elect AB, SE-43137 Molndal, Sweden
[2] Mid Sweden Univ, SE-85170 Sundsvall, Sweden
关键词
Point dose profile measurement; CT dose; CT QA; Solid dose detector;
D O I
10.1016/j.nima.2009.03.159
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Existing solid-state detectors for point dose measurements are typically mounted in a package with a high Z material acting as the back contact causing unsymmetrical response. In the Computed Tomography (CT) dose profile application this is a major drawback. A new silicon detector has been constructed without this disturbing element. The new detector has been mounted in a cylindrical tube and its characteristics are presented in this paper. The conclusion is that this new detector design is a major improvement for correct dose and dose profile measurements in CT systems. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:223 / 225
页数:3
相关论文
共 7 条
[1]  
CEDERQUIST B, 2008, THESIS GOTEBORG U GO
[2]  
HERRNSDORF L, 2006, EUR RADIOL, V16, pC699
[3]  
IEC, 2005, 61267 IEC
[4]   National Conference on Dose Reduction in CT, with an emphasis on pediatric patients [J].
Linton, OW ;
Mettler, FA .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2003, 181 (02) :321-329
[5]  
VRIEZE TJ, 2006, DOSIMETRY COMP MEASU
[6]  
WICKMAN G, 1991, THESIS UMEA U UMEA
[7]  
2001, STRALSKYDD 118 RIKTL