Double-lumen tracheal tubes and bougies: a bench study to investigate factors that influence the risk of shearing

被引:6
作者
Luther, D. G. P. [1 ]
Robertson, H. F. [2 ]
Suchett-Kaye, I. [2 ]
Birch, A. [3 ]
Molyneux, M. [1 ]
机构
[1] Univ Hosp Bristol NHS Fdn Trust, Bristol, Avon, England
[2] Univ Bristol, Med Sch, Bristol, Avon, England
[3] Bristol Sch Anaesthesia, Anaesthet, Bristol, Avon, England
关键词
double-lumen tube placement; endobronchial intubation; SIMULATED DIFFICULT INTUBATION; ENDOTRACHEAL-TUBE; MULTIPLE-USE; OBSTRUCTION; MANAGEMENT; INTRODUCER;
D O I
10.1111/anae.14672
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Tracheal intubation with a double-lumen tube can be more challenging than with a single-lumen tube. A bougie can be used to facilitate intubation. Case reports have described fragment shearing from bougies when they are removed from the tube after intubation. These fragments have the potential to cause harm. It has even been suggested that bougies and double-lumen tubes should not be used together. We conducted a benchtop trial to investigate factors that influence the risk of shearing. We investigated three brands of double-lumen tube (each in three sizes and both lateralities) and four brands of bougie. We simulated one intubation and 29 further insertions/removals of bougie with every bougie-double-lumen tube combination. We inspected the inside of the tube for evidence of shearing after first, tenth and thirtieth removals. We found that brand of bougie, brand of double-lumen tube and size of double-lumen tube (but not its laterality) all influenced the degree of shearing. Certain bougie-double-lumen tube combinations produced a particularly high degree of shearing, so these should be avoided.
引用
收藏
页码:891 / 895
页数:5
相关论文
共 26 条
[1]   A comparison of simulated difficult intubation with multiple-use and single-use bougies in a manikin [J].
Annamaneni, R ;
Hodzovic, I ;
Wilkes, AR ;
Latto, IP .
ANAESTHESIA, 2003, 58 (01) :45-49
[2]  
Anon, 1837, LANCET, V29, P54
[3]   A practical approach to adult one-lung ventilation [J].
Ashok, V. ;
Francis, J. .
BJA EDUCATION, 2018, 18 (03) :69-74
[4]   The obstruction of an endotracheal tube by the plastic coating sheared from a stylet: A revisit [J].
Bhargava, M ;
Pothula, SNM ;
Joshi, S .
ANESTHESIOLOGY, 1998, 88 (02) :548-549
[5]   Management of unanticipated difficult intubation: a survey of current practice in the Oxford region [J].
Bokhari, A ;
Benham, SW ;
Popat, MT .
EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2004, 21 (02) :123-127
[6]   Which device should be considered the best for lung isolation: double-lumen endotracheal tube versus bronchial blockers [J].
Campos, Javier H. .
CURRENT OPINION IN ANESTHESIOLOGY, 2007, 20 (01) :27-31
[7]   OBSTRUCTION OF AN ENDOTRACHEAL-TUBE BY THE PLASTIC COATING SHEARED FROM A STYLET [J].
COOK, WP ;
SCHULTETUS, RR .
ANESTHESIOLOGY, 1985, 62 (06) :803-804
[8]  
Gardiner D, 2009, ANAESTHESIA, V64, P219, DOI [10.1111/j.1365-2044.2008.05838.x, 10.1111/j.1365-2044.2008.05837_2.x]
[9]  
Graham JM, 2013, ANAESTH INTENS CARE, V41, P127
[10]   Bougie-assisted difficult airway management in a manikin - the effect of position held on placement and force exerted by the tip [J].
Hodzovic, I ;
Wilkes, AR ;
Latto, IP .
ANAESTHESIA, 2004, 59 (01) :38-43