Effects of hemolysis and lipemia interference on kaolin-activated thromboelastography, and comparison with conventional coagulation tests

被引:12
|
作者
Tang, Ning [1 ]
Jin, Xi [1 ]
Sun, Ziyong [1 ]
Jian, Cui [1 ]
机构
[1] Huazhong Univ Sci & Technol, Tongji Hosp, Tongji Med Coll, Clin Lab, Wuhan, Hubei, Peoples R China
关键词
Coagulation test; hemolysis; lipemia; quality management; thromboelastography; FACTOR-VII; BLOOD; SPECIFICATIONS; QUALITY; PLASMA;
D O I
10.1080/00365513.2016.1271906
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
The effects of hemolysis and lipemia on thromboelastography (TEG) analysis have been scarcely evaluated in human samples, and neglected in clinical practice. We aimed to investigate the effects of in vitro mechanical hemolysis and lipemia on TEG analysis and conventional coagulation tests. Twenty-four healthy volunteers were enrolled in the study. Besides the controls, three groups with slight, moderate and severe mechanical hemolysis were constituted according to free hemoglobin (Hb) concentrations of 0.5-1.0, 2.0-6.0 and 7.0-13.0g/L, respectively; and three groups with mild, moderate and high lipemia were established according to triglyceride concentrations of approximate to 6.0, approximate to 12.0, and approximate to 18.0mmol/L, respectively. Four TEG parameters, reaction time (R), coagulation time (K), angle (), and maximum amplitude (MA), were measured alongside conventional plasma tests including prothrombin time (PT), activated partial thromboplastin time (APTT) and fibrinogen (FIB) by mechanical method, and platelet count by optical method. Results showed that the median R and MA values at moderate and severe hemolysis and K at severe hemolysis exceeded respective reference intervals, and were considered unacceptable. Median values of TEG parameters in lipemic samples were all within reference intervals. Bias values of conventional plasma tests PT, APTT and FIB in hemolyzed or lipemic samples were all lower than the Clinical Laboratory Improvement Amendments (CLIA) allowable limits. Bias values of platelet count at moderate to severe hemolysis and lipemia exceeded the CLIA allowable limits. In conclusion, the detection of TEG was in general more affected by mechanical hemolysis than plasma coagulation tests. Pre-analytical variables should be taken into account when unexpected TEG results are obtained.
引用
收藏
页码:98 / 103
页数:6
相关论文
共 35 条
  • [1] Effects of sedation on haematological, biochemical, coagulation profile, and kaolin-activated thromboelastography in rabbits
    Bassan, Tiziana
    Pastor, Josep
    Martinez-Caro, Javier
    Agulla, Beatriz
    Martorell, Jaume
    VETERINARY JOURNAL, 2025, 310
  • [2] Effects of hemolysis, bilirubin, and lipemia interference on coagulation tests detected by two analytical systems
    Nougier, Christophe
    Jousselme, Emilie
    Sobas, Frederic
    Pousseur, Valentin
    Negrier, Claude
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2020, 42 (01) : 88 - 94
  • [3] A comparison of kaolin-activated versus nonkaolin-activated thromboelastography in native and citrated blood
    Thalheimer, Ulrich
    Triantos, Christos K.
    Samonakis, Dimitrios N.
    Zambruni, Andrea
    Senzolo, Marco
    Leandro, Gloacchino
    Patch, David
    Burroughs, Andrew K.
    BLOOD COAGULATION & FIBRINOLYSIS, 2008, 19 (06) : 495 - 501
  • [4] Interference in Coagulation Testing: Focus on Spurious Hemolysis, Icterus, and Lipemia
    Lippi, Giuseppe
    Plebani, Mario
    Favaloro, Emmanuel J.
    SEMINARS IN THROMBOSIS AND HEMOSTASIS, 2013, 39 (03) : 258 - 266
  • [5] Evaluation of kaolin-activated thromboelastography and sample stability in healthy horses
    Machackova, K.
    Boselova, M.
    Vanova, I.
    Drabkova, Z.
    Doubek, J.
    VETERINARNI MEDICINA, 2018, 63 (05) : 203 - 209
  • [6] Kaolin-activated thromboelastography and standard coagulation assays in cyanotic and acyanotic infants undergoing complex cardiac surgery: a prospective cohort study
    Rizza, Alessandra
    Ricci, Zaccaria
    Pezzella, Chiara
    Favia, Isabella
    Di Felice, Giovina
    Ranucci, Marco
    Cogo, Paola
    PEDIATRIC ANESTHESIA, 2017, 27 (02) : 170 - 180
  • [7] Reference values for kaolin-activated thromboelastography in volunteers of Anhui Province in China
    Sun, Jun-Bo
    Bian, Mao-Hong
    Zhong, Tao
    Lu, Ying-Yu
    Zhu, Bang-Qiang
    Wen, Hui-Qin
    Hu, Hai-Liang
    JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2017, 31 (06)
  • [8] Comparison between thromboelastography and conventional coagulation test: Should we abandon conventional coagulation tests in polytrauma patients?
    Tur Martinez, Jaume
    Petrone, Patrizio
    Axelrad, Alexander
    Marini, Corrado P.
    CIRUGIA ESPANOLA, 2018, 96 (07): : 443 - 449
  • [9] Reference Range of Kaolin-Activated Thromboelastography (TEG) Values in Healthy Pet Rabbits (Oryctolagus cuniculus)
    Bassan, Tiziana
    Pastor, Josep
    Agulla, Beatriz
    Jornet, Oriol
    Martorell, Jaume
    ANIMALS, 2023, 13 (14):
  • [10] Dynamic Modelling of Obstetric Patient Coagulation from Kaolin-Activated Thromboelastogram Data
    Fitzmaurice, Kieran P.
    Pressly, Michelle A.
    Clermont, Gilles
    Waters, Jonathan H.
    Parker, Robert S.
    IFAC PAPERSONLINE, 2020, 53 (02): : 16329 - 16334