Pre-operative Percutaneous Nephrolithotripsy Characterisation of Kidney Stones with Second-Generation Dual-Source Dual-Energy Computed Tomography

被引:1
作者
Ali, D. K. Mella Mohd [1 ,2 ]
Mahmud, Mohd Hafizi [1 ]
Mohamad, Noor Shafini [1 ]
机构
[1] Univ Teknol MARA UiTM, Fac Hlth Sci, Ctr Med Imaging, Selangor Branch, Puncak Alam Campus, Shah Alam, Selangor, Malaysia
[2] Hosp Tengku Ampuan Afzan Pahang, Radiol Dept, Pahang, Malaysia
来源
MALAYSIAN JOURNAL OF MEDICAL SCIENCES | 2020年 / 27卷 / 05期
关键词
dual-source computed tomography; chemical composition; kidney stone; percutaneous nephrolithotripsy; URINARY CALCULI; URIC-ACID; CT; DIFFERENTIATION; MANAGEMENT; FILTRATION; DIAGNOSIS;
D O I
10.21315/mjms2020.27.5.5
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Background: The current clinical practice to manage kidney stone requires knowledge of the stone composition. However, it is often difficult to determine the actual stone composition before a stone is operatively removed from the patient. Dual-energy computed tomography (DECT) can predict urinary stone composition, but it is not widely adopted. The purpose of the study was to investigate the use of a second-generation DECT with tin or stannum (Sn) filter for characterising the kidney stones composition. Methods: Thirty-three kidney stones were scanned ex vivo using a dual-source (DS)-DECK seamier with dual-energy (DE) mode of 80/140 kVp with and without 4 nmi Sn filtration. DE ratio was calculated to determine the kidney stones composition (uric acid, calcium oxalate, calcium phosphate and cystine). The median DE ratio of the stones was compared using Wilcoxon signed rank test and the results were further correlated with semi-quantitative Fourier transform infrared (FTIR) spectroscopy analysis using Kendall's Tau test with P < 0.05 deemed to be statistically significant. Results: Second-generation DS-DECT could significantly discriminate the stones composition with and without Sn filtration (P < 0.001). The median DE ratio of uric acid, calcium oxalate and cystine stones were significantly higher with Sn filtration than those without filtration (P < 0.05). DECT results revealed significant correlation with FTIR spectroscopy analysis (r = 0.716, P < 0.001). DECT with Sn filtration showed increased performance (100% sensitivity, 0% specificity) than those without filtration (48.5% sensitivity, 0% specificity) in the detection of the kidney stone subtypes. Conclusion: In the second-generation DECT with additional Sn filtration, DECT has shown a significant performance in characterising and discriminating the kidney stone composition. This may improve diagnostic and therapy management in kidney stones cases.
引用
收藏
页码:43 / 52
页数:10
相关论文
共 32 条
  • [1] Can a dual-energy computed tomography predict unsuitable stone components for extracorporeal shock wave lithotripsy?
    Ahn, Sung Hoon
    Oh, Tae Hoon
    Seo, Ill Young
    [J]. KOREAN JOURNAL OF UROLOGY, 2015, 56 (09) : 644 - 649
  • [2] Alatab S, 2016, IRAN J KIDNEY DIS, V10, P51
  • [3] Dual-layer spectral detector CT: non-inferiority assessment compared to dual-source dual-energy CT in discriminating uric acid from non-uric acid renal stones ex vivo
    Ananthakrishnan, Lakshmi
    Duan, Xinhui
    Xi, Yin
    Lewis, Matthew A.
    Pearle, Margaret S.
    Antonelli, Jodi A.
    Goerne, Harold
    Kolitz, Elysha M.
    Abbara, Suhny
    Lenkinski, Robert E.
    Fielding, Julia R.
    Leyendecker, John R.
    [J]. ABDOMINAL RADIOLOGY, 2018, 43 (11) : 3075 - 3081
  • [4] Renal Stone Assessment with Dual-Energy Multidetector CT and Advanced Postprocessing Techniques: Improved Characterization of Renal Stone Composition-Pilot Study
    Boll, Daniel T.
    Patil, Neil A.
    Paulson, Erik K.
    Merkle, Elmar M.
    Simmons, W. Neal
    Pierre, Sean A.
    Preminger, Glenn M.
    [J]. RADIOLOGY, 2009, 250 (03) : 813 - 820
  • [5] Dawoud MM, 2017, EGYPT J RADIOL NUC M, V48, P717, DOI 10.1016/j.ejrnm.2017.03.020
  • [6] Characterization of Urinary Stone Composition by Use of Third-Generation Dual-Source Dual-Energy CT With Increased Spectral Separation
    Duan, Xinhui
    Li, Zhoubo
    Yu, Lifeng
    Leng, Shuai
    Halaweish, Ahmed F.
    Fletcher, Joel G.
    McCollough, Cynthia H.
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2015, 205 (06) : 1203 - 1207
  • [7] Clinical utility of computed tomography Hounsfield characterization for percutaneous nephrolithotomy: a cross-sectional study
    Gallioli, Andrea
    De Lorenzis, Elisa
    Boeri, Luca
    Delor, Maurizio
    Zanetti, Stefano Paolo
    Longo, Fabrizio
    Trinchieri, Alberto
    Montanari, Emanuele
    [J]. BMC UROLOGY, 2017, 17
  • [8] Unenhanced third-generation dual-source chest CT using a tin filter for spectral shaping at 100 kVp
    Haubenreisser, Holger
    Meyer, Mathias
    Sudarski, Sonja
    Allmendinger, Thomas
    Schoenberg, Stefan O.
    Henzler, Thomas
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2015, 84 (08) : 1608 - 1613
  • [9] Dual-energy CT for the evaluation of urinary calculi: Image interpretation, pitfalls and stone mimics
    Jepperson, M. A.
    Cernigliaro, J. G.
    Sella, D.
    Ibrahim, E.
    Thiel, D. D.
    Leng, S.
    Haley, W. E.
    [J]. CLINICAL RADIOLOGY, 2013, 68 (12) : E707 - E714
  • [10] In Vivo Comparison of Radiation Exposure of Dual-Energy CT Versus Low-Dose CT Versus Standard CT for Imaging Urinary Calculi
    Jepperson, Maria A.
    Cernigliaro, Joseph G.
    Ibrahim, El-Sayed H.
    Morin, Richard L.
    Haley, William E.
    Thiel, David D.
    [J]. JOURNAL OF ENDOUROLOGY, 2015, 29 (02) : 141 - 146