Failure of less invasive stabilization system plates in the distal femur - A report of four cases

被引:87
作者
Button, G [1 ]
Wolinsky, P [1 ]
Hak, D [1 ]
机构
[1] Univ Calif Davis, Dept Orthopaed Surg, Sacramento, CA 95817 USA
关键词
less invasive stabilization system; LISS; failure; distal femur fractures;
D O I
10.1097/00005131-200409000-00015
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Four cases of failure of distal femoral Less Invasive Stabilization System plates are reported. Two modes of failure were observed: in 2 cases, the Less Invasive Stabilization System plates broke, and in 2 cases, the plates gradually displaced from the femur as the shaft screws lost purchase in the diaphyseal femoral cortex. Delayed union/nonunion, early weight bearing, and improper plate placement were identified as the etiologies for plate failure.
引用
收藏
页码:565 / 570
页数:6
相关论文
共 11 条
  • [1] Operative stabilization of supracondylar femur fractures above total knee arthroplasty - A comparison of four treatment methods
    Althausen, PL
    Lee, MA
    Finkemeier, CG
    Meehan, JP
    Rodrigo, JJ
    [J]. JOURNAL OF ARTHROPLASTY, 2003, 18 (07) : 834 - 839
  • [2] The development of the distal femur Less Invasive Stabilization System (LISS)
    Frigg, R
    Appenzeller, A
    Christensen, R
    Frenk, A
    Gilbert, S
    Schavan, R
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2001, 32 : 24 - 31
  • [3] Indirect reduction and bridge plating of supracondylar fractures of the femur
    Huang, HT
    Huang, PJ
    Su, JY
    Lin, SY
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2003, 34 (02): : 135 - 140
  • [4] New technique for treatment of unstable distal femur fractures by locked double-plating: Case report and biomechanical evaluation
    Jazrawi, LM
    Kummer, FJ
    Simon, JA
    Bai, B
    Hunt, SA
    Egol, KA
    Koval, KJ
    [J]. JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2000, 48 (01) : 87 - 92
  • [5] Treatment of multifragmentary fractures of the femur by indirect reduction (biological) and plate fixation
    Kesemenli, C
    Subasi, M
    Necmioglu, S
    Kapukaya, A
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2002, 33 (08): : 691 - 699
  • [6] Distal femoral fracture fixation utilizing the Less Invasive Stabilization System (LISS): The technique and early results
    Kregor, PJ
    Stannard, J
    Zlowodzki, M
    Cole, PA
    Alonso, J
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2001, 32 : 32 - 47
  • [7] Fixation of distal femoral fractures above total knee arthroplasty utilizing the Less Invasive Stabilization System (LISS)
    Kregor, PJ
    Hughes, JL
    Cole, PA
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2001, 32 : 64 - 75
  • [8] Biomechanical evaluation of the less invasive stabilization system for the internal fixation of distal femur fractures
    Marti, A
    Fankhauser, C
    Frenk, A
    Cordey, J
    Gasser, B
    [J]. JOURNAL OF ORTHOPAEDIC TRAUMA, 2001, 15 (07) : 482 - 487
  • [9] Mechanical comparison of endosteal substitution and lateral plate fixation in supracondylar fractures of the femur
    Prayson, MJ
    Datta, DK
    Marshall, MP
    [J]. JOURNAL OF ORTHOPAEDIC TRAUMA, 2001, 15 (02) : 96 - 100
  • [10] Distal femoral fractures and LISS stabilization
    Schandelmaier, P
    Partenheimer, A
    Koenemann, B
    Grün, OA
    Krettek, C
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2001, 32 : 55 - 63