Evaluation of a point-of-care glucose meter for general use in complex tertiary care facilities

被引:19
作者
Chan, Pak Cheung R. [1 ,2 ]
Rozmanc, Mary [1 ]
Selden-Long, Isolde [2 ,3 ]
Kwan, Jason [1 ]
机构
[1] Sunnybrook Hlth Sci Ctr, Dept Clin Pathol, Toronto, ON M4N 3M5, Canada
[2] Univ Toronto, Dept Pathobiol & Lab Med, Toronto, ON M5S 1A1, Canada
[3] Gamma Dynacare Med Labs, Brampton, ON, Canada
关键词
Glucose meter; Method evaluation; Glycemic control; Tertiary care; INTENSIVE INSULIN THERAPY; BLOOD-GLUCOSE; PERITONEAL-DIALYSIS; QUALITY; PATIENT; INTERFERENCE; HEMATOCRIT; TEMPERATURE; PERFORMANCE; ATTITUDES;
D O I
10.1016/j.clinbiochem.2009.03.023
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objective: To evaluate the performance of the StatStrip (TM) glucose meter from Nova Biomedical for use in complex tertiary care facilities. Methods: Performance evaluation was conducted in 6 clinical locations involving nurse end-users. Imprecision (CV)<5% was considered acceptable. Ten substances were examined for potential analytical interferences at two levels and 3 glucose concentrations. Inaccuracy was determined by 386 paired glucose differences between glucose meter and laboratory reference analyses. Acceptability was based on CSLI/ISO15197 criteria. Potential clinical impact was classified under a 5-level severity scheme. Results: Imprecision varied from CV 2.4 to 4.9%. Significant interference was observed only for free hemoglobin at 10 g/L. 97% of paired glucose meter-laboratory differences met CSLI/ISO15197 criteria. All discrepant results had low severity scores. Conclusion: The StatStripT (TM) glucose meter demonstrated acceptable imprecision and inaccuracy, and was relatively free from common interferences. It should be a good candidate for general use in complex tertiary care facilities. (C) 2009 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:1104 / 1112
页数:9
相关论文
共 52 条
  • [1] *AM DIAB ASS, 2000, DIABETES CARE S1, V23, pS24
  • [2] Amer Diabet Assoc, 2000, DIABETES CARE, V23, pS80
  • [3] [Anonymous], 1992, Fed Regist, V57, P7002
  • [4] [Anonymous], 1996, DIABETES CARE, V19, pS62
  • [5] [Anonymous], 2003, 151972003E ISO
  • [6] FINGERSTICK GLUCOSE DETERMINATION IN SHOCK
    ATKIN, SH
    DASMAHAPATRA, A
    JAKER, MA
    CHOROST, MI
    REDDY, S
    [J]. ANNALS OF INTERNAL MEDICINE, 1991, 114 (12) : 1020 - 1024
  • [7] Boyd JC, 2001, CLIN CHEM, V47, P209
  • [8] *CAN DIAB ASS CLIN, 2003, CAN J DIABETES S2, V27, pS113
  • [9] *CAN DIAB ASS CLIN, 2003, CAN J DIABETES S2, V27, pS21
  • [10] Reference change values and power functions
    Canadell, NI
    Petersen, PH
    Jensen, E
    Ricós, C
    Jorgensen, PE
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2004, 42 (04) : 415 - 422