Comparison of Radiation Dose and Image Quality with Various Computed-Tomography Scout Views: The Angular Modulation Technique Based on Information Calculated from Scout Views

被引:0
作者
Kim, Myeong Seong [1 ,2 ,3 ]
Lee, Jong-Woong [4 ]
Kim, Sun Geun [5 ]
Kweon, Dae Cheol [6 ]
机构
[1] Korean Natl Canc Ctr Goyang, Dept Radiol, Goyang, South Korea
[2] Seoul Natl Univ, Grad Sch Publ Hlth, Seoul, South Korea
[3] Seoul Natl Univ, Inst Hlth & Environm, Seoul, South Korea
[4] Kyung Hee Univ Hosp, Dept Radiol, Seoul, South Korea
[5] Baekseok Culture Univ, Dept Radiol Technol, Cheonan, South Korea
[6] Shinhan Univ, Dept Radiol Sci, Uijongbu, South Korea
关键词
Automatic Exposure Control; Computed Tomography; Radiation Dose; Scout View; TUBE-CURRENT MODULATION; CT; REDUCTION; EXPOSURE; STRATEGIES; PHANTOM;
D O I
10.5812/iranjradiol.35606
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: While using the automatic exposure control technique in computed tomography (CT), the scout view determines the radiation dose for the CT scan by taking anatomical morphology into consideration. Objectives: The goal of this study was to estimate radiation doses during scout scanning and to evaluate the hypothesis that the radiation dose during CT depends on the location and order of the X-ray tubes. Materials and Methods: A CT scanner (GE VCT 64-channel) with the X-ray tubes placed in the anteroposterior (AP), lateral, and posteroanterior (PA) positions was used to acquire the scout views. The effective doses were assessed using an anthropomorphic chest phantom (RSD Inc., Long Beach, CA), an AAPM CT performance phantom (CIRS model 610), and CT dose indices (CTDI) head phantom. Both single and two scout views were tested with the three views combined according to the order of scanning. The CTDIs obtained from the CT unit dose report were used to compare the chest CT radiation doses, while the image quality was evaluated based on the root-mean-square error values calculated using the Image program. Results: The results of this study indicate that single AP (0 degrees) scout views obtained with the z-axis modulation technique (AutomA) during chest CT scanning resulted in CTDIs that were lower than those generated by using lateral (90 degrees) or PA (180 degrees) scout views. The analysis of two-scout views using AutomA alone revealed the lowest CTDI in the 90 degrees - 0 degrees scout views. However, based on the CT image noise analysis, a scout angle order of 0 degrees-90 degrees resulted in the optimal scout view combination with the minimal dose. Conclusion: Scout views may be useful in determining dose and image quality in CT scanning; however, it is necessary to first determine the most appropriate scout view for CT scans.
引用
收藏
页数:11
相关论文
共 16 条
  • [1] Automatic tube-current modulation in CT -: A comparison between different solutions
    Althén, JN
    [J]. RADIATION PROTECTION DOSIMETRY, 2005, 114 (1-3) : 308 - 312
  • [2] Current concepts - Computed tomography - An increasing source of radiation exposure
    Brenner, David J.
    Hall, Eric J.
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (22) : 2277 - 2284
  • [3] ICRP, 2010, ICRP PUBL, V87
  • [4] Dose reduction in CT by anatomically adapted tube current modulation. II. Phantom measurements
    Kalender, WA
    Wolf, H
    Suess, C
    [J]. MEDICAL PHYSICS, 1999, 26 (11) : 2248 - 2253
  • [5] Reducing radiation dose in emergency computed tomography with automatic exposure control techniques
    Kalra M.K.
    Rizzo S.M.R.
    Novelline R.A.
    [J]. Emergency Radiology, 2005, 11 (5) : 267 - 274
  • [6] Strategies for CT radiation dose optimization
    Kalra, MK
    Maher, MM
    Toth, TL
    Hamberg, LM
    Blake, MA
    Shephard, JA
    Saini, S
    [J]. RADIOLOGY, 2004, 230 (03) : 619 - 628
  • [7] Radiation dose modulation techniques in the multidetector CT era: From basics to practice
    Lee, Chang Hyun
    Goo, Jin Mo
    Lee, Hyun Ju
    Ye, Sung-Joon
    Park, Chang Min
    Chun, Eun Ju
    Im, Jung-Gi
    [J]. RADIOGRAPHICS, 2008, 28 (05) : 1451 - 1459
  • [8] Impact of new technologies on dose reduction in CT
    Lee, Ting-Yim
    Chhem, Rethy K.
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2010, 76 (01) : 28 - 35
  • [9] Radiation Dose Reduction at CT Enterography: How Low Can We Go While Preserving Diagnostic Accuracy?
    Leng, Shuai
    Yu, Lifeng
    McCollough, Cynthia H.
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 195 (01) : 76 - 77
  • [10] Patient dose reduction in CT
    Lewis, MA
    Edyvean, S
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2005, 78 (934) : 880 - 883