Comparison of 2-Dimensional Shear Wave Elastographic Measurements Using ElastQ Imaging and SuperSonic Shear Imaging: Phantom Study and Clinical Pilot Study

被引:12
作者
Hwang, Jisun [1 ]
Yoon, Hee Mang [2 ,3 ]
Jung, Ah Young [2 ,3 ]
Lee, Jin Seong [2 ,3 ]
Cho, Young Ah [2 ,3 ]
机构
[1] Hallym Univ, Med Ctr, Dongtan Sacred Heart Hosp, Dept Radiol, Hwaseong, South Korea
[2] Univ Ulsan, Asan Med Ctr, Coll Med, Dept Radiol, 88 Olymp Ro 43 Gil, Seoul 05505, South Korea
[3] Univ Ulsan, Asan Med Ctr, Res Inst Radiol, Coll Med, 88 Olymp Ro 43 Gil, Seoul 05505, South Korea
关键词
liver; phantom study; shear wave elastography; ultrasound; LIVER FIBROSIS; ULTRASOUND ELASTOGRAPHY; SPEED MEASUREMENTS; HEPATITIS-C; STIFFNESS; CHILDREN; REPEATABILITY; ACCURACY; US;
D O I
10.1002/jum.15108
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Objectives To compare the feasibility and clinical applicability of ElastQ imaging (Philips Healthcare, Best, the Netherlands) with that of SuperSonic shear imaging (SSI; SuperSonic Imagine, Aix-en-Provence, France) using an elastographic phantom and a pilot study of patients. Methods Two-dimensional shear wave elastography measurement was performed by ElastQ imaging and SSI by 2 radiologists. An elastographic phantom with 5 target elasticities at 2 acquisition depths was used. The coefficients of variation and intraclass correlation coefficients (ICCs) were evaluated for repeatability and interobserver agreement, respectively. The mean elasticities of the systems at each target were compared. The proportions of measurements that were out of the range of expected values and measurement errors were calculated to determine accuracy. Liver stiffness (LS) was measured by both systems in 27 children and young adult patients with various liver diseases. Results Both systems provided high repeatability in elasticity measurements of phantom targets (coefficients of variation, 0.69%-15.82%), and there was excellent interobserver agreement (ICC, 0.992). Most (90%) mean elasticities of targets were significantly different between the techniques (P <= .002) and acquisition depths (P <= .004). ElastQ imaging had significantly lower proportions of out-of-range measurements and measurement errors (P <= .003) than SSI. In patients with liver disease, LS measurements of the systems were strongly correlated (rho = 0.955; P < .001) and had excellent agreement (ICC, 0.951; P < .001). Conclusions ElastQ imaging had comparably good results in terms of repeatability, interobserver agreement, and accuracy in the phantom model compared with SSI. The pilot patient study showed strong correlations in LS values between the systems.
引用
收藏
页码:311 / 321
页数:11
相关论文
共 24 条
[1]   NONINVASIVE IN VIVO LIVER FIBROSIS EVALUATION USING SUPERSONIC SHEAR IMAGING: A CLINICAL STUDY ON 113 HEPATITIS C VIRUS PATIENTS [J].
Bavu, Eric ;
Gennisson, Jean-Luc ;
Couade, Mathieu ;
Bercoff, Jeremy ;
Mallet, Vincent ;
Fink, Mathias ;
Badel, Anne ;
Vallet-Pichard, Anais ;
Nalpas, Bertrand ;
Tanter, Mickael ;
Pol, Stanislas .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2011, 37 (09) :1361-1373
[2]   Shear-wave Elastography Improves the Specificity of Breast US: The BE1 Multinational Study of 939 Masses [J].
Berg, Wendie A. ;
Cosgrove, David O. ;
Dore, Caroline J. ;
Schaefer, Fritz K. W. ;
Svensson, William E. ;
Hooley, Regina J. ;
Ohlinger, Ralf ;
Mendelson, Ellen B. ;
Balu-Maestro, Catherine ;
Locatelli, Martina ;
Tourasse, Christophe ;
Cavanaugh, Barbara C. ;
Juhan, Valerie ;
Stavros, A. Thomas ;
Tardivon, Anne ;
Gay, Joel ;
Henry, Jean-Pierre ;
Cohen-Bacrie, Claude .
RADIOLOGY, 2012, 262 (02) :435-449
[3]   A Comparative Study of Strain and Shear-Wave Elastography in an Elasticity Phantom [J].
Carlsen, Jonathan F. ;
Pedersen, Malene R. ;
Ewertsen, Caroline ;
Saftoiu, Adrian ;
Lonn, Lars ;
Rafaelsen, Soren R. ;
Nielsen, Michael B. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2015, 204 (03) :W236-W242
[4]   Variability of ShearWave Velocity using Different Frequencies in Acoustic Radiation Force Impulse (ARFI) Elastography: A Phantom and Normal Liver Study [J].
Chang, S. ;
Kim, M. -J. ;
Kim, J. ;
Lee, M. -J. .
ULTRASCHALL IN DER MEDIZIN, 2013, 34 (03) :260-265
[5]  
Dietrich CF, 2017, ULTRASCHALL MED, V38, pE52, DOI [10.1055/a-0641-0076, 10.1055/s-0043-103952]
[6]   Ultrasound shear wave speed measurements correlate with liver fibrosis in children [J].
Dillman, Jonathan R. ;
Heider, Amer ;
Bilhartz, Jacob L. ;
Smith, Ethan A. ;
Keshavarzi, Nahid ;
Rubin, Jonathan M. ;
Lopez, M. James .
PEDIATRIC RADIOLOGY, 2015, 45 (10) :1480-1488
[7]   Superficial ultrasound shear wave speed measurements in soft and hard elasticity phantoms: repeatability and reproducibility using two ultrasound systems [J].
Dillman, Jonathan R. ;
Chen, Shigao ;
Davenport, Matthew S. ;
Zhao, Heng ;
Urban, Matthew W. ;
Song, Pengfei ;
Watcharotone, Kuanwong ;
Carson, Paul L. .
PEDIATRIC RADIOLOGY, 2015, 45 (03) :376-385
[8]   LIVER ULTRASOUND ELASTOGRAPHY: AN UPDATE TO THE WORLD FEDERATION FOR ULTRASOUND IN MEDICINE AND BIOLOGY GUIDELINES AND RECOMMENDATIONS [J].
Ferraioli, Giovanna ;
Wong, Vincent Wai-Sun ;
Castera, Laurent ;
Berzigotti, Annalisa ;
Sporea, Loan ;
Dietrich, Christoph F. ;
Choi, Byung Ihn ;
Wilson, Stephanie R. ;
Kudo, Masatoshi ;
Barr, Richard G. .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2018, 44 (12) :2419-2440
[9]   Evaluation of Inter-System Variability in Liver Stiffness Measurements [J].
Ferraioli, Giovanna ;
De Silvestri, Annalisa ;
Lissandrin, Raffaella ;
Maiocchi, Laura ;
Tinelli, Carmine ;
Filice, Carlo ;
Barr, Richard G. .
ULTRASCHALL IN DER MEDIZIN, 2019, 40 (01) :64-75
[10]   Accuracy of Real-Time Shear Wave Elastography for Assessing Liver Fibrosis in Chronic Hepatitis C: A Pilot Study [J].
Ferraioli, Giovanna ;
Tinelli, Carmine ;
Dal Bello, Barbara ;
Zicchetti, Mabel ;
Filice, Gaetano ;
Filice, Carlo .
HEPATOLOGY, 2012, 56 (06) :2125-2133