Realization of the spectral responsivity scale based on an absolute cryogenic radiometer

被引:7
|
作者
Hong, Kee-Suk [1 ]
Park, Seongchong [1 ]
Kim, Seung-Kwan [1 ]
Park, Seung-Nam [1 ]
Lee, Dong-Hoon [1 ]
机构
[1] Korea Res Inst Stand & Sci, Ctr Photometry & Radiometry, Daejeon 34186, South Korea
关键词
Spectral responsivity; Cryogenic radiometer; Trap detector; Measurement uncertainty; TRAP DETECTORS; NM;
D O I
10.3938/jkps.67.2045
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
We present the concept of and the procedure for realizing the primary scale of the spectral responsivity based on an absolute cryogenic radiometer at the Korea Research Institute of Standards and Science (KRISS). A detailed error and uncertainty analysis is performed by calibrating the standard detectors in the wavelength range from 300 nm to 1000 nm. The relative expanded uncertainty of the spectral responsivity calibration for a Si trap detector is 0.06% at 632.8 nm and 0.4% for other wavelengths (k = 2). The calibration results are validated by comparing with the results from a theoretical model of quantum efficiency for a transmission-type trap detector.
引用
收藏
页码:2045 / 2058
页数:14
相关论文
共 50 条
  • [1] Realization of the spectral responsivity scale based on an absolute cryogenic radiometer
    Kee-Suk Hong
    Seongchong Park
    Seung-Kwan Kim
    Seung-Nam Park
    Dong-Hoon Lee
    Journal of the Korean Physical Society, 2015, 67 : 2045 - 2058
  • [2] Absolute spectral responsivity of silicon trap detector based on absolute cryogenic radiometer
    Shi, Xueshun
    Liu, Changming
    Liu, Yulong
    Yang, Lechen
    Zhao, Kun
    Chen, Haidong
    INTERNATIONAL CONFERENCE ON PHOTONICS AND OPTICAL ENGINEERING (ICPOE 2014), 2015, 9449
  • [3] Long-Wave Infrared Absolute Spectral Responsivity Scale by Using an Absolute Cryogenic Radiometer
    Liu Hong-bo
    Shi Xue-shun
    Xu Wen-bin
    Liu Chang-ming
    Liu Hong-yuani
    Wang Heng-fei
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2020, 40 (12) : 3680 - 3685
  • [4] Mid-infrared absolute spectral responsivity scale based on an absolute cryogenic radiometer and an optical parametric oscillator laser
    Zhao, Kun
    Shi, Xueshun
    Chen, Haidong
    Liu, Yulong
    Liu, Changming
    Chen, Kunfeng
    Li, Ligong
    Gan, Haiyong
    Ma, Chong
    METROLOGIA, 2016, 53 (03) : 981 - 985
  • [5] The technique of measuring absolute spectral responsivity of InGaAs detector based on cryogenic radiometer
    Fan, JH
    Yang, ZJ
    Hou, XQ
    Qin, Y
    Li, ZQ
    2ND INTERNATIONAL SYMPOSIUM ON ADVANCED OPTICAL MANUFACTURING AND TESTING TECHNOLOGIES: OPTICAL TEST AND MEASUREMENT TECHNOLOGY AND EQUIPMENT, PTS 1 AND 2, 2006, 6150
  • [6] Mid-infrared spectral responsivity scale based on an absolute cryogenic radiometer and a tunable quantum cascade laser
    Liu, Wende
    Xu, Nan
    He, Yingwei
    Gan, Haiyong
    Feng, Guojin
    Wang, Yanfei
    Liu, Xiangliang
    Fu, Yangting
    Wang, Yingce
    Lin, Yandong
    METROLOGIA, 2021, 58 (02)
  • [7] Precise scale of spectral responsivity for InGaAs detectors based on a cryogenic radiometer and several laser sources
    Werner, L
    Friedrich, R
    Johannsen, U
    Steiger, A
    METROLOGIA, 2000, 37 (05) : 523 - 526
  • [8] Calibration of absolute spectral responsivity at 1 064 nm of transfer detector against cryogenic radiometer
    Pang, Weiwei
    Zheng, Xiaobing
    Li, Jianjun
    Shi, Xueshun
    Wu, Haoyu
    Xia, Maopeng
    Gao, Dongyang
    Shi, Jianmin
    Qi, Tao
    Kang, Qing
    Hongwai yu Jiguang Gongcheng/Infrared and Laser Engineering, 2015, 44 (12): : 3812 - 3818
  • [9] Filter radiometer based realization of photometric scale traceable to cryogenic radiometer at UME
    Durak, M
    Samedov, F
    Optical Measurement Systems for Industrial Inspection IV, Pts 1 and 2, 2005, 5856 : 972 - 979
  • [10] A METHOD OF REALIZING SPECTRAL IRRADIANCE BASED ON AN ABSOLUTE CRYOGENIC RADIOMETER
    JOHNSON, BC
    CROMER, CL
    SAUNDERS, RD
    EPPELDAUER, G
    FOWLER, J
    SAPRITSKY, VI
    DEZSI, G
    METROLOGIA, 1993, 30 (04) : 309 - 315