Transintercostal-Evoked Spinal Cord Potential in Thoracic Aortic Surgery

被引:4
作者
Takahashi, Shinya [1 ]
Orihashi, Kazumasa [2 ]
Imai, Katsuhiko [1 ]
Mizukami, Taketomo [1 ]
Takasaki, Taiichi [1 ]
Sueda, Taijiro [1 ]
机构
[1] Hiroshima Univ Hosp, Dept Cardiovasc Surg, Hiroshima, Japan
[2] Kochi Univ, Fac Med, Dept Surg 2, Kochi 780, Japan
关键词
THORACOABDOMINAL AORTA; ANEURYSM REPAIR; ANGIOGRAPHY; ADAMKIEWICZ; PARAPLEGIA; ARTERY;
D O I
10.1016/j.avsg.2014.04.008
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background: We have developed a novel method of measuring spinal cord-evoked potentials with stimulation and recording at the intercostal nerves (transintercostal-evoked spinal cord potential: Tic-ESCP). The purpose of this study was to examine the feasibility and accuracy of Tic-ESCP during thoracic aortic surgery. Methods: In addition to the conventional electrodes (cranial and intrathecal), stimulating and recording electrodes were placed on the intercostal nerves that were located at a cephalad and caudal level relative to the aneurysm after the pleura on the intercostal nerves was opened. Specially designed hook-type electrodes were anchored to the nerves and surroundings atraumatically and fixed on the pleura. The conventional transcranial motor-evoked potential (Tc-MEP) and Tic-ESCP were recorded simultaneously. Eight patients were examined in this study. Results: In all patients, Tic-ESCP could be clearly recorded with biphasic waveforms consisting of first a positive wave and a subsequent negative wave. In all 8 patients, the waveform of Tc-MEP and Tic-ESCP changed during aortic reconstruction. In 2 cases, the waveform of Tc-MEP and Tic-ESCP decreased below 50% of baseline during aortic clamping and the intercostal arteries were reconstructed with no resultant spinal cord injury. In 1 case with a shaggy aorta, Tc-MEP and Tic-ESCP had different values and each evoked potential could have reflected that regional spinal cord infarction and paraplegia had occurred. Conclusions: Tic-ESCP was clinically feasible and changes were compatible with the conventional Tc-MEP. The Tic-ESCP waveforms were simple and appeared to be specific to the spinal cord within the target range, in contrast to the other evoked potentials which are multimorphic and reflect the amplitudes at the brain and multiple levels of the spinal cord.
引用
收藏
页码:1775 / 1781
页数:7
相关论文
共 50 条
[41]   Spinal cord protection during thoracoabdominal aortic replacement: spinal cord perfusion maintenance [J].
Uchino, Gaku ;
Yunoki, Keiji ;
Sakoda, Naoya ;
Hattori, Shigeru ;
Kawabata, Takuya ;
Saiki, Munehiro ;
Fujita, Yasufumi ;
Hisamochi, Kunikazu ;
Yoshida, Hideo .
INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY, 2017, 24 (05) :708-713
[42]   Prevention of reperfusion injury of the spinal cord in aortic surgery:: An experimental study [J].
Kocogullari, Cevdet Ugur ;
Becit, Necip ;
Erkut, Bilgehan ;
Keles, M. Sait ;
Ceviz, Munacettin ;
Ates, Azman ;
Gundogdu, Cemal ;
Kaygin, Mehmet Ali ;
Kocak, Hikmet .
SURGERY TODAY, 2008, 38 (03) :237-244
[43]   Spinal cord protection in thoracoabdominal aortic aneurysm surgery: a multimodal approach [J].
Haunschild, Josephina ;
Von Aspern, Konstantin ;
Misfeld, Martin ;
Davierwala, Piroze ;
Borger, Michael A. ;
Etz, Christian D. .
JOURNAL OF CARDIOVASCULAR SURGERY, 2021, 62 (04) :316-325
[44]   Prevention of reperfusion injury of the spinal cord in aortic surgery: An experimental study [J].
Cevdet Ugur Koçogullari ;
Necip Becit ;
Bilgehan Erkut ;
M. Sait Keleş ;
Munacettin Ceviz ;
Azman Ates ;
Cemal Gündoğdu ;
Mehmet Ali Kaygin ;
Hikmet Koçak .
Surgery Today, 2008, 38 :237-244
[45]   Prevention of spinal cord ischaemic complications after thoracoabdominal aortic surgery [J].
Cambria, RP ;
Giglia, JS .
EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 1998, 15 (02) :96-109
[46]   Current strategies of spinal cord protection during thoracoabdominal aortic surgery [J].
Tanaka, Akiko ;
Safi, Hazim J. ;
Estrera, Anthony L. .
GENERAL THORACIC AND CARDIOVASCULAR SURGERY, 2018, 66 (06) :307-314
[47]   Commentary: A conceptual roadmap to spinal cord protection in thoracoabdominal aortic surgery [J].
Rosati, Carlo ;
Di Luozzo, Gabriele .
JTCVS TECHNIQUES, 2021, 8 :18-19
[48]   Intrathecal Lactate Concentration and Spinal Cord Injury in Thoracoabdominal Aortic Surgery [J].
Casiraghi, Giuseppina ;
Poli, Davide ;
Landoni, Giovanni ;
Buratti, Luca ;
Imberti, Roberto ;
Plumari, Valentina ;
Turi, Stefano ;
Mennella, Roberta ;
Messina, Melissa ;
Covello, Remo Daniel ;
Carozzo, Andrea ;
Motta, Andrea ;
Zangrillo, Alberto .
JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2011, 25 (01) :120-126
[49]   The monitoring and preventing of spinal cord ischemic injury during thoracic descending aortic operating [J].
Liu, Feng ;
Guan, Yulong ;
Wan, Caihong ;
Dong, Peiqing .
SCANDINAVIAN CARDIOVASCULAR JOURNAL, 2012, 46 (04) :239-244
[50]   Rescue therapy for symptomatic spinal cord ischemia after thoracic endovascular aortic repair [J].
Spratt, John R. ;
Walker, Kristen L. ;
Neal, Dan ;
Arnaoutakis, George J. ;
Martin, Tomas D. ;
Back, Martin R. ;
Zasimovich, Yury ;
Franklin, Michael ;
Shahid, Zain ;
Upchurch Jr, Gilbert R. ;
Scali, Salvatore T. ;
Beaver, Thomas M. .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2024, 168 (01)