Ionophore- Nafion™ modified gold-coated electrospun polymeric fibers electrodes for determination of electrolytes

被引:17
作者
Aldea, Anca [1 ,2 ]
Matei, Elena [1 ]
Leote, Ricardo J. B. [1 ,3 ]
Rau, Ileana [2 ]
Enculescu, Ionut [1 ]
Diculescu, Victor C. [1 ]
机构
[1] Natl Inst Mat Phys, Atomistilor 405A, Magurele 077125, Romania
[2] Univ Politehn Bucuresti, Fac Appl Chem & Mat Sci, Gheorghe Polizu 1-7, Bucharest 011061, Romania
[3] Univ Bucharest, Fac Phys, Atomistilor 405, Magurele 077125, Romania
关键词
Electrospinning; Ionophore; Sensor; Flexible; Sweat; ION-SELECTIVE ELECTRODES; BILAYER-LIPID MEMBRANES; AMMONIUM; SENSOR; CA2+;
D O I
10.1016/j.electacta.2020.137239
中图分类号
O646 [电化学、电解、磁化学];
学科分类号
081704 ;
摘要
The present work describes the electrochemical properties of ionophores immobilized at the surface of electrodes obtained from electrospun polymeric fibers, in order to develop sensors for the analysis of electrolytes. Poly(methyl methacrylate) submicrometer fibers were prepared by electrospinning, coated with a gold layer by magnetron sputtering and then transferred on polyethylene terephthalate (PET) in order to obtained flexible electrodes. The ionophores were immobilized at the surface of these electrodes by drop-casting a ionophore-Nafion mixed solution. The sensor surface was investigated by scanning electron microscopy coupled with energy-dispersive X-ray spectroscopy in order to understand the morphology and distribution of a model Ca2+ ionophore over the electrode surface. Also, Fourier-transformed infrared spectroscopy was performed and demonstrated that the model Ca2+ ionophore can be immobilized in the nafion matrix maintaining its conformation, while cyclic voltammetry and electrochemical impedance spectroscopy demonstrated that the Ca2+ ionophore allows the diffusion of target ions through this this type of membrane. In order to prove the concept of ionophore-based sensors for the analysis of some electrolytes, Ca2+, NH4+, Cl- and H+ ionophore immobilized in a nafion matrix at the surface of these flexible electrodes were tested and the determination of the target ions performed by potentiometry in different media including artificial sweat. Finally, sensitivities, limits of detection, selectivity coefficients were determined. (C) 2020 Elsevier Ltd. All rights reserved.
引用
收藏
页数:11
相关论文
共 45 条
[1]   A Ca2+ selective membrane electrode based on calcium-imprinted polymeric nanoparticles [J].
Alizadeh, Taher ;
Shamkhali, Amir Naser ;
Hanifehpour, Younes ;
Joo, Sang Woo .
NEW JOURNAL OF CHEMISTRY, 2016, 40 (10) :8479-8487
[2]   Review on Wearable Technology Sensors Used in Consumer Sport Applications [J].
Aroganam, Gobinath ;
Manivannan, Nadarajah ;
Harrison, David .
SENSORS, 2019, 19 (09)
[3]   PVC membrane ion-selective electrodes for the determination of hyoscyamine in pure solution and in pharmaceutical preparations under batch and flow modes [J].
Badawy, SS ;
Issa, YM ;
Mutair, AA .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2005, 39 (1-2) :117-124
[4]   Smart Health Monitoring Systems: An Overview of Design and Modeling [J].
Baig, Mirza Mansoor ;
Gholamhosseini, Hamid .
JOURNAL OF MEDICAL SYSTEMS, 2013, 37 (02)
[5]  
Bakker E, 1999, ELECTROANAL, V11, P915, DOI 10.1002/(SICI)1521-4109(199909)11:13<915::AID-ELAN915>3.0.CO
[6]  
2-J
[7]   Potentiometric ion sensors [J].
Bobacka, Johan ;
Ivaska, Ari ;
Lewenstam, Andrzej .
CHEMICAL REVIEWS, 2008, 108 (02) :329-351
[8]   Wearable flexible sweat sensors for healthcare monitoring: a review [J].
Chung, Michael ;
Fortunato, Giuseppino ;
Radacsi, Norbert .
JOURNAL OF THE ROYAL SOCIETY INTERFACE, 2019, 16 (159)
[9]   Ion selective textile organic electrochemical transistor for wearable sweat monitoring [J].
Coppede, Nicola ;
Giannetto, Marco ;
Villani, Marco ;
Lucchini, Valeria ;
Battista, Edmondo ;
Careri, Maria ;
Zappettini, Andrea .
ORGANIC ELECTRONICS, 2020, 78
[10]   Wearable Potentiometric Sensors for Medical Applications [J].
Cuartero, Maria ;
Parrilla, Marc ;
Crespo, Gaston A. .
SENSORS, 2019, 19 (02)